Extraction of Oil from Oilseeds Using Accelerated Solvent Extraction (ASE)


The Application Notebook

Jennifer Peterson, Richard Carlson, Bruce Richter, and David Knowles, Dionex Corporation


Figure 1: Rate of oil extraction from oilseeds.
Oils for foods and cooking are derived from oilseeds like canola, soybeans, corn, flax, and cotton. More recently, oil from oilseed crops has been used in the production of biodiesel. As a result, many agricultural companies are researching potential oilseed crops for maximum output of oil to be used for biodiesel. It is not only important to grow oilseeds that produce the best yields but also those crops that produce the best type of oil.

Traditional solvent extraction techniques use large volumes of solvent (typically several hundred milliliters) and long extraction times (8–16 h) to remove oil from the seeds. Once the oil is removed, the weight percent of the oil seeds can be determined, and the composition of the oils can be studied.

Accelerated Solvent Extraction (ASE ) uses high temperature and pressure to reduce the extraction time and solvent usage for solvent extraction. ASE has proven to be an effective technique for the extraction of oil from oilseeds.

Experimental


Table I: AOCS Method AM 2-93
Canola seeds which contain approximately 45 weight percent oil were ground and weighed into an extraction cell. The ground seeds were extracted using 100% petroleum ether and 105 C with three 10 min cycles. The extracts were evaporated to dryness, and the collection vials were weighed to determine the percent of oil.

Results and Discussion


Table II: ASE System Method
The ASE results were compared to the traditional AOCS (American Oil Chemist Society) Official Method AM 2-93 which is based on the FOSFA (Federation of Oil Seeds and Fat Association) Official Method (Table I).

The weight percent of oil in the seeds was determined to be 44.9% with 0.31% RSD (n = 3) with ASE as compared to 45.2% with 0.24% RSD (n = 12) for the AOCS method. The weight of oil extracted per unit time is contrasted for ASE and the FOSFA procedure (see Figure 1).

Conclusions

The data shows that ASE can be used effectively to extract oil from oilseeds. The data generated are equivalent to those obtained by existing methods, but time and solvent usage are significantly less with ASE (Table II).

ASE is a registered trademark of Dionex Corporation.




Dionex Corporation
1228 Titan Way, P.O. Box 3603, Sunnyvale, CA 94088
tel. (408)737-0700; fax (408)730-9403
Website: http://www.dionex.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors:

LCGC Columnist Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments in existing technology lines.

LATEST: The Role of Selectivity in Extractions: A Case Study

History of Chromatography | Industry Veterans:

With each installment of this column, a different industry veteran covers an aspect of the evolution and continued development of the science of chromatography, from its birth to its eventual growth into the high-powered industry we see today.

LATEST: Georges Guiochon: Separation Science Innovator

MS — The Practical Art| Kate Yu:
Kate Yu is the editor of 'MS-The Practical Art' bringing her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers.

LATEST: Mass Spectrometry for Natural Products Research: Challenges, Pitfalls, and Opportunities


LC Troubleshooting | John Dolan:

LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan, Vice President of LC Resources and world renowned expert on HPLC, is able to highlight common problems and provide remedies for them.

LATEST: LC Method Scaling, Part I: Isocratic Separations

More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here