HPLC Mobile Phases — 10 Bad Habits to Avoid



1. Measuring the pH of the mobile phase after the organic has been added: pH meters are calibrated to give the correct pH readback in aqueous solution — the buffers you verify this with are aqueous. If you measure the pH with the organic added, the pH will be different to that of measuring before organic addition. However, the most important point is to be consistent. If you do always measure pH after the organic is added, make sure you state this in the method so that everyone does it the same way. It won’t be 100% accurate, but at least it will be consistent. This is probably more important than having the exact pH.

2. Not using a buffer: Buffers are present to control pH and resist a change in pH. Many other parts of method (for example, sample matrix, CO2 in air, source of water used for your mobile phase) can change the pH of the mobile phase causing shifts in retention, peak shape, and peak response. (Formic acid, TFA and so on are not buffers).

3. Not using the buffer in its correct pH range: Each buffer salt has a two pH unit wide range over which it provides the optimal pH stability. Outside this window the salt is ineffective at resisting change in pH. Either use your buffer within the correct range or pick a buffer whose range covers the pH you require.

4. Adding buffer to organic: Mixing aqueous buffer into the organic phase carries a high risk of the buffer being precipitated — in many cases so finely that it may not be obvious it has happened. ALWAYS add the organic to the aqueous phase as this greatly reduces the risk of buffer precipitation.

5. Using the pump to mix gradients from 0%: Modern pumps are very effective at mixing mobile phases and degassing online, however not everyone has a high quality pump. Premix your A and B starting mix to a single solution that runs at 100% on line A. For example, prepare the starting mixture by mixing 950 ml Aqueous with 50 ml organic, then filter and degas. This reduces variability between HPLCs, reduces the risk of bubbles and precipitation in the system. Note: 95:5 mixed on the pump will not give the same retention time as 95:5 premixed in the bottle — you normally need to add a few more percent organic when premixing.

The complete article is available to CHROMacademy Lite and Premier members here >>

CHROMacademy Lite Membership is FREE and it only takes two minutes to register.

With a Lite Membership you are given access to:
∙ This month's webcast & tutorial
∙ Selected eLearning modules
∙ Featured CHROMacademy Content
∙ The CHROMacademy forum

Test drive CHROMacademy and Check out more great content available FREE to our Lite members »

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>



Click here