Dynamic Gas Chromatography Meets Dynamic Mass Spectrometry - In this column, we explore the interaction of dynamic gas chromatography with the dynamic aspects of mass spectrometry and discuss some key
Dynamic Gas Chromatography Meets Dynamic Mass Spectrometry
In this column, we explore the interaction of dynamic gas chromatography with the dynamic aspects of mass spectrometry and discuss some key assumptions that can affect scientific results if incorrect.


Spectroscopy
Volume 28, Issue 1, pp. 14-23

Consider that a reference-quality mass spectrum for a compound is measured when a few "ideal conditions" are fulfilled, or we assume that they are. For instance, we assume that the instrument measures the spectrum across the entire mass range with no mass discrimination. Ideally, ions of all masses are analyzed and detected with equal probability and the recorded ion intensities accurately reflect the extent of the various fragmentation processes. We assume that the resolving power of the instrument is sufficient to distinguish ions of adjacent masses. There are other assumptions that often "go without saying." But we are going to say them here because when the assumptions are not true, our measurement data can be skewed. We assume that the sample for which the reference mass spectrum is recorded is a pure sample, and purity may include enantiomeric composition, or for larger molecules, higher-order structures. Finally, we assume that the amount of sample in the ionization source of the mass spectrometer remains constant during the time required to record the mass spectrum. Each of these desiderata is, of course, an "ideal condition" not usually fully met. Sometimes, the assumed condition is not even slightly achieved. To achieve at least a measure of sample purity, chromatography is often used to introduce samples into the mass spectrometer. Many forms of column chromatography, however, elute samples into the mass spectrometer within a varying concentration centered around a retention time. The elution profile defines the peak shape. Here, we explore the interaction of dynamic chromatography with the dynamic aspects of mass spectrometry (MS).

Dynamic GC


Figure 1: This symmetrical shape is the idealized representation of a peak eluted from a GC column. The x-axis is time after injection of the sample onto the column, and the y-axis is a detector response proportional to the amount of sample eluted from the column.
Let's start with a leading and hopefully thought-provoking question: If a pure sample is eluted from a gas chromatography (GC) column operated in its normal fashion, and there is no detector, what is the peak shape at the end of the column? The dynamic processes of GC are well understood, and we can model the processes of molecular diffusion in the mobile and stationary phases to arrive at an answer. If the sample is a pure compound and the diffusion processes are unperturbed, then the sample will be eluted within a symmetrical peak shape as shown in Figure 1, with the x-axis representing time after sample injection at the front of the column and the y-axis representing some parameter proportional to the amount of sample eluted from the column. The time point at which the mass flow rate from the column reaches a maximum is called the retention time, and this is labeled on the figure. On the same column, using the same GC conditions, the retention time should be reproducible, and it can be used as a factor in identifying the sample. One analytical principle to keep in mind here is that of "necessary and sufficient." More precisely, if the retention time of a reference standard is known, then if the unknown sample is the same compound as the reference it will be eluted at the same retention time. However, if the unknown sample is eluted at the same retention time, it may be the same compound as the reference. To be sure of our identification we need more information, and that brings us to information provided by the detector other than just a determination of the retention time. We want a detector that provides additional and characteristic information, which brings us directly to MS. Although chromatography coupled with mass spectrometry seems like the marriage made in heaven, we should always be aware of the mutual give-and-take in the relationship. In this column, we discuss the dynamic characters of our techniques, and we will start very simply.

Detector Characteristics

The characteristics of the detector determine the nature of the y-axis parameter in Figure 1. If the detector operates to produce a response that is directly and linearly proportional to the mass flow of the sample entering it, and if it operates to produce a measureable signal in a time frame much shorter than incremental changes in mass flow, then the measured peak shape with this detector is very close to the modeled peak shape. If the detector produces a response that is concentration-dependent, then the relevant proportionality factor will change the appearance of the peak, although the peak shape may still be symmetric overall. If the detector operates in such a fashion that it provides a signal from some types of eluted sample molecules and not others (such as with an electron-capture detector), then the proportionality factor for the y-axis response changes from sample to sample and peak to peak. Most modern GC detectors produce a near-instantaneous response, so that the x-axis in Figure 1 remains directly proportional to time. We can foresee that if the detector itself has dynamic characteristics that have a scale that is similar to the dynamics of the elution of the GC peak, then a proportional factor must be applied to the x-axis as well.

The quintessential GC detector that provides the wealth of information needed to identify compounds as they are eluted from the column is the mass spectrometer. Accordingly, we need to evaluate the role of MS as a GC detection method in terms of its response characteristics that may provide a proportionality factor that changes the y-axis (response) or the x-axis (time) in Figure 1. Many types of ionization methods, mass analyzers, and detectors used in MS have been previously described in this column. The definitive matrix for MS ionization methods and their mass- or concentration-dependent response factors across all types of compounds is a construct that will probably never be completed. It cannot be predicted with specificity, and therefore the calibration curve for the particular sample of interest remains a reliable method to establish the proportionality on the y-axis. In the remainder of this column, we discuss the dynamic nature of mass analysis using various types of instruments and describe the effects of that analysis on the measured elution profile of a GC peak (the x-axis).


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
Sharing new techniques
Viewing the latest products and applications
Networking
Learning
Sharing new techniques
36%
Viewing the latest products and applications
27%
Networking
9%
Learning
27%
View Results
Source: Spectroscopy,
Click here