Trace Analysis of Greenhouse Gases Using a Barrier Ionization Discharge Detector - - Chromatography Online
Trace Analysis of Greenhouse Gases Using a Barrier Ionization Discharge Detector


The Application Notebook
pp. 701-702

According to the recent IPCC report (1), climate change as a result of global warming is estimated to proceed at a faster rate than previously assumed. Consequently, easy and efficient monitoring of the greenhouse gases methane (CH4), carbon dioxide (CO2), and nitrous oxide (N2O) is becoming an increasingly important issue. The analytical challenges that surround this approach are the varying concentration ranges of the components in ambient air, ranging from high ppm levels for carbon dioxide, low ppm to ppb levels for methane, and trace levels below 100 ppb for nitrous oxide. So far, a rather complex GC system including two or three detectors was needed to fulfill all requirements. With the new Barrier Ionization Discharge Detector (BID), this approach can be simplified to a gas chromatography (GC) system with one detector and one column only.

Experimental

The BID detector belongs to the family of helium ionization detectors. It is a universal detector able to measure all types of organic and inorganic compounds except for helium and neon. Because of the novel plasma technology, the BID detector combines the high detection sensitivity of the helium ionization technique with robustness and long-term stability.


Figure 1: Ambient air chromatogram sampled at the southern edge of the city of Duisburg (Germany). Results are shown in Table 1.
Gas injection was carried out using a two-position six-port VALCO valve with a 1 mL sample loop. Chromatographic separation was performed on a 2 m Restek ShinCarbon ST micropacked column with a 1 mm inner diameter and 1/16" outer diameter. The column was selected because of its sample capacity, and therefore its ability to obtain sufficient separation of methane from the ambient air matrix peak, even with a 1 mL injection volume. Figure 1 shows the chromatogram of an ambient air sample taken in the vicinity of a larger city. This is why carbon monoxide is also present in the chromatogram.


Table 1: Quantitation results from the chromatogram in Figure 1.
The separation of matrix air requires a temperature programme starting at 30 C, then hold for 5 min, and a temperature increase at a rate of 10 C/min to 120 C. Helium with a purity of 6.0 or better was used as carrier gas and discharge gas for the BID detector. The carrier gas flow over the ShinCarbon ST column was 14 mL/min. The BID was operated at 250 C with a discharge gas flow of 80 mL/min.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors:

LCGC Columnist Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments in existing technology lines.

LATEST: The Role of Selectivity in Extractions: A Case Study

History of Chromatography | Industry Veterans:

With each installment of this column, a different industry veteran covers an aspect of the evolution and continued development of the science of chromatography, from its birth to its eventual growth into the high-powered industry we see today.

LATEST: Georges Guiochon: Separation Science Innovator

MS — The Practical Art| Kate Yu:
Kate Yu is the editor of 'MS-The Practical Art' bringing her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers.

LATEST: Mass Spectrometry for Natural Products Research: Challenges, Pitfalls, and Opportunities


LC Troubleshooting | John Dolan:

LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan, Vice President of LC Resources and world renowned expert on HPLC, is able to highlight common problems and provide remedies for them.

LATEST: LC Method Scaling, Part I: Isocratic Separations

More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here