Detergent Removal Using Amberchrom CG161 - - Chromatography Online
Detergent Removal Using Amberchrom CG161


The Application Notebook

Introduction


Figure 1: Structure of Triton X-100.
Detergents are commonly used in hydrophobic cell membrane protein purification and viral removal/inactivation. These detergents must eventually be removed during the purification process. If a detergent has low critical micelle concentration (CMC), it can be difficult to remove from solution using techniques such as dialysis or gel filtration. Therefore, a detergent with a low CMC like Triton X-100 requires different removal techniques. Triton X-100 is a non-ionic detergent with an approximate molecular weight 625 and CMC value of 0.24 mM. The structure of Triton X-100 is shown in Figure 1.


Figure 2: Dynamic binding capacity of Amberchrom CG161 for cephalosporin C and BSA.
Amberchrom CG161 was chosen as a chromatographic solution for the removal of Triton X-100. It is a polystyrene divinylbenzene polymer used for the purification of peptides and antibiotics. This resin has been used in commercial purification processes for several decades and is an excellent choice for detergent removal because of its unique pore structure and hydrophobic nature. As shown in Figure 2, Amberchrom CG161 has a high capacity for small molecules such as cephalosporin C (479 MW), and virtually no capacity for larger proteins such as BSA (67000 Da). Therefore, detergents should effectively be removed from protein feedstocks using Amberchrom CG161.

Results and Conclusions


Table 1: Capacity comparison for Triton X-100.
For detergent capacity experiments, Amberchrom CG161 was compared with another commercially available resin which is sold specifically for detergent removal, SDR HyperD (Pall Life Sciences, East Hills, New York, USA). Amberchrom CG161M and SDR HyperD were packed into separate glass 1.0 cm ID × 7.0 cm L columns and the columns were used to determine the dynamic binding capacities for Triton X-100. A phosphate buffered solution containing 5 mg/mL of Triton X-100 was passed through the column at a flow-rate of 2 mL/min (152 cm/hr) and the effluent was collected. After complete saturation of the column (200 mL of effluent collected), the experiment was terminated. The 1% breakthrough capacity and total capacity were then calculated.

Conditions:
Column: 1cm ID × 7cm L
Flow-rate: 2 mL/minute
Sample: 5 mg/mL Triton X-100
Mobile phase: Sodium phosphate buffer, pH 7.4
Detection: UV @ 280nm

As can be seen in Table 1, Amberchrom CG161M demonstrated higher 1% and total capacity values for Triton X-100 as compared to SDR HyperD. Amberchrom CG161 chromatographic resin is an effective tool for the removal of detergents from biological samples. It provides high capacity for non-ionic detergents such as Triton X-100.

ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All Comments submitted are subject to review, and may be delayed before posting. Comments will be posted at the discretion of the editor.
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
Do you use countercurrent chromatography?
Yes, frequently
Yes, sometimes
No, never
Yes, frequently
66%
Yes, sometimes
5%
No, never
29%
View Results
Source: The Application Notebook,
Click here