Extraction of Antiepileptic Drugs from Serum and Urine Using ISOLUTE® SLE+ Prior to LC–MS–MS Analysis - - Chromatography Online
Extraction of Antiepileptic Drugs from Serum and Urine Using ISOLUTE® SLE+ Prior to LC–MS–MS Analysis


The Application Notebook
pp. 162

Antiepileptic drugs (AEDs) are prescribed to suppress seizures in epilepsy patients. The ability to therapeutically monitor these drugs in patients is necessary for maintaining optimal medical care and managing any adverse effects of the drug. In this application note ISOLUTE SLE + is demonstrated as an effective way to extract AEDs from serum and urine. The method was developed using a 96-well plate format to facilitate a high throughput workflow model.

Methodology

Format: ISOLUTEŽ SLE+ 400 μL 400 Supported Liquid Extraction plate, part number 820-0400-PO1

Sample pre-treatment: Pipette serum/urine (blank, calibrator, and patient [100 μL]) into a container. Add 50% aqueous formic acid (100 μL). Add up to 100 μL of internal standard.

Sample processing: Load up to 300 μL of pre-treated serum sample onto the ISOLUTE SLE+ 96-well plate. Apply a pulse of positive pressure or vacuum and allow samples to sit for 5 min.

Analyte elution: Elute analytes with methyl tert-butyl ether containing 1% (v/v) trifluoroacetic acid (conc) solution (2 × 700 μL). Allow sample to flow through by gravity and collect eluent. Apply positive pressure or vacuum as needed to facilitate a constant flow of approximately 1 mL/min (10–12 drops).


Table 1: Multiple reaction monitoring (MRM) transitions for AED drugs in positive mode Turbo Ionspray.
Post extraction: Evaporate to dryness (45 °C for 15 min) and reconstitute sample in mobile phase.

HPLC Conditions

Instrument: Agilent 1200 Liquid Handling System (Agilent Technologies, Berkshire, UK)

Column: Phenomenex Gemini C18, 150 mm × 4.6 mm (5 μm)

Mobile phase: Solvent A: 5 mM ammonium formate with 0.01% (v/v) formic acid; Solvent B: methanol:acetonitrile (50:50 m v/v)




Gradient:

Mass Spectrometry Conditions

Instrument: Applied Biosystems/MDS Sciex 4000 Q-Trap triple quadrupole mass spectrometer (Applied Biosystems, Foster City, California, USA) equipped with a Turbo IonsprayŽ interface for mass analysis.

Ion source temperature: 700 °C

Results and Discussion


Figure 1: Plot of average recoveries for neutral and zwitterionic antiepileptic drugs at 20 ng/mL in serum and urine, pre-treated with 50% aqueous formic acid.
Use of an aqueous formic acid pre-treatment increases recovery of the zwitterionic gabapentine (Figure 1), compared with other pre-treatment strategies, but did not improve recovery of vigabatrine, because of its more polar nature. This results in a robust method giving high, reproducible analyte recoveries.

Biotage AB
Vimpelgatan 5, Uppsala, Sweden
Tel: +46 18 56 59 00 fax: +46 18 59 19 22
E-mail:

Website: http://www.biotage.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Standardized Testing of Silica as a Base Material for Difficult Bonded-Phase Preparative Applications


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Retention Time Variation Is Normal?


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here