Increasing Column Efficiency with Slip Flow Technology - - Chromatography Online
Increasing Column Efficiency with Slip Flow Technology

LCGC Europe eNews

LCGC recently spoke with Dr. Mary Wirth, the W. Brooks Fortune Distinguished Professor in the Department of Chemistry at Purdue University, about her groundbreaking work involving protein column efficiency. Wirth and colleagues guest authored the October instalment of “Column Watch,” in LCGC North America which explains how “slip flow” works and presents a practical example using this concept.

Your paper in the October issue of LCGC North America discusses the phenomenon of “slip flow.” What is slip flow?

Slip flow has the effect of flattening the parabolic flow profile to reduce band broadening. Slip flow is a nanoscale phenomenon that enhances the flow rate of a liquid when the interactions between the liquid and the walls are weak. Such weak interactions occur in reversed-phase liquid chromatography between polar solvents and hydrocarbon surfaces. Submicrometer particles have the nanoscale interstitial dimensions needed for significant slip flow.

Is this phenomenon important for separations of both proteins and small molecules?

Our present results point to importance for both. Proteins diffuse slowly, so the parabolic flow profile can impart quite a bit of broadening in conventional media, and slip flow reduces this broadening. The advantage is likely to extend to small molecules. These require smaller particles. We have preliminary results for particles as small as 110 nm in diameter showing very large flow enhancements, so the advance could eventually impact both small-molecule pharmaceuticals and protein separations.

What impact do you expect this discovery to have on the separations community?

People are interested in something new, so it will likely stimulate discussion and research, and enable new applications. Ultimately, new instrumentation will be needed to take full advantage of the narrower peaks from slip flow. Instruments always impart some broadening, so any advances in column efficiency necessitate advances in the instrumentation.

Your research web site says that you work at the interface of chemistry and medicine, with the goal of creating technology for earlier detection of disease. How does your work on slip flow fit into that vision?

Liquid chromatography (LC) already plays a role in biomarker detection for disease diagnostics, giving reproducible separations of proteins, peptides and metabolites, all with mass spectrometry (MS) compatibility. Earlier detection of diseases requires higher sensitivity because aberrant species would generally be at lower concentration. Slip flow should increase sensitivity because narrower peaks mean more intense peaks. Also, in biomarker discovery, the complexity of biological samples is a challenge, so the higher resolution from slip flow ought to be a benefit as well. Finally, throughput has been a big limitation in biomarker discovery, and the higher speed should help throughput.

Have your collaborations with medical researchers and industrial scientists had an impact on the way you approach your research?

Yes! It is essential to talk with practitioners frequently because they know what the real problems are and they know what the real needs are. In academics, we are free to pursue a new discovery, and slip flow in chromatographic media is a good example of that. We have one foot in the nanofluidics world so that we can advance the understanding of the phenomenon of slip flow. We have the other foot in the medical world, where our discussions with clinicians and industrial scientists help us direct these advances toward real problems.

What research questions do you hope to answer in the next few years?

Since we have not optimized flow rate yet, we would like to know how far slip flow can be pushed in terms of plate numbers. We would also like to know how small the particles can be to address column capacity.

We are beginning to use a commercial nanoLC instrument. This will reveal how much benefit can be gained from existing instruments and what instrumental advances are needed to enjoy the full advantages of slip flow.

Since the peaks are so narrow, our published work used on-column detection of labeled proteins by fluorescence microscopy. We want to know if nanospray MS detection will preserve the high gain in column efficiency from slip flow.

Absorbance detectors are more common than MS, but they require a long pathlength, which would give enormous broadening for our capillaries. We would like to see if label-free, on-column detection can be done. This is possible because the submicrometer particles form colloidal crystals, and many publications have shown that colloidal crystals act as diffraction gratings, which are sensitive to the composition of the liquid inside. So it is conceivable that the absorbance detector can be avoided in general use.

Related content:

http://www.chromatographyonline.com/SlipFlow

ADVERTISEMENT

LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
CHROMacademy
Are you aware of CHROMacademy and the learning tools it offers?
Yes, I would like to know more
Yes, I am not interested
No, I would like to know more
No, I am not interested
Yes, I would like to know more
70%
Yes, I am not interested
8%
No, I would like to know more
17%
No, I am not interested
4%
View Results

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC Europe eNews,
Click here