Quadrupole mass analysers were first described and developed in 1953 by the West German physicists Wolfgang Paul and Helmut
Steinwedel while they were working at the University of Bonn. Electric fields are used to separate ions according to their
mass-to-charge ratio (m/z), the ratio of mass in daltons (Da) to the integer number of charges (z), as they pass along the central axis of parallel and equidistant poles or rods. Each rod has two voltages applied, one of
which is a fixed direct current and the second is an alternating current that cycles with a superimposed radio frequency (10
kHz is not uncommon).
The magnitude of the applied electric field can be ordered such that only ions with a specific m/z ratio can travel through the quadrupole, prior to being detected. Ions with all other m/z values are deflected onto trajectories that would cause them to collide with the quadrupole rods and discharge, or be ejected
from the mass analyser field and removed via the vacuum. The quadrupole is often referred to as an exclusive detector because
only ions with a specific m/z are stable in the quadrupole at any one time. Those ions with a stable trajectory are often referred to as having noncollisional,
resonant or stable trajectories.
A typical quadrupole mass analyser consists of four rods with a hyperbolic cross section. The quadrupole rods are typically
constructed using molybdenum alloys because of their inherent inertness and lack of activity. Very high degrees of accuracy
and
 Figure 1: Schematic diagram showing the construction and applied voltages for a typical quadrupole mass analysing device.
|
precision (in the micrometre region) in rod machining and relative positioning are required to achieve unit mass accuracy
(Figure 1). For clarity the figure shows the rods with a much smaller diameter than in reality.
It is a common misconception that the quadrupole mass analyser consists of a pair of positive and negative rods. Because of
the voltage oscillating at a radio frequency (commonly known as the "RF voltage"), each pair of rods will be successively
positive then negative and so forth. In essence, there will always be a pair of positive and negative rods; however, they
will alternate at the radio frequency. The misnomer arises as one rod pair has a negative direct current (DC) voltage offset (-U) and the other a positive DC offset voltage (+U).
An ion traveling through the quadrupole will successively be attracted and then repelled from each rod until it reaches what
is known as a "saddle" field.
At certain values of U and V (DC and RF as they are colloquially referred to), ions of a particular mass-to-charge ratio will oscillate with a trajectory
that is "within" the space between rods (often called the "tunnel radius"). When this occurs, the ion, which is also accelerated
through the mass analysing device with an applied voltage between the two quadrupole ends, reaches the detector. The relationship
between the DC and alternating current (RF) voltages and the mass-to-charge ratio of stable ions can be plotted on a Mathieu
diagram.
Quadrupole rods may be "tuned" using a compound that reproducibly fragments to give ions of particular mass-to-charge ratio.
A popular compound for tuning the quadrupole in electron ionization gas chromatography–mass spectrometry (GC–MS) is perfluorotributylamine,
which fragments very reproducibly over a wide mass range with relative fragment intensities that are known and reproducible.
The magnitude of the applied voltages to allow the passage of specific fragment ions can then be assigned and the "mass axis"
can be calibrated. Furthermore, the DC and RF voltages can be adjusted to alter the resolution and sensitivity of the device.
This operation is typically carried out using an automated instrument algorithm, although learning to tune the device manually
can result in much higher sensitivity for particular target ion masses.
Typically, a single spectral experiment involves a range of DC to RF values being scanned to sequentially allow ions across
the full mass range to pass through the analyser. This is called a "scan" function, and a scanning rate of 10 Hz is considered
very good across a mass range of 50–500 m/z. The total ion abundance of each successive scan is summed to produce a total ion current (TIC), which functions as the pseudo
chromatogram — of course the contributing spectrum for each data point can then be extracted if required.
For higher sensitivity, the scan range of the quadrupole can be limited or fixed to certain DC to RF values that correspond
to known high abundance ions produced by the analyte of interest. In this way, the frequency of measurement is increased and
consequently the signal-to-noise ratio increases; this is known as selected ion monitoring.
Typical commercial quadrupole instruments can achieve unit mass resolution; that is, mass 201 and mass 202 can be sufficiently
resolved from each other, although under ideal conditions the quadrupole is capable of much higher resolution. Commercially
available instruments typically achieve resolution values of up to 5000 (full width half maximum), with mass accuracy of higher
than 100 ppm, meaning that automated derivation of elemental composition is not possible. Single-quadrupole liquid chromatography
LC–MS and GC–MS instruments are used in laboratories all over the world and provide a cost effective, but relatively sensitive
and selective means of ion detection.