High Temperature GPC Analysis of Polyolefins with Infrared Detection - - Chromatography Online
High Temperature GPC Analysis of Polyolefins with Infrared Detection


The Application Notebook


Introduction

The analysis of the molar mass distribution of polyethylene and polypropylene resins by GPC/SEC has always been considered a demanding task because of the requirement of high temperature operation for dissolution and complex hardware design, which often results in high maintenance cost, in particular related to the autosampler/injector and detector units, and in other problematic and consuming tasks such as solvent handling added to column fragility, sample degradation or detector sensitivity–stability.

All those considerations were kept in mind in the design of a new high temperature GPC instrumentation.

Instrumentation and Experimental

The analyses are performed in a newly developed modular GPC instrument from Polymer Char, the GPC-IR, which consists of a main body, an autosampler and a pump.

The main body incorporates two ovens, preventing the columns from suffering during maintenance tasks and, therefore, extending their life; an Agilent GC oven is used for the columns, which can be independently heated up to 220 C with a temperature control of 0.01 C. On top of the columns' oven, there is a separate oven, where injector and detectors are located. This design allows preventive maintenance to be performed and the loop volume to be changed without affecting the columns' temperature.

The autosampler is a separate unit with 70 vials of 10 mL attached to the detectors/injector oven through a heated interface. The vials are filled automatically before entering for the dissolution process into a small heated carousel oven. After dissolution time (different dissolution times can be selected for each sample), the sample is automatically filtered meanwhile filling the loop before injection. Vials stay in this carousel oven just the needed time for dissolution, before injection, to minimize sample oxidation.

Additionally, optional nitrogen purging of the vial can be used to prevent further oxidation. Dissolution is facilitated with a gentle shaking of the vial to prevent shear degradation. Up to 3 injections can be performed in the same vial without any dilution among injections.

The concentration detector used is an infrared detector, which is very appropriate for polyolefin analysis as per its demonstrated high sensitivity for this kind of polymers, as well as fast baseline stabilization. Additionally, by incorporating a second wavelength sensor to the same detector, concentration and composition are measured simultaneously, as the number of branches per 1000C (optionally, a carbonyl sensor can be used for the composition distribution of EVA or EAA resin analysis).

For triple detector GPC analysis, the light-scattering detector HELEOS from Wyatt is easily incorporated on the right side, as well as a four capillary viscometer integrated in the injector/detectors oven.

The analyses have been performed with trichlorobenzene stabilized with 300 ppm of BHT at 1 mL/min as solvent. A sample solution of 18 mg in 9 mL and injection of 200 μL is used with integrated filtration and automated backflush inside the injector/detector oven. Three columns (30 cm each with 13 μm particles packing) were used at 150 C. Injector/detectors oven temperature was set at 160 C.

Results


Figure 1: (a) IR4 results of a metallocene polyethylene resin. (b) IR5 duplicated results from three samples: PP and PE homopolymers and a copolymer.
The analysis of a metallocene polyethylene resin is shown in Figure 1(a) with the infrared detector signals. The orange curve corresponds to a selective absorbance with emphasis in the CH3 region. The red curve corresponds to the concentration signal based on a broadband absorbance of CH2 and CH3, which is used for all the molar mass moments calculation. The ratio of those two signals is directly proportional to the amount of branches or CH3/1000C as shown in purple as a flat line in this homogeneous resin. For high levels of branching or for the composition analysis of propylene-ethylene copolymers the calibration can be obtained with the injection of polypropylene and polyethylene homopolymers as shown in Figure 1(b). The molar mass distribution curves are shown together with the branch content for duplicated analysis of the two homopolymers and a PE copolymer (metallocene resin) with 20 CH3/1000C.


Figure 2: Triple detector results of a LDPE resin (NBS1476).
The analysis of a LDPE resin with triple detector (c+visc+LS) is shown in Figure 2, with the molar mass distribution curves as well as the corresponding Mark-Houwink plot.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here