Separation of Nucleobases Using TSKgel SuperSW mAb HTP Column in HILIC Mode - - Chromatography Online
Separation of Nucleobases Using TSKgel SuperSW mAb HTP Column in HILIC Mode


The Application Notebook
pp. 12-13

Hydrophilic interaction liquid chromatography (HILIC) is one of the fastest growing modes of separation, in which any polar chromatographic surface can be used. Chemically bonded diol coated phases, as found in TSKgel SW size exclusion chromatography (SEC) columns, demonstrate high polarity and hydrogen bonding properties and do not contain ionizable groups other than the unreacted residual silanols, making them appropriate for HILIC mode.

For many years, SEC columns have been used to separate various nucleic acid species such as DNA, RNA, and tRNA as well as their constituent bases, adenine, guanine, thymine, cytosine, and uracil. In medicine, several primary nucleobases are the basis for the nucleoside analogues and other synthetic analogs which are used as anticancer and antiviral agents. Nucleobase modifications are the basis of oligonucleotide-based therapeutics, making their purification very important.

The TSKgel SuperSW mAb HTP column is a newly introduced SEC column designed for the high throughput separation of monoclonal antibodies from their high and low molecular mass variants. TSKgel SuperSW mAb HTP has a diol coating to minimize secondary interactions which may occur in SEC separations. This note demonstrates the benefits of using a TSKgel SuperSW mAb HTP column in HILIC mode for the superior resolution of four nucleobases, as opposed to using the column in SEC mode or using a HILIC column.

Materials and Methods



Results and Discussion


Figure 1: Separation of four nucleobases using TSKgel SuperSW mAb HTP column in HILIC mode at pH 7.4.
Figure 1 illustrates the separation of four nucleobases using the TSKgel SuperSW mAb HTP column in HILIC mode with 15 mmol/L ammonium bicarbonate, pH 7.4 as mobile phase B. It is important to note that the order of elution of the analytes does not correlate with their molecular mass (as in SEC separations), but instead is based on their relative hydrophilicity.


Figure 2: Separation of nucleobases using the TSKgel SuperSW mAb HTP column under conventional SEC conditions.
Figure 2 illustrates the separation of the four nucleobases on the TSKgel SuperSW mAb HTP column using conventional SEC conditions. As expected, due to the similarities in molecular masses between the four compounds, significant interference is observed amongst the peaks of interest, particularly the three pyrimidine derivatives, when separated on the TSKgel SuperSW mAb HTP column under SEC conditions. The late elution of adenine (relative to the other three compounds) may be attributed to possible interactions between the stationary phase and the derivatized purine compound, leading to a shift towards longer retention time.


Figure 3: Separation of nucleobases using the TSKgel Amide-80 HILIC column.
In an effort to explore the novelty of the separation of nucleobases using the TSKgel SuperSW mAb HTP column in HILIC mode, the same separation was carried out using a TSKgel Amide-80 HILIC column. The use of the TSKgel Amide-80 column yields very poor separation of the four nucleobases with virtually no retention of any of the components (Figure 3).

Conclusions

This work illustrates the novelty and utility of the TSKgel SuperSW mAb HTP column as a diol-functionalized HILIC column for the high resolution separation of nucleobases on the basis of their relative hydrophilicity, rather than differences in their relative molecular mass. As shown, markedly different separation profiles are observed with the use of the TSKgel Amide-80 HILIC column under identical chromatographic conditions. Additionally, nucleobase separation using the TSKgel SuperSW mAb HTP under conventional SEC conditions yielded poor resolution of all components, making it an ineffective mode of separation for this application. The TSKgel SuperSW mAb HTP column, while designed for SEC separation of monoclonal antibodies, is an extremely effective tool in HILIC mode that should be considered for the fast separation of nucleobases.

Tosoh Bioscience and TSKgel are registered trademarks of Tosoh Corporation.



Tosoh Bioscience LLC
3604 Horizon Drive, Suite 100, King of Prussia, PA 19406
tel. (484) 805-1219, fax (610) 272-3028
Website: http://www.tosohbioscience.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here