Veterinary Drug Residue Analysis Using the AutoMate-Q40: An Automated Solution to QuEChERS - - Chromatography Online
Veterinary Drug Residue Analysis Using the AutoMate-Q40: An Automated Solution to QuEChERS


The Application Notebook
pp. 24

QuEChERS is a quick-easy-cheap-effective-rugged-safe extraction method developed in 2003 for the extraction of pesticide residues in agricultural commodities (1,2). Modifications to the method have expanded the scope to include many additional matrices and target analytes such as veterinary drug residues.

With the amounts of samples being required for residue analysis continually increasing, Teledyne Tekmar has developed the AutoMate-Q40. This revolutionary system is designed to automate the QuEChERS extraction workflow, allowing laboratories to be more efficient and timely in meeting their customer requirements for fast and reliable results.

The intent of this application is to evaluate the performance of the AutoMate-Q40 by monitoring veterinary drug residues in bovine liver. The target compounds (sulfonamides) will be analyzed using LC–MS-MS.

Extraction and Clean-Up


Figure 1: QuEChERS extraction for bovine liver.
Figure 1 shows the flow chart for the QuEChERS extraction procedure for bovine liver using the AutoMate-Q40.

Results and Discussion

A precision and accuracy study was performed on the bovine liver samples using the AutoMate-Q40. A 2.0 g/mL stock solution was used to fortify the liver samples. The AutoMate-Q40 spiked the samples with 50.0, 100.0, and 250.0 L of the standard to produce 50.0, 100.0, and 250.0 ng/g check samples. These QC samples were quantitated against a corresponding matrix matched calibration.


Table I: Data table for veterinary drug residues
Table I shows that when using the AutoMate-Q40 to extract veterinary drug residues from bovine liver samples, it exhibits recoveries ranging from 74.2% to 103.6%. By using the AutoMate-Q40 for the QuEChERS extraction, it showed great precision ranging from 1.2% to 9.4% RSD for the spiked QC samples.

Conclusion

This study demonstrates the feasibility of automating the QuEChERS extraction method using the AutoMate-Q40. The extraction process is faster, more reliable, and easier. This enables time and labor savings, while improving consistency and repeatability of the extraction. As shown above in Table I combined veterinary residues spikes recovered at 89.5% with an average RSD of 4.5%. These numbers indicate superb precision and accuracy thus validating the performance of the AutoMate-Q40 and its use as an excellent analytical tool.

References

(1) European Committee for Standardization/Technical Committee CEN/TC275, "Foods of plant origin: Determination of pesticide residues using GC–MS and/or LC–MS-MS following acetonitrile extraction/partitioning and cleanup by dispersive SPE QuEChERS-method" (2008).

(2) AOAC Official Method 2007.07 Pesticide Residues in Food by Acetonitrile Extraction and Partitioning with Magnesium Sulfate. Gas Chromatography–Mass Spectrometry and Liquid Chromatography–Tandem Mass Spectrometry, First Action (2007).




Teledyne Tekmar
4736 Socialville Foster Rd., Mason, OH 45040
tel. (513) 229-7000
Website: http://www.teledynetekmar.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here