Simultaneous Extraction of Ethyl Glucuronide (EtG) and Ethyl Sulphate (EtS) from urine with EVOLUTE AX prior to LC–MS-MS analysis - - Chromatography Online
Simultaneous Extraction of Ethyl Glucuronide (EtG) and Ethyl Sulphate (EtS) from urine with EVOLUTE AX prior to LC–MS-MS analysis


The Application Notebook



Table 1: Positive ions acquired in multiple reaction monitoring (MRM) mode.
This application note describes the extraction of both ethyl glucuronide (EtG) and the lower concentration metabolite ethyl sulphate (EtS) from human urine using EVOLUTE AX columns followed by LC-MS-MS analysis. EtG and EtS are specific metabolites formed within the body after the ingestion of alcohol. EVOLUTE AX is a resin-based mixedmode sorbent with an optimized combination of non-polar (hydrophobic), polar (hydrophilic) and weak anion exchange interactions ideally constructed to extract acidic analytes such as ethyl glucuronide.

Extraction Conditions

Sample pre-treatment: Dilute urine samples with acetonitrile (1:9, v/v)

Column conditioning: Methanol (3 mL)

Column equilibration 1: Water (3 mL)

Column equilibration 2: Acetonitrile (3 mL)

Sample loading: Load pre-treated sample (2 mL total volume) onto EVOLUTE AX 100 mg/3 mL column (part number 6130010-B)

Interference elution 1: Acetonitrile (3 mL)

Interference elution 2: Methanol (3 mL)

Analyte elution: 2% HCl in acetonitrile (3 mL)

Evaporation: Evaporate to dryness at 40 °C

Reconstitution: Reconstitute in HPLC grade water (250 µL), vortex mix and add acetonitrile (250 µL), vortex mix

Analytical Conditions

Instrument: Waters Acquity UPLC with 20 µL injection loop interfaced to Premier XE triple quadrupole mass spectrometer equipped with an electrospray interface for mass analysis.

Column: UPLC HSS C18 (150 mm × 1.8 µm × 2.1 mm i.d.)

Mobile phase: A = 0.05% Formic acid (aq) B = Acetonitrile

Flow rate: 0.2 mL/min

Injection volume: 10 µL (partial loop with overfill)

Sample temp: 10 °C

Column temp: 50 °C

Desolvation temp: 450 °C

Source temp: 100 °C


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 
Survey
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
If you are you planning to attend HPLC2013 Amsterdam what is your main reason?
Sharing new techniques
Viewing the latest products and applications
Networking
Learning
Sharing new techniques
36%
Viewing the latest products and applications
27%
Networking
9%
Learning
27%
View Results
Source: The Application Notebook,
Click here