How Much Retention Time Variation is Normal? - - Chromatography Online
How Much Retention Time Variation is Normal?


LCGC Europe
Volume 27, Issue 8, pp. 409414

Small changes in retention time of a liquid chromatography (LC) method are normal. At what point is a problem suggested? Retention time shifts can be frustrating when you can't figure out if a shift is something you caused or if there is another reason for it.

It is interesting how problems tend to cluster — in the last few weeks I've received several questions related to retention time shifts with liquid chromatography (LC) methods. Some of these correspond to the preparation of new batches of mobile phase, some are from one day to another, some are from within a batch of samples, and some result from a change in instruments. In this month's "LC Troubleshooting," I will discuss some of the factors that influence small changes in retention.

Most LC methods run on modern instrumentation with a good quality column will have quite stable retention times. I generally expect to see run-to-run variation in the second decimal place of the retention time, such as 0.02–0.05 min. However, the historical behaviour of the method should be used to determine what is normal for that method. For example, with large biological molecules, such as proteins, that use shallow gradients, the variability can be an order of magnitude or more larger.

Mobile Phase Organic Concentration

One of the most common causes of shifts in retention time in reversed-phase LC separations is a minor change in the concentration of the organic solvent, usually methanol or acetonitrile. This can happen from a minor error in formulation or a change in the mobile-phase composition if one solvent evaporates over time.

For small molecules (arbitrarily <1000 Da), we can use the "Rule of Three" to estimate the effect of a change in organic solvent, or %B. This rule states that the retention factor, k, changes approximately threefold for a 10% change in %B. This rule derives from plots, such as that in Figure 1, where k is plotted versus %B on a semi-log plot. The red line in Figure 1 represents the retention behaviour for a 500 Da analyte. Retention changes from ~29 min at 40% B to ~4 min at 60%. For practical purposes, this relationship can be considered linear and represented in the standard y = mx + b format as



where kw is the (theoretical) retention in 100% water, S is the slope of the plot, and Φ is the %B as a decimal (0.5 = 50%). Values of S can be determined from two experimental runs using equation 1. Empirical observations indicate that S can be estimated as



where MW is the molecular weight (Da). Thus for a 500 Da analyte, S ˜≈ 5.6. If we consider S ≈˜ 5 as an average for sub-1000 Da molecules, we can then estimate how k changes with %B with



where Δk is the change in k value for a Φ change in organic. If S = 5, a 10% change in organic gives Δk ≈˜ 1050.1 = 3.16 ≈˜ 3. This is the basis of the Rule of Three.



As an example, our 500 Da analyte in Figure 1 has k = 5 at 50% B. We can convert this to retention time, t R, by rearranging the equation for k, where t0 is the column dead time (also abbreviated tM). If we assume a 150 mm 4.6 mm column operated at 1 mL/min, t0 ≈˜ 1.5 min, so tR = 1.5(1 + 5) = 9.0 min. The Rule of Three would suggest that k ˜≈ 3 5 = 15 for a change to 40% B, which would correspond to tR ≈˜ 24 min. In fact, S = 5.16 for our analyte, so k = 18.1 and tR = 28.7 min — close enough for a rule of thumb.


Table 1: Effect of small errors in mobile-phase composition on the retention for analytes of different S values.
Mobile-phase formulation errors should not be in the 10% region if you are at all careful, so what happens for smaller changes? We can use equations 3 and 5 to determine the effect of a 0.1%, 0.5%, and 1% error in formulating our 50% B mobile phase for our "average" S = 5 small molecule. I'll leave the calculations to you, but the results summarized in Table 1 show that these correspond to retention shifts of ≈˜ 0.1, 0.4, and 0.9 min, respectively (I have rounded values for display, so if you try to repeat my calculations, your results may vary somewhat). So, you can see that it takes only a minor error in mobile-phase formulation to shift retention times enough to notice the change.

With larger molecules, the problem is magnified, because S increases markedly with an increase in molecular weight. This makes the log k versus %B plots steeper, as is seen for 5000 Da (blue) and 50,000 Da (green) compounds, such as peptides or proteins, respectively. Steeper plots mean that these compounds are much more sensitive to minor changes in the mobile-phase composition. Our Rule of Three for <1000 Da samples becomes the Rule of 60 at 5000 Da, and the Rule of 400 at 50,000 Da. This extreme sensitivity of the retention of large molecules to small changes in %B means that isocratic separation is not practical for most separations of these analytes.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC Europe,
Click here