GC–MS Leak Detection


Recently a CHROMacademy user contacted our experts as their GC-MS system was not maintaining vacuum.

Normally, the vacuum was set at 1.2x10-7 Torr; however, it was dropping to 2.3x10-4 Torr. 
The user carried out some maintenance by tightening and then subsequently replacing the nut and ferrule at the transition line which solved the problem initially.

Unfortunately this did not fully remedy the problem. The source was capped and the pressure held, however, this still left the user scratching their head as to why they were seeing gross pressure drops.

The vacuum not holding in a GC-MS system would indicate a leak. Leaks in a GC–MS system can be particularly frustrating; they can cause spurious peaks in the mass spectrum, a decrease in sensitivity, loss of carrier gas (this could be expensive), and damage to the instrument.

Often leaks can occur where the column enters the MS or at point(s) where the vacuum system was disassembled and reassembled (i.e. for any routine maintenance operations). If it is expected that there is a leak in a GC-MS system there are a few things that can be done to try and identify the source of the leak in order to repair it.

1. When the system is running verify that peaks at m/z 28 and 32 are present in the mass spectrum.  These peaks are indicative of nitrogen and oxygen; if they are present in a ratio of 4:1 (nitrogen:oxygen) this would indicate that there is a leak to the atmosphere.  If the peaks for water (m/z 18), nitrogen (m/z 28), and oxygen (m/z32) rise above 10% (relative to m/z 69 peak — typical base peak from PFTBA used during auto-tune) this is also indicative of a leak.

The complete article is available to CHROMacademy Lite and Premier members here >>

CHROMacademy Lite Membership is FREE and it only takes two minutes to register.

With a Lite Membership you are given access to:
∙ This month's webcast & tutorial
∙ Selected eLearning modules
∙ Featured CHROMacademy Content
∙ The CHROMacademy forum

Test drive CHROMacademy and Check out more great content available FREE to our Lite members »

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Avoiding Reversed-Phase Chromatography Problems Through Informed Method Development Practices: Choosing the Stationary-Phase Chemistry


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Estimating Resolution for Marginally Separated Peaks


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>



Click here