Measuring Antibody Molecular Weight by SEC-MALS - - Chromatography Online
Measuring Antibody Molecular Weight by SEC-MALS


The Application Notebook
pp. 29

A purified polyclonal antibody (IgG) is separated and fully characterized using the Viscotek SEC-MALS 20, allowing calculation of molecular weight and radius of gyration (Rg).

Therapeutic recombinant antibodies represent a growing proportion of biopharmaceuticals and are primarily classed as Immunoglobulin G (IgG). However, proteins have a tendency to aggregate over time and one challenge for biologic drugs is that the presence of aggregates will stimulate an immune response. Size-exclusion chromatography (SEC) is a powerful tool that is commonly used to look at the aggregation of proteins. While most SEC systems use a single concentration detector such as ultraviolet (UV), the addition of light scattering allows the molecular weight of the protein to be measured independent of its retention volume. The new SEC-MALS 20 detector, which uses multi-angle light scattering (MALS), is ideal for this application. In addition, the MALS detector makes it possible to measure the radius of gyration (Rg) of molecules that scatter light anisotropically.


Table I: Measured molecular weights of the different peaks of the IgG sample
In this application note, a purified polyclonal antibody (IgG) is separated using SEC and characterized using the Viscotek SECMALS 20.

Experimental Conditions


Figure 1: Overlay of MALS detector responses for IgG.
Samples were analyzed using a Viscotek TDAmax system connected to Viscotek SEC-MALS 20. The mobile phase was phosphate buffered saline, which was also used to prepare the IgG for analysis.

Results

The SEC-MALS results are presented in Table I. The monomer (15.80 mL) and dimer (14.00 mL) peaks are clearly identified by the measured molecular weights and low polydispersity (Mw/Mn). No size (Rg) can be measured for these peaks as they are below the isotropic scattering threshold of 10–15 nm. Studying Figure 1, it is just possible to see that the SEC-MALS 20 show the same response for the monomer peaks at all angles. The aggregate peak (13.23 mL) is clearly different. The molecular weight is higher and more polydisperse, which shows that there is a variable composition of molecules within the aggregate peak. Because it is large, the light scattering response varies with angle and can be used to measure the Rg.




Malvern Instruments Ltd.
Enigma Business Park, Grovewood Road, Malvern, UK
tel. +44 (0) 1684 892456
E-mail:

Website: http://www.malvern.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here