High Resolution LC–MS Analysis of Reduced IgG1 Monoclonal Antibody Fragments Using HALO Protein C4 - - Chromatography Online
High Resolution LC–MS Analysis of Reduced IgG1 Monoclonal Antibody Fragments Using HALO Protein C4


The Application Notebook
pp. 10

A new fused-core particle designed specifically for the separation of proteins and monoclonal antibodies with 3.4 m particle size and a thin 0.2 m porous shell demonstrates high resolution of the heavy and light chain fragments of IgG1.

The focus of many pharmaceutical companies has shifted to larger biotherapeutic molecules as potential treatments for disease. These biomolecules present a new set of challenges compared to their small molecule predecessors when complete characterization is considered. Variants can be found through different glycans, deamidations, enzymatic clipped polypeptides, and so on, not to mention the potential for aggregate formation. Rapid reversed-phase LC analysis is convenient for its ability to be interfaced to online mass spectrometry. The HALO Protein C4 bonded phase is ideal for this application with its optimized design for high recovery and stability at the high temperatures and low pH required by the analysis.

Experimental Conditions

Column: 100 mm 2.1 mm Halo Protein C4; gradient: 29–32% B in 20 min; mobile phase A: 0.5% (v/v) formic acid with 20 mM ammonium formate; mobile phase B: 45% acetonitrile, 45% isopropanol/0.5% (v/v) formic acid with 20 mM ammonium formate; temperature: 80 C; flow rate: 0.4 mL/min; instrument: Shimadzu LCMS 2020; detection: 280 nm; MS-2020 single quadrupole MS instrument using ESI at +4.5 kV capillary voltage, 2 pps scan rate from 500 to 2000 amu m/z. Component masses for these measurements were determined by deconvolution using MagTran software, v. 1.02, based on the ZScore algorithm developed by Zhang and Marshall (1).

Results


Figure 1: LC–MS separation of reduced IgG1. Column: 2.1 mm i.d 100 mm HALO Protein C4; Flow rate: 0.4 mL/min. Mobile phase A: 0.5 % formic acid with 20 mM ammonium formate; mobile phase B: 45% AcN/45% IPA/ 0.5 % formic acid with 20 mM ammonium formate; gradient: 29–32% B in 20 min.; temperature: 80 C; detection: 280 nm; MS conditions: Shimadzu LCMS-2020, ESI +4.5 kV, 2 pps, 500–2000 m/z.
The addition of a small amount of ammonium formate to formic acid in the mobile phase improves the peak shapes. An optimized, shallow gradient allows resolution of the variants of both the heavy and light chain fragments of IgG1 as shown in Figure 1. The MS spectra can then be deconvoluted to determine the masses associated with each peak.

Conclusions

HALO Protein C4 demonstrates the low pH and high temperature stability that is required to analyze reduced and alkylated IgG1 using mass-spec friendly mobile phase. The use of 80 C enables improved peak shape while the high resolution allows complete analysis of the IgG1 fragments that are present.

References

(1) Z. Zhang and A.G. Marshall, J. Am. Soc. Mass Spectrom. 9, 225 (1998).




MAC-MOD Analytical, Inc.
103 Commons Court, Chadds Ford, PA 19317
tel. (800) 441-7508; fax (610) 358-5993
Website: http://www.mac-mod.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here