1) Not Taking Enough Care at the Inlet

Any leak in GC will cause significant problems with chromatography and the inlet is the most likely area. Don't take the risk of having to troubleshoot a leak, change the septum regularly and make sure the column is correctly installed and leak-free.

2) Turning the oven on before the carrier gas

If there's one thing that will permanently damage a column more quickly than anything else it's heating the column without any carrier gas flowing. Get into the habit of making sure the carrier gas is flowing before you switch on the oven AND make sure the oven is cool before switching the carrier off. When switching the oven off, set the oven temperature to 35oC before switching the oven off as this will cool much quicker.

3) Not programming in column dimensions

Most modern GCs use an EPC (electronic pneumatic controller) to accurately deliver carrier flows and this relies totally on the correct column dimensions being entered by the operator. Make sure anytime you change the column you enter the correct dimensions into the GC or else your flowrate and split ratios will be wildly inaccurate.

4) Not using optimal flowrate for carrier

Each carrier gas has an optimal linear velocity, which gives the best efficiency (Nitrogen c.12cm/sec, Helium c.35cm/sec and Hydrogen 40 - 50cm/sec). To obtain this optimal linear velocity the flowrate is matched to the internal diameter of the column in use. e.g. for a 0.32mm ID column using helium the optimal flowrate will be about 2.0ml/min. If the optimal flowrate is not used, column efficiency can be significantly reduced.

The complete article is available to CHROMacademy Lite and Premier members here >>

CHROMacademy Lite Membership is FREE and it only takes two minutes to register.

With a Lite Membership you are given access to:
∙ This month's webcast & tutorial
∙ Selected eLearning modules
∙ Featured CHROMacademy Content
∙ The CHROMacademy forum

Test drive CHROMacademy and Check out more great content available FREE to our Lite members »

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: ChromAcademy,
Click here