Extraction of Acrylamide from Fried Potato Chips (Crisps) Using ISOLUTE® SLE+ Prior to LC–MS–MS Analysis - - Chromatography Online
Extraction of Acrylamide from Fried Potato Chips (Crisps) Using ISOLUTE® SLE+ Prior to LC–MS–MS Analysis


The Application Notebook
pp. 704

Acrylamide has been found to occur in many cooked carbohydrate-rich foods and is of concern as a possible carcinogen. The supported liquid extraction (SLE) method described in this application note achieves high recoveries of acrylamide in fried potato chips (crisps). The method is sensitive enough to potentially measure levels as low as 10 ppb in a popular brand and flavour and has also been tested in flavoured varieties that were both machine and hand fried.

Extraction Procedure

Format: ISOLUTE SLE+ 1 mL Sample Volume Columns, part number 820-0140-C
Sample pre-treatment: Add 10 mL water to 1 g crushed sample (previously spiked with 13C3 acrylamide internal standard), and mix for 1 h. Centrifuge and remove a 0.65 mL aliquot of the aqueous layer, taking care not to include any of the thin upper oil layer.
Sample loading: Load the pre-treated sample (0.65 mL) onto the ISOLUTE SLE+ column. Apply a pulse of vacuum or positive pressure to initiate flow. Allow the sample to absorb for 5 min.
Analyte elution: Elute with ethyl acetate–tetrahydrofuran, (1:1, v/v, 2 2.5 mL) and allow to flow under gravity into tubes containing 2 L ethylene glycol. Apply vacuum or positive pressure to elute any remaining extraction solvent.
Post elution: Dry the volatile constituents of the eluate in a stream of air or nitrogen. Reconstitute in water (200 L).

HPLC Conditions


Table 1: Positive ion mode – MRM parameters.
Instrument: Waters Acquity
Column: Phenomenex Hydro 4 m 50 mm 2 mm C18 column with a C18 guard cartridge and on-line filter
Mobile phase A: 0.1% formic acid in water
Mobile phase B: 0.1% formic acid in methanol
Flow rate: 0.3 mL min-1
Injection volume: 10 L
Gradient: Initial 100% A, hold for 0.6 min; Linear ramp to 100% B over 0.25 min (0.85 min), hold 1.65 min (2.5 min); Linear ramp to 100% A in 0.01 min (2.51 min), hold 2.49 min (5 min)
Column temperature: 40 C
Sample temperature: 20 C

MS Conditions

Ions were selected on the basis of achieving maximum sensitivity using multiple reaction monitoring.


Figure 1: Chromatogram of untreated potato chip (crisp) extract showing level of incurred acrylamide.
Instrument: Waters Quattro Premier
Ionization mode: ES+
Desolvation temperature: 450 C
Source temperature: 120 C

Results and Conclusion

A method has been developed which measures acrylamide from a challenging matrix at highly sensitive levels. Despite acrylamide being a relatively polar molecule, excellent separation was demonstrated between this and matrix interferences on the ISOLUTE SLE+ material. The method delivers good recovery (90%), low %RSD (<10%), and is significantly easier, quicker, and more cost-effective to perform than SPE-based procedures, making it ideal for routine analysis of fried and baked foods.



Biotage AB
Vimpelgatan 5, Uppsala, Sweden
Tel: +46 18 56 59 00 fax: +46 18 59 19 22
E-mail:

Website: http://www.biotage.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors:

LCGC Columnist Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments in existing technology lines.

LATEST: The Role of Selectivity in Extractions: A Case Study

History of Chromatography | Industry Veterans:

With each installment of this column, a different industry veteran covers an aspect of the evolution and continued development of the science of chromatography, from its birth to its eventual growth into the high-powered industry we see today.

LATEST: Georges Guiochon: Separation Science Innovator

MS — The Practical Art| Kate Yu:
Kate Yu is the editor of 'MS-The Practical Art' bringing her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers.

LATEST: Mass Spectrometry for Natural Products Research: Challenges, Pitfalls, and Opportunities


LC Troubleshooting | John Dolan:

LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan, Vice President of LC Resources and world renowned expert on HPLC, is able to highlight common problems and provide remedies for them.

LATEST: LC Method Scaling, Part I: Isocratic Separations

More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here