2D polymer liquid chromatography is a powerful tool for the deformulation of complex samples; however, it is considered to
be very specialized and time-consuming. This article shows how recent hardware and software improvements have led to the technique
becoming a method for routine analysis.
The demands on chromatography, regarding the information density of the results, increase with the complexity of the samples
under investigation. Every single separation method suffers from limited chromatographic resolution (peak purity, limited
peak capacity, determination of multiple property distributions etc.). Many attempts have been made to overcome the intrinsic
limitations of the chromatographic process.
A powerful approach is to combine different separation techniques into a single experiment (multidimensional chromatography;
also known as 2D chromatography, orthogonal chromatography and cross-fractionation).
The combination of independent separation techniques improves the chromatographic resolution because of increased peak capacity,
which is not the sum but the product of the peak capacities of each technique:1
The separation "angle" between two techniques that separate strictly according to different properties is considered as
90° and the separation is referred to as "orthogonal".
An advantage of chromatographic hyphenation over performing two isolated separation methods is the ability to correlate and
identify the properties of different sample components.
 Figure 1
|
The mapping of four 2D experiments is illustrated in Figure 1 — the separation according to one property is plotted versus
the separation according to a second property. As can be seen, projections to the axes of each "contour plot" are the same
for all examples. The traces of those projections would also result from isolated, one-dimensional separations. Therefore,
any traditional separation optimization strategies would fail in all of these situations. Nevertheless, the direct combination
of both separation techniques can easily show the sample differences. The 2D separation is achieved by a consecutive separation
of fractions eluting from the 'first dimension' using the 'second dimension' technique. The resulting contour plots can be
read in a similar way to a geographical map that shows the first dimension on the y-axis and the second dimension on the x-axis.
This type of separation is extremely useful for any kind of complex samples, which are difficult to separate, identify comprehensively
or fully deformulate. This is most often the situation for synthetic, natural and biopolymers (monodisperse proteins are one
well-known exception), which possess coexisting multiple property distributions [such as molar mass distribution, functionality
type distribution, chemical composition distribution, molecular architecture (shape) distribution, to name just a few].
This article will focus on comprehensive 2D polymer liquid chromatography (LC) coupling a liquid adsorption mode technique
(LAC) with size-exclusion chromatography/gel permeation chromatography (SEC/GPC), and on the options to use them as a high-throughput
application.
Since the 1980s, several groups have tried to implement this technique for polymer analysis. However, it was seen as a very
time-consuming technique that could take days for a single experiment — depending on the degree of automation (off-line or
stopped flow). In the last decade of the twentieth century, the first fully automated 2D LC experiments were reported that
could further reduce the time consumption to less than one day. This led to an increased use of 2D techniques in R&D laboratories.
More information on other aspects of 2D chromatography may be taken from a review about 2D chromatography of synthetic polymers.2 Multidimensional separations are also performed in other areas, such as in a current review that focuses on biomedical and
pharmaceutical applications.3
Experimental
The implementation of 2D chromatography is straightforward if robust separation methods are available for each dimension.
Two existing chromatography systems (consisting of a minimum of two pumps, two columns, one injector and one detector) can
be used for the separation. A transfer valve for automated fractionation and appropriate software are also required.
Comprehensive 2D work employs complete transfer of injected mass from the first to the second dimension. This ensures that
the complete sample is analysed and detected in the second separation method. The advantages and limitations of different
approaches will be discussed.2
 Figure 2
|
Full automation is a basic requirement for short analysis times. However, in the last couple of years it has become possible
to reduce the total analysis time for a 2D experiment, including data evaluation, to one or two hours and in some situations
to even shorter times. The following considerations are based on the most commonly used set-up for 2D chromatography in polymer
analysis with a dual loop transfer injection valve (8-port, 2-position valve), an HPLC separation in the first dimension and
a SEC/GPC (size-exclusion chromatography) separation in the second.4 A schematic of this set-up is shown in Figure 2.
The effluent of the first dimension (HPLC column) is collected in a loop of the transfer valve and then injected automatically
into the second dimension system (GPC column). The two loops are filled and emptied into the second dimension alternately.
The interval between two valve switchings is determined by the time needed for one second-dimension analysis. To keep the
cycle times as short as possible, the second dimension flow-rate must be high. The first-dimension flow-rate is usually very
low and can easily be calculated, dividing the loop volume through the above mentioned time interval (for 100% sample transfer).
Using this procedure, the sample eluting from the first dimension will consecutively be fractionated and analysed completely
in the second dimension.
There are three important technical improvements that can be used to "speed up" 2D chromatography: column technology, detection
options and chromatography data systems.
Column technology: It is important to keep the volume that is going to be fractionated as low as possible, so this should be considered during
the development of the first dimension separation method. Recent introductions of short HPLC columns with particles smaller
than 2 μm might help to transfer existing methods with long columns and bigger particles to those that will reduce the volume
consumption but retain the resolution.
The approach to speed up GPC separations is a different one. The size-exclusion separation is based on the accessible pore
volume and if the porosity of the stationary phase is increased above a critical value, the matrix will no longer be stable.
Therefore, the column volume should not be changed if a comparable resolution is to be achieved. In 2D chromatography, SEC
is usually used as the second dimension technique that must be very fast to have a short interval between the transfer of
two fractions. Normal analytical SEC columns (300 mm × 8 mm) are used with an eluent flow-rate of 1 mL/min, which leads to
an analysis time of about 12 min. This requires a very low flow-rate in the first dimension (and, therefore, a very long analysis
time for the complete experiment), since the storage loop of the transfer valve must not be overfilled before the next transfer
to the second dimension takes place. This interval could be shortened by increasing the flow-rate of the SEC, but normal analytical
columns have an optimum flow-rate of about 1 to 1.5 mL/min. Newer HighSpeed SEC columns can allow the flow-rate to be increased
up to 10 mL/min, resulting in associated decreases in analysis time, but without a loss in resolution. This could be achieved
by changing column dimensions to 50 mm × 20 mm and further optimization of the stationary phase. As the total column volume
is similar, the quality of the separation is maintained.5
Detection options: Even if columns allow high-speed analysis, the detector must be able to detect the sample under the given conditions. For
several reasons evaporative light-scattering detectors (ELSD) are often used for 2D applications. During detection, the eluent
will be evaporated, which is a challenge at high flow-rates. This was often a limitation for high-speed measurements because
the ELSDs were not powerful enough to evaporate many solvents with flow-rates higher than 3–5 mL/min. Most manufacturers of
such detectors have optimized their instruments in recent years allowing the HighSpeed columns to be used at their optimum
flow-rates.