Concentrating Phenolic Compounds Using the XcelVap Concentration System - - Chromatography Online
Concentrating Phenolic Compounds Using the XcelVap Concentration System


The Application Notebook
pp. 36

Natural and synthetic phenolic compounds have been used in a wide range of applications, from antiseptics and fungicides to food additives. With the adoption of the Clean Water Act (CWA), many phenolic compounds were placed on a list of "Priority Pollutants." This list encompasses highly manufactured compounds which were found in water with a frequency of 2.5% or more (1). In addition, the entire class of compounds or specific compounds may be written into National Pollution Discharge Elimination (NPDES) permits to control release into public waterways (2).


Table I: XcelVap conditions
Although the general analysis process contains extraction, extract drying, and evaporation steps prior to chromatography, this note will describe a procedure for the evaporation and concentration of organic solvent extracts containing phenolic compounds. It will make use of an automated pressure ramp during the concentration procedure to ensure that a maximum throughput of samples is achieved.

Experimental

The XcelVap benchtop, heated water bath, nitrogen blow-down system (Horizon Technology) was used with 200-mL nipple tipped concentrator tubes. The following procedure was followed:

1. Set the XcelVap conditions using the parameters given in Table I and allow the bath to equilibrate.

2. Place 200 mL of dichloromethane (DCM) into each of the six 200-mL XcelVap concentrator tubes.

3. Using a spiking mix of phenolic compounds and surrogates prepared at 100 g/mL, add 0.5 mL of the mix to each of the six concentrator tubes.

4. Start the XcelVap.

5. When complete, remove the rack from the XcelVap unit to prevent excess heat transfer to the extracts.

6. Using a pipette, transfer the concentrated extract to an auto-sampler vial, being sure to rinse the concentrator tube with DCM and bring the final volume to 1 mL.

Results


Table II: Average recovery and precision of 12 replicate evaporations
Table II shows the results of 12 replicate extracts concentrated using the procedure as given above. Each was concentrated using a single, automated pressure ramp and resulted in 0.5–1 mL of solvent. The extract was then brought up to a final volume of 1 mL prior to being injected onto the GC–MS.

Conclusions

The XcelVap Concentration System yields excellent recoveries for phenolic compounds when used in a fully automated capacity. With its large volume concentration tubes, the XcelVap System can be used with sample extracts from a wide range of extraction procedures including solid phase extraction (Method 3535), liquid-liquid extraction (Method 3510), continuous liquid-liquid extraction (Method 3520), TCLP extracts (Method 1311), pressurized fluid extraction (Method 3545), and ultrasonic extraction (Method 3550).

References

(1) http://water.epa.gov/scitech/methods/cwa/pollutants-background.cfm/.

(2) http://www.ecfr.gov/cgi-bin/retrieveECFR?gp=&SID=37bb5003c18921554522c0c1e206c8e3&r=PART&n=40y30.0.1.1.19/.




Horizon Technology, Inc.
16 Northwestern Drive, Salem, NH 03079
tel. (603) 893-3663, fax (603) 893-4994
Website: http://www.horizontechinc.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here