Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification or Calibration, Part I - - Chromatography Online
Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification or Calibration, Part I


LCGC North America
Volume 28, Issue 12

In this two-part article series I'll be discussing the role of system suitability tests (SSTs) in the context of analytical instrument qualification (AIQ) and the generation of quality analytical results. I'll look at this first from the perspective of regulatory requirements based upon the United States Pharmacopeia (USP) general chapter 1058 (1058) on AIQ (1). As much of this discussion is also good analytical science I will also look at this from the perspective of ISO 17025 (2), the interpretation of the ISO 9000 quality management system for testing and calibration laboratories. I'll use the ISO 17025 term of calibration to be equivalent to AIQ and discuss the reason why I have taken this approach.


Table I: Selected U.S. good manufacturing practice (GMP) regulations for instruments and equipment
Specifically, I want to discuss the question why SSTs are not a substitute for AIQ for the initial qualification or requalification of a chromatograph (that is, a periodic check or operational qualification) under GMP or calibration under ISO 17025. The reason for this is that there is a common misconception in some laboratories that SSTs can be used to qualify a chromatograph. This is wrong. Using SSTs as the sole instrument qualification approach will leave any laboratory exposed to regulatory action or nonconformance as the chromatographs cannot be demonstrated as being fit for their intended purposes. Therefore, we will explore the reasons for this from both the regulatory and quality perspectives.

What Is a System Suitability Test?

An SST is simply a check to see if the chromatographic system (instrument modules, column, and mobile phase) is capable of undertaking the analysis you require. It is performed immediately before the samples you have prepared in the laboratory are committed for analysis. At it's simplest, an SST is the single injection of a standard solution of reference compounds to see if the separation is "OK." The only criterion used for assessing if the system is acceptable is a visual check by the chromatographer to assess if the separation is within expected boundaries. However, there are no measurements made: Just a visual assessment that things are right.

At the other end of the scale are the formal SSTs described in the pharmacopeias used in pharmaceutical QC laboratories. Here a minimum of five replicate injections of a standard are injected and the calculated peak areas and other chromatographic criteria (retention time window, peak shape, peak tailing, resolution between peaks, and so forth) are objectively compared with predefined specifications defined by the pharmacopeia (peak area repeatability) and the laboratory (all other chromatographic criteria). If the SST injections meet all the predefined criteria then the chromatographic system is deemed acceptable and the samples committed for analysis.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC North America,
Click here