Rapid Improvements for LC–MS-MS Analysis Using the New Phree Phospholipid Removal Plates - - Chromatography Online
Rapid Improvements for LC–MS-MS Analysis Using the New Phree Phospholipid Removal Plates


The Application Notebook

When performing LC–MS-MS analysis, phospholipids are perhaps one of the most troublesome components of bioanalytical samples. The presence of phospholipids not only reduces the column lifetime and sensitivity but can also cause a phenomenon known as ion suppression.

This work compares the presence of phospholipids in plasma samples after two different sample preparation techniques, protein precipitation and simultaneous protein precipitation and phospholipid removal using a new product, Phree.

Experimental Conditions


Table 1: Sample preparation protocols.
Plasma samples from the same lot were prepared (Table 1) and were injected on a Kinetex® 2.6 µm C18 core–shell column coupled with an API 3000 MS (AB SCIEX) (Figure 1). A total phospholipid profile was monitored using m/z 184–184.

Results and Discussion


Figure 1: Total phospholipid profile.
When monitoring the total phospholipid profile, the protein precipitated plasma showed a large amount of phospholipids. In comparison, the plasma sample that was prepared using Phree showed virtually no phospholipids (Figure 1).


Figure 2: Column sensitivity after 250 injections.
We also studied analyte response by comparing diclofenac spiked plasma samples. After sample preparation, 20 µL samples were injected on a Kinetex® 2.6 µm C18 core–shell column. The diclofenac signal in the protein precipitated samples was immediately lower than the signal in the Phree extracted samples. The signal in the protein precipitated sample rapidly decreases. The Phree extracted sample shows a steadier signal and is able to withstand > 250 injections (Figure 2). The higher signal strength of the Phree extracted sample is due to the reduction of phospholipid induced ion suppression. The significant reduction in phospholipid build up on the column and MS source also resulted in an increase in column lifetime and a decrease in the amount of MS maintenance required.

Conclusion

By preparing samples with Phree, analysts can remove proteins and phospholipids in 4 short steps resulting in immediate improvements to their chromatography work.

For more information about Phree, visit http://www.phenomenex.com/Phree.

Phenomenex Inc.
411 Madrid Avenue, Torrance, California 90501, USA
tel. +1 (310) 212 0555 +1 (310) 328 7768
Website: http://www.phenomenex.com

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
Sharing new techniques
Viewing the latest products and applications
Networking
Learning
Sharing new techniques
36%
Viewing the latest products and applications
27%
Networking
9%
Learning
27%
View Results
Source: The Application Notebook,
Click here