HPLC Gradient Elution – Baseline Drift - - Chromatography Online
HPLC Gradient Elution – Baseline Drift


The CHROMacademy technical team was recently posed with the question of the causes of baseline drift in gradient high performance liquid chromatography (HPLC) (Figure 1).

The primary cause of baseline drift in gradient HPLC is due to changes in the refractive index of the eluent. During gradient elution the composition of the eluent will change and, hence, so will its refractive index. This usually manifests itself as a gradual increase in response during the gradient time. To compensate for the change in refractive index properties a reference wavelength should always be set otherwise drifting baselines will occur.

Positive baseline drift is commonly seen in gradient elution and is often caused by differences in the absorbance properties of the A and B solvents.  In the case of reversed phase HPLC gradient elution the eluent composition will change from a high percent of A (aqueous) to a high percent B (organic solvent).  The common organic solvents used in reversed phase HPLC have a greater UV absorbance (Table 1) than aqueous (water) which results in the rising baseline when monitoring at low wavelengths (~200 nm).  The problem of baseline drift caused by the disparity in UV absorbance can be remedied in three ways.

The complete article is available to CHROMacademy Lite and Premier members here >>

CHROMacademy Lite Membership is FREE and it only takes two minutes to register.

With a Lite Membership you are given access to:
∙ This month's webcast & tutorial
∙ Selected eLearning modules
∙ Featured CHROMacademy Content
∙ The CHROMacademy forum

Test drive CHROMacademy and Check out more great content available FREE to our Lite members »

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>



Click here