Utilization of CESI Technology for Comprehensive Characterization of Biologics - - Chromatography Online
Utilization of CESI Technology for Comprehensive Characterization of Biologics


The Application Notebook
pp. 8-9

Monoclonal antibodies (mAbs) form a major class of biologics and recently biosimilars and biobetters are being added to the growing inventory of therapeutics. In-depth characterization of mAbs at various stages of development and manufacturing is essential to maintain product safety and efficacy. However, analysis of mAbs is challenging due to their high molecular weight, the microheterogeneity presented by the glycans, and degradative modifications that occur during production. Any analytical technique that provides greater depth of information without a time penalty is an advantage. A recent advancement to meet this need was the introduction of CESI–MS. CESI is the integration of capillary electrophoresis (CE) and electrospray ionization (ESI) in a dynamic process, within the same device. In this technology, the analytes are separated inside an open nontapered capillary based on their electrophoretic mobility, and electrosprayed directly into the MS (2). At operating flow rates less than 30 nL/min, very efficient desolvation and, thus, ionization is achieved.

Though high speed CESI separations reduce analysis time, it also necessitates the use of high speed MS to preserve the separation efficiency. The TripleTOF® 5600+ system (AB Sciex) offers the necessary high acquisition speed, high resolution, and high mass accuracy, in both MS and MS-MS modes. CESI performance was evaluated by analyzing a tryptic digest of trastuzumab using the SCIEX CESI 8000 - TripleTOF® 5600+ MS platform.

Experimental Conditions

Trastuzumab was reduced, alkylated, and digested with trypsin. After drying, it was resuspended in the leading electrolyte (100 mM ammonium acetate at pH 4) and 50 nL (100 fmol) was injected into the separation capillary. The background electrolyte used was 10% acetic acid and a separation voltage of 20 kV (normal polarity) was applied for 60 min. Information dependent acquisition (IDA) was utilized to trigger MS-MS. IDA parameters were optimized so that the duty cycle of the MS was matched to the high speed CE separation. Data analysis was performed using BioPharmaView™ software (AB Sciex, Massachusetts).

Results and Discussion


Figure 1: (a) Extracted ion electropherogram of N-terminal peptide with Glu and pyroGlu separated by CESI and (b) MS-MS identification of N-terminal peptide with pyroGlu.
Primary Sequence Coverage: 100% primary sequence coverage of both the light and heavy chains of the antibody were obtained. Peptides ranging from 4 to 63 amino acids in length without any missed cleavages were detected. Electrophoretic separation is based on the charge-to-mass ratio of the peptides and is not dependent on relative hydrophobicity. Thus, small hydrophilic peptides often lost in the LC void volume, and large hydrophobic peptides, which tend to be retained on the column, can be identified by CESI–MS, resulting in the high sequence coverage observed.


Figure 2: MS-MS identification of asparagine deamidation with (a) showing unmodified peptide and (b) deamidation at Asn55 in the heavy chain of trastuzumab.
PTM Characterization: Data from CESI-MS analysis showed the presence of several PTM hotspots such as N-terminal pyroGlu formation, methionine oxidation, and asparagine deamidation. Pyroglutamination leads to loss of a positive charge which results in the electrophoretic mobility of the modified peptide being lower than the unmodified one. This is advantageous since the modified and unmodified forms can be separated by CESI–MS and the MS-MS spectra confirmed the presence of the pyroGlutamate residue (Figure 1). Oxidative degradations at Met255 and Met431 and deamidations at Asn55 and Asn387 in the heavy chain and at Asn30 in the light chain were also identified. A typical identification from the CESI–MS data is shown in Figure 2.


Figure 3: Extracted ion electropherograms of peptide TKPREEQYNSTYR with G0F, G1F, and G2F modifications.
Glycosylation Heterogeneity: Trastuzumab possesses one N-glycosylation site at Asn300 in the HC where different glycoforms such as a-fucosylated or fucosylated glycans can be present (3). By using CESI-MS, the G0F, G1F, and G2F forms of the peptide TKPREEQYNSTYR were separated well as shown in Figure 3. Furthermore, the identification of G0F, G1F, and G2F forms of peptide EEQYNSTYR without the missed cleavage (at the arginine residue in the peptide TKPREEQYNSTYR) also confirmed the presence of these glycoforms. In addition, the a-fucosylated forms of this peptide, such as G0 and G1, were identified, but it has to be further confirmed that the a-fucosylated forms were not generated due to source fragmentation of fucosylated counterparts.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here