Simultaneous Quantitative and Qualitative Measurements in a Single Workflow to Increase Productivity in Primary Drug Metabolism Investigations - - Chromatography Online
Simultaneous Quantitative and Qualitative Measurements in a Single Workflow to Increase Productivity in Primary Drug Metabolism Investigations

The Application Notebook
pp. 167-168

The ability to simultaneously collect quantitative and qualitative information from a DMPK analysis has the potential to significantly increase productivity in pharmaceutical drug discovery and development. We present a single workflow allowing P450 drug clearance values to be determined as well as metabolites identified, profiled, and their structures elucidated. To be able to do all of this on a high throughput UHPLC chromatographic timescale is essential for the high levels of productivity required for today's DMPK screening laboratories. Haloperidol provides a good example of what can be achieved.



Haloperidol

Workflow and Protocol


Figure 1: In a single workflow, data dependent MS–MS spectra identify and elucidate metabolite structures and drug clearance is measured.
Microsomal incubations were carried out by Unilabs Bioanalytical Solutions at 1 M drug concentration and a protein concentration of 0.5 mg/mL. Aliquots were taken and quenched with acetonitrile containing propranolol as an internal standard at eight time points over a period of 60 min.

Chromatography

Column: Fortis, 1.7 m, H2O, 2.10 mm 30 mm
Column temperature: 30 C
MPA: 0.1% formic acid in 95% H2O/CH3CN
MPB: 100% CH3CN
Gradient: 0.0 0.3 2.0 2.5 2.6 3.0 min
MP %: 95 95 5 5 95 95 %
Flow rate: 300 L/min
Injection volume: 5 L


Figure 2: Metabolite detection software compares the data file for the drug (in this case t60) with the corresponding control sample. A base peak chromatogram of the difference is created allowing the metabolites to be easily observed and their mass determined to four decimal places.
The high surface area and lipophilic ligand combined with a hydrophilic end cap give this stationary phase a broad selectivity and resolving power for the target drug and the metabolites. The use of small particles allows UHPLC to compress the peak into a tighter and taller peak, therefore enhancing detection of very low level analytes.

Metabolite Detection


Figure 3: Time profiles for the disappearance of haloperidol and the appearance of three metabolites.
Metabolite detect software compares the data file for the drug (in this case t60) with the corresponding control sample. A base peak chromatogram of the difference is created, allowing metabolites m/z 354, 212, and even 392 to be easily observed.


Figure 4: Linear calibration of 50 pg/mL to 50 ng/mL (3 decades) was achieved using the XIC for the measured m/z of each metabolite +/- 0.005 Da. R2= 0.9974.
Metabolite detection software is able to detect the m/z = 392 metabolite even though it coelutes with the internal standard.

Drug and Metabolite Profiles


Figure 5: The structure of metabolite m/z = 354 is easily identified using SmartFormula 3D to understand the fragmentation pattern.
Integration is carried out on the XIC for the measured m/z of each metabolite +/- 0.005 Da. Plotting the ratio of metabolite to internal standard (M/IS) versus time produces the metabolite profiles. Half-life and clearance values are determined from the natural log (ln) of the drug profile versus time plot.

Linearity



MS–MS data was not available for m/z = 392 because of coelution with the internal standard. The high quality data available, even for such a small peak, means SmartFormula is still able to predict the formula and deduce that it is a mono-oxidative metabolite.

Comparison with 3Q


Figure 6: The structure of metabolite m/z = 392 is easily identified using SmartFormula 3D to understand the fragmentation pattern.
Both the AB Sciex API 5000 and Bruker impact QTOF yield equivalent results for the clearance values. This can be clearly seen by comparing the ln [Drug]/[IS] versus time plots.


Figure 7: Clearance data from impact.
The linearity and gradients of these plots are nearly identical and result in values for t1/2 of 45 and 47 min, respectively.


Figure 8: Clearance data from 3Q.
The difference in y intercept is a result of a difference in relative response of the internal standard and has no influence on the clearance results.

Conclusions

The quan–qual workflow is effective and robust using a rapid analytical method suitable for high throughput screening at 1 M drug concentrations.

Metabolite detection software allows metabolites to be rapidly identified and profiled even when compounds coelute.

Bruker Daltonics Inc.
40 Manning Road, Billerica, Massachusetts 01821, USA
Tel: (978) 663 3660 fax (978) 667 5993
Website: http://www.bruker.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here