Immunosorbents for Selective Sample Preparation of Complex Mixtures - - Chromatography Online
Immunosorbents for Selective Sample Preparation of Complex Mixtures


LCGC Asia Pacific


Despite the advances that have been made in the sensitivity and selectivity of analytical instrumentation, for complex samples encountered in environmental and food matrices, some form of sample pretreatment is almost always required to isolate the analytes of interest. For liquid samples, the use of liquid–liquid extraction (LLE) and solid-phase extraction (SPE) have been the most popular approaches to pretreat these samples. Many methods have been published using one or both of these techniques to simplify complex mixtures before chromatographic or spectroscopic analysis. In some instances, LLE has given way to SPE because there has been a trend to eliminate or reduce the consumption of organic solvents, especially chlorinated solvents. Even though there have been many new SPE sorbents introduced that provide more selective interactions, workers still prefer to use nonselective hydrophobic sorbents for their sample preparation needs.1 Even in sorbents considered to be more selective, such as ion exchange SPE packings, analyte retention is often highly influenced by hydrophobic interactions with the organic portion of the sorbent surface. When these interactions occur, coextraction of other analytes and matrix interferences generally occurs, often a big problem when the analytes of interest are in trace concentrations. Then, additional clean-up procedures are required, increasing the chance of analyte loss and contamination by the introduction of other chemical species.

Thus, there is a need for clean-up methods that will result in more selective extraction of the analytes of interest, hopefully in a single step. In the past decade, there has been an increasing level of interest in the development of highly selective sorbents based upon immunoaffinity interactions. Such sorbents involve antigen–antibody interactions that are based upon molecular recognition and, thus, are very specific. The antibodies are immobilized onto solid supports such as silica or polymeric adsorbents that are classified as immunosorbents (ISs). Then, these sorbents are used in a manner similar to SPE cartridges. The analyte isolation process is termed immunoextraction or immunoaffinity extraction. Sometimes, when the experiment is performed in a column format, it is referred to as an immunoaffinity chromatography (IAC) but the process involved is not really a chromatographic mechanism.

There have been many papers published on the subject of ISs and immunoextraction but only recently have commercial products become available that can be applied conveniently to real world applications. This instalment of "Sample Prep Perspectives" will be devoted to an introduction to ISs and their potential applications for small molecules that might be encountered as pollutants in environmental and food samples. For those who would like in-depth coverage, there are a number of excellent review articles2–4 and most recently a textbook that has chapters devoted to immunoaffinity and other bioanalytical techniques.5

In an earlier article, immunoaffinity products that were designed to remove high-abundance proteins from biological samples were covered.6 In principle, these products are similar to those discussed here but are optimized for the immunoaffinity interactions for large molecules. In addition, an earlier article discussed molecular imprinted polymers (MIPS),7 that, in principle, mimic antibodies but involve the formation of a specific molecular imprint (cavity) on the surface of the sorbent formed by a template molecule while ISs are prepared by the immobilization of an antibody onto the surface of a solid support.


ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All Comments submitted are subject to review, and may be delayed before posting. Comments will be posted at the discretion of the editor.
LCGC E-mail Newsletters
Select each newsletter you wish to subscribe for:
Monthly
| Weekly
| Monthly
Survey
How do you make an LC column last longer?
Filter your sample and mobile phase
Adjust the pH of the mobile phase to within the working range of the column
Flush the column with methanol or acetonitrile if it's not going to be used for a while
All of the above
Some of the above
None of the above
Filter your sample and mobile phase
5%
Adjust the pH of the mobile phase to within the working range of the column
3%
Flush the column with methanol or acetonitrile if it's not going to be used for a while
8%
All of the above
64%
Some of the above
18%
None of the above
2%
View Results
Source: LCGC Asia Pacific,
Click here