Solid Phase Extraction of Organophosphorus Pesticides and Triazine Herbicides in Water Using a New Polymeric Sorbent - - Chromatography Online
Solid Phase Extraction of Organophosphorus Pesticides and Triazine Herbicides in Water Using a New Polymeric Sorbent


The Application Notebook
pp. 18

This application describes a solid phase extraction (SPE) method for EPA method 8141B analytes, including organophosphorus compounds, and triazine herbicides in water. Target analytes are extracted onto Styre Screen HL DVB polymeric sorbent from water samples at neutral pH. Proper sample pH is critical for good analyte recoveries as organophosphorus esters hydrolyze under acidic or basic conditions. The retained analytes are eluted with acetone and dichloromethane (DCM), the eluate is then dried, concentrated, and exchanged into n-hexane or 1:1 acetone:n-hexane if a loss of polar analytes occurs.

SPE Procedure


Table I: SPE materials
a) Adjust 1 L (or less) water sample to pH 7.
b) Connect sample transfer tubes to the top of the SPE cartridges (ECHLD156) and attach the cartridges to a SPE manifold.
c) Wash the SPE cartridges with 3 5 mL DCM, condition with 2 5 mL methanol, and equilibrate with 2 5 mL DI water.
d) Insert the stainless steel ends of the transfer tubes into the sample containers, and draw the entire sample through the SPE cartridge at a fast drop-wise fashion (about 15 mL/min).
e) Remove the transfer tubes from the SPE cartridges, and dry the cartridges under full vacuum for 10 min.
f) Insert collection vials into the manifold for eluate collection.
g) Rinse the sample bottles with 5 mL acetone and, using the transfer tubes, slowly draw the rinsates through the SPE cartridges. Repeat with 10 mL DCM.
h) Remove the transfer tubes from the SPE cartridges. Continue the elution with 5 mL DCM.
i) Dry the eluates with 10–15 g anhydrous sodium sulfate (ECSS25K) held in 15 mL reservoirs (RFV1F15P) or glass funnels with glass wool.
j) Rinse the collection vials with 2 5 mL DCM, pass the rinses through the sodium sulfate and collect.
k) Concentrate the dried eluates to about 0.5 mL under a gentle stream of nitrogen at 40 C.
l) Rinse the surface of the eluate containers with 3 mL n-hexane; continue concentrating to 2 mL.
m) Adjust the final volume to 2 mL with n-hexane. Add internalstandards. The samples are ready for GC analysis.

Instrumental

GC–MS: Agilent 6890N GC with 5975C MSD Injector: 250 C Injection Vol.: 2 L
GC column: Restek Rxi -5sil MS 30 m 0.25 mm, 0.25 m, integrated with 10-m guard column
Carrier gas: Helium at a constant flow of 1.2 mL/min
Oven: Initial temperature at 60 C, hold for 1 min; ramp at 10 C/min to 300 C, hold for 2
Tune: dftpp.u
Full Scan: 45–450 amu

Results


Table II: Accuracy and precision data

Conclusion

Excellent recoveries ranging from 85.5% to 120% and relative standard deviations of less than 7.3% were obtained using UCT's new Styre Screen HL DVB polymeric sorbent (ECHLD156) for the extraction of EPA 8141B analytes in water.




UCT, LLC
2731 Bartram Road, Bristol, PA 19007
tel. (800) 385-3153; Email:

Website: http://www.unitedchem.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here