Significant Costs Savings Achieved Through the Development of a New Analytical Method for Residual API Analysis, with a Kinetex® 1.7 μm Core-Shell UHPLC Column - - Chromatography Online
Significant Costs Savings Achieved Through the Development of a New Analytical Method for Residual API Analysis, with a Kinetex® 1.7 μm Core-Shell UHPLC Column


The Application Notebook

This study demonstrates how the ultrahigh-performance of 1.7 μm core-shell columns can be utilized in the laboratory in order to realize significant cost savings.

The pharmaceutical industry has been undergoing extensive changes in recent years, primarily brought upon by the need to reduce operational costs. This project resulted in an annual cost saving of €250,000 to €350,000 Euros, based upon the reduction in analysts, solvents, wastes, and equipment overheads.



Flow Rate: 0.4 mL/min

Temperature: 50 °C

Detector: PDA 210-300 nm, extracted channels 225 nm and 280 nm

Instrument: Waters® ACQUITY® equipped with PDA

Sample: 1. Antidepressant drug (containing an HCl salt); 2. Hormone therapy #1 (containing a salt); 3. SERM drug (containing a basic functional group); 4. CNS drug (containing basic functional group); 5. PPI drug (containing basic functional group); 6. CNS drug (containing basic functional group); 7. CNS drug (containing basic functional group); 8. Hormone therapy #2 (neutral); 9. Oral contraceptive hormone #1 (neutral); 10. Hormone therapy #3 (neutral); 11. Oral contraceptive hormone #2 (neutral); 12. Hormone therapy #4 (neutral); 13. Oral contraceptive hormone (neutral); 14. Hormone therapy #5 (neutral); 15. Hormone therapy #6 (acetate salt of 14); 16. Immunosuppressant drug (macromolecule, containing basic functional group)

Swab Extraction Procedure

A swab containing residual API from different production surfaces was extracted with 2 mL of a solution containing a mixture of 1:1 acetonitrile:water and shaken for 5 min.

Results


Figure 1: Sample chromatogram at 225 nm (a) and 280 nm (b) using a Kinetex C18 1.7 μm, 100 × 2.1 mm.
The active pharmaceutical ingredients (APIs) used in these experiments are a mixture of organic salts, organic bases, and neutral organic compounds. The analysis was performed on a Phenomenex® Kinetex® 1.7 μm C18 column under gradient conditions with a flow rate of 0.4 mL/min. Two different UV wavelengths were used for quantification, 225 nm as shown in Figure 1a and 280 nm as shown in Figure 1b.

A reduced run time of 6 min compared to 10 to 20 min in the original method was achieved. The critical pair for separation was compounds 5 and 6, as the closest eluting peaks. However, molecules 5 and 6 are not in the same class of drug and would not routinely be manufactured on the same equipment set and hence seen in the same sample. Therefore, complete resolution was not required.

Conclusion

A new analytical method was developed which is capable of analyzing 16 different API residues from production surfaces. It has been shown that the 1.7 μm Kinetex 100 × 2.1 mm column was capable of resolving 16 different chemical entities with a 6 min run time. This new analytical method will be used to replace 16 older methods, thereby facilitating an annualized cost saving for the site of €320,000.

For complete details request Laboratory Case Study by Andrew Charles and colleagues.

Phenomenex, Inc.
411 Madrid Avenue, Torrance, CA 90501
tel. (310) 212-0555, fax (310) 328-7768
Website: http://www.phenomenex.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Avoiding Reversed-Phase Chromatography Problems Through Informed Method Development Practices: Choosing the Stationary-Phase Chemistry


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Estimating Resolution for Marginally Separated Peaks


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here