HILIC with Increased Sensitivity for the Analysis of Sugars - - Chromatography Online
HILIC with Increased Sensitivity for the Analysis of Sugars


The Application Notebook
pp. 23

Hydrophilic interaction chromatography (HILIC) provides an alternative approach for the separation of highly polar substances. Using a high polarity packing material, separation is based on the hydrophilic interactions of the analyte on the stationary phase surface. Shodex Asahipak NH2P series is packed with a durable polymer based packing material modified with chemically stable amino functional groups. Shodex Asahipak NH2P-40 3E column is suitable for saccharide analysis, has demonstrated the separation of sugars at low concentration, and has achieved 2–3 times improved detection sensitivity even with RI detection.

Experimental Conditions

The separation was carried out by Shodex Asahipak NH2P-40 3E (3.0 mm i.d. 250 mm, 4 m), a polymer-based amino HILIC column. The separation is compared to the same analysis using Shodex Asahipak NH2P-50 4E (4.6 mm i.d. 250 mm, 5 m). Column temperature was set at 30 C and flow rate was 0.35 mL/min for the NH2P-40 3E analysis and 1.0 mL/min for the NH2P-50 4E analysis. Eluent conditions are 75% acetonitrile in water. Injection volume of 10 L of 0.2% each sugar used for each experiment. HPLC system was coupled with RI detector.

Results


Figure 1: Increased sensitivity using NH2P-40 3E for the analysis of sugars. Column: Shodex NH2P-40 3E (left) and NH2P-50 4E (right); column temperature: 30 C; injection volume: 10 L; eluent: CH3CN/H2O = 75/25; flow rate: 0.35 mL/min and 1.0 mL/min, respectively; detector: Shodex RI.
The saccharides, arabinose, mannose, glucose, sucrose, and lactose, were analyzed successfully by HPLC with RI detection with NH2P-40 3E and NH2P-50 4E (Figure 1). Shodex NH2P-40 3E column has smaller dimensions of 3.0 mm i.d. 250 mm length with a smaller particle size of 4 m. The decrease in column dimensions requires a decrease flow rate of 0.35 mL/min. The decrease in particle size allows for an increase in the number of theoretical plates allowing for increased sensitivity of analytes. Of note, the NH2P-40 3E analysis uses one third the amount of eluent and provides up to three times improved detection sensitivity.

Conclusions

Shodex Asahipak NH2P-40 3E, a polymer-based hydrophilic interaction (HILIC) chromatography column suitable for saccharide analysis, has demonstrated the separation of these sugars at low concentration and achieved 2–3 times improved detection sensitivity even with RI detection. The presented method using NH2P-40 3E provides an efficient analysis of saccharides that can be detected in small quantities while saving the amount of eluent used per analysis.




Shodex™/Showa Denko America, Inc.
420 Lexington Avenue Suite 2335A, New York, NY, 10170
tel. (212) 370-0033 x109 , fax (212) 370-4566
Website: http://www.shodex.net/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here