Techniques for Structure Elucidation of Unknowns: Finding Substitute Active Pharmaceutical Ingredients in Counterfeit Medicines - - Chromatography Online
Techniques for Structure Elucidation of Unknowns: Finding Substitute Active Pharmaceutical Ingredients in Counterfeit Medicines


LCGC Asia Pacific


A recurring topic in this column is structural characterization including the underpinnings of mass accuracy and resolution and spectral interpretation.1–3 The need to determine the exact structure, especially of an unknown, is at the core of why we employ mass spectrometers in all the variations and "flavours". Mass spectrometry (MS) has developed into a mainstream analytical technique in the pharmaceutical industry where one of its major roles is identifying and characterizing new chemical entities by elucidating the structure of unknown compounds.

Jean-Claude Wolff, a primary contributor to this discussion as a scientist with GSK (Medicines Research Centre, Stevenage, Hertfordshire, UK) has found characterizing unknowns in potential counterfeit medicines both a challenging and excellent focus for his talents.

I would not suggest that elucidating the structure of impurities is easy. My purpose is merely to show how combined mass spectrometric techniques can gather the information necessary to elucidate the structure of an unknown compound. In so doing, I suggest a pathway that leads to identifying the structure of the molecule analysed. As an example, I offer the exercise of determining the substitute, active pharmaceutical ingredients in counterfeit medicines.

Often the only information available when considering a novel chemical entity might be that the new, unknown compound of interest belongs to a particular class of compounds — in natural product research alkaloids, for instance. Additional information ("background" information) becomes available after performing structure elucidation on impurities or degradants structurally related to a pharmaceutically active compound.

The World Health Organization (WHO) defines counterfeit medicines as:

Those medicines that are deliberately and fraudulently mislabelled with respect to identity and/or source. Counterfeiting can apply to both branded and generic products. Counterfeit products may include products with the correct ingredients or with the wrong ingredients, without active ingredients, with insufficient active ingredient, or with fake packaging.4

"I focus on products with the wrong active pharmaceutical ingredient whose identity I then determine using MS techniques. My sole guidance is that a compound [to be a probable suspect] be readily available commercially and that it be cheap." These are the key elements making the compound attractive to the counterfeiters. Note, however, that "substitute actives" are not necessarily so common that they will be found in libraries. Hence, the reason for structure elucidation.

Counterfeits with an incorrect active ingredient are commonplace. A study of antibiotics and antimalarials by the German Pharma Health Fund showed about 16% of cases investigated contained incorrect ingredients.5

The MS tools available to determine incorrect pharmaceutical ingredients are numerous: you can use a series of sample introduction techniques such as direct inlet or probe, gas chromatography (GC) and liquid chromatography (LC). The choice of inlet technique depends on a sample's characteristics. Inlet technique is also linked to the ionization method you choose — electron ionization (EI), chemical ionization (CI), atmospheric pressure chemical ionization (APCI), electrospray ionization (ESI) — or the newer techniques such as desorption electrospray ionization (DESI) or direct analysis in real time (DART). All of these techniques have been reported in this column in recent years — though admittedly we treated GC techniques rather lightly, an imbalance we shall soon redress.

Structural elucidation of unknowns (in this instance substitute, active pharmaceutical ingredients) demands using MS instruments capable of tandem MS and accurate mass measurements to determine the elemental composition of an analyte. Here the choice is large, ranging from ion trap to Fourier-transform (FT) ion cyclotron resonance via quadrupole time of flight (Q-TOF). Regardless of the instrument or ionization technique used, however, you can elucidate an unknown compound's structure by using the arsenal of MS techniques available. The goal here is simply to show a possible pathway and discuss its merits.

Structure elucidation of a substitute, active pharmaceutical ingredient in an antimalarial tablet: The most effective approach for identifying and characterizing counterfeit medicines is to analyse a suspect sample using LC coupled to accurate-mass, tandem MS. This approach is more precise and quantitative than direct-inlet analysis (probe or infusion) of a sample. LC provides distinct separation of the active or substitute-active pharmaceutical ingredient from the excipients. Moreover, using LC reduces the potential for ion suppression in the ion source. Another advantage of introducing a sample via LC is the ability to compare LC–UV and LC–MS profiles of a suspect sample with a genuine one.


ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All Comments submitted are subject to review, and may be delayed before posting. Comments will be posted at the discretion of the editor.
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
Do you use countercurrent chromatography?
Yes, frequently
Yes, sometimes
No, never
Yes, frequently
66%
Yes, sometimes
5%
No, never
29%
View Results
Source: LCGC Asia Pacific,
Click here