Antibody therapeutics are enjoying high growth rates in the biopharmaceutical market, with the major areas of therapeutic
application being cancer and immune/inflammation-related disorders including arthritis and multiple sclerosis. In 2010, four
of the top ten best-selling global drug brands were monoclonal antibodies (mAbs). The characterization of these complex biomolecules
is a major challenge in process monitoring and quality control (QC). The main product characteristics to be monitored are
aggregate and fragment content, glycosylation pattern and charged isoforms. The standard method used in biopharmaceutical
QC for mAb aggregate and fragment analysis is size exclusion chromatography (SEC). A new series of silica-based HPLC columns
can be applied to either increase speed or improve resolution of the separation of antibody fragments, monomers and dimers.
Experimental
IgG was digested with papain over 24 h. The fragmentation process was monitored by analysing 10 or 5 μl aliquots of the sample.
Mobile phase: 200 mmol/L phosphate buffer + 0.05% NaN3, pH 6.7
Flow rate: a & b: 1.0 mL/min; c: 0.35 mL/min
Injection vol.: a & b: 10 μl; c: 5 μL
Temperature: 25 °C
Detection: UV @ 280 nm
Samples: 10 g/L IgG digested with papain for 0–24 h
Columns: a: TSKgel G3000SWXL, 7.8 mm i.d. × 30 cm
b: TSKgel SuperSW mAb HR, 7.8 mm i.d. × 30 cm
c: TSKgel SuperSW mAb HTP, 4.6 mm i.d. × 15 cm
Results
 Figure 1: Separation of antibody fragments, monomers and dimers by size exclusion chromatography (SEC). a: TSKgel G3000SWXL,
7.8 mm i.d. × 30 cm; b: TSKgel SuperSW mAb HR, 7.8 mm i.d. × 30 cm; c: TSKgel SuperSW mAb HTP, 4.6 mm i.d. × 15 cm.
|
Figure 1a shows the separation of a papain digested immunoglobulin G sample on a TSKgel G3000SWXL column, which is applied
as the standard SEC column in routine analysis of aggregates in many QC and R&D labs. Figure 1b demonstrates that the resolution
of the separation can be improved by using the new TSKgel SuperSW mAb HR (High Resolution) with 4 micron silica particles.
This column provides higher resolution than the conventional column at the same analysis time. Using the TSKgel SuperSW mAb
HTP (High Throughput), a short semi-micro column packed with the same 4 micron particles as SuperSW mAb HR, dimer, monomer
and fragments were separated at the same resolving power as on the conventional column but in half the analysis time (Figure
1c).
Summary
SEC is a common method for the separation of antibody monomer from dimer, aggregates or degradation products on the basis
of molecular size. Two novel SEC columns designed for antibody separation exhibit reduced analysis time while achieving baseline
separation or enhanced resolution between monomer and dimer.