Profiling of Essential Oil VOCs and SVOCs by GC–TMS - - Chromatography Online
Profiling of Essential Oil VOCs and SVOCs by GC–TMS

Christopher R. Bowerbank 1 , Tai V. Truong 2 , Michael Vincent 3 , Ann M. Hoffman 1 , Milton L. Lee 2 , and Douglas W. Later 1 , 1 Torion Technologies Inc., 2 Brigham Young University, and 3 Rocky Mountain Oils


Figure 1
Acceptance of raw materials in food and fragrance industries historically is performed by sensory evaluation (taste, smell, appearance) without definitive analytical data to support quality control. Portable analytical instruments for on-site characterizationby gas chromatography/mass spectrometry have only recently become available. Product-specific volatile and semivolatile organic compounds (VOCs, SVOCs) can be used to verify the identification, source, or quality of a commodity. For rapid screening, a product profile containing marker compounds can be created to match against a new lot of material prior to acceptance. This pre-acceptance approval process has numerous benefits including decreased rejection rates due to counterfeit/bait-and-switch products. Coiled wire filament (CWF) sampling combined with portable gas chromatography-toroidal ion trap mass spectrometry (GC-TMS) is an effective tool for on-site analysis of VOCs and SVOCs at the point of origin or upon receipt of a commodity.

Experimental Conditions

Frankincense, sandalwood, and corn mint were provided by Rocky Mountain Oils (Springville, UT). Essential oil extracts were prepared by dissolution in dichloromethane (~1% v/v). Sampling was performed by placing the CWF into the sample solution, with ~1 μL uptake by capillary action. The solvent was evaporated in air for 5 s, and the CWF syringe was inserted into the GUARDION®-7 GC-TMS injection port, desorbing the target analytes into a low thermal mass injector (270 °C) coupled with a capillary GC column (MXT-5, 5 m × 0.1 mm, 0.4 μm df [Restek, Bellefonte, PA]). After an initial 30 s hold at 40 °C, the GC temperature was increased at 0.3 °C/s to 250 °C, with a final hold for 100 s. The capillary GC is coupled to a TMS detector with a mass range of 45–500 m/z.

Results

Figure 1 shows the GC-TMS chromatograms for two different sources of frankincense. Each sample is discernibly different based on the presence/absence of chemical compounds at specific retention times. Source #2 does not contain major compounds found at 6 min (p-allyl Anisole) and 10.2 min (caryophyllene oxide). Likewise, caryophellene (8.6 min) in source #2 is not found in as appreciable an amount as in source #1. This likely indicates that the samples are either from different sources, or that the original lot has undergone atypical aging or improper storage.

ConclusionsVOCs and SVOCs from essential oils can be quickly analyzed on-site using GC-TMS. Profiling of these compounds using the GUARDION-7 GC-TMS can support acceptance decisions of commodities at the source. The short GC-TMS analysis time allows the user to quickly analyze and evaluate multiple samples. VOC and SVOC marker screening can be used to verify raw material quality and/or identification.

Acknowledgements GUARDION® is a registered trademark of Torion Technologies Inc.




Torion Technologies Inc.
796 East Utah Valley Dr, Suite 200, American Fork, UT 84003
tel. (801)705-6600; fax (801)705-6649
Website: http://www.torion.com/; Email:

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here