Tips & Tricks GPC/SEC: How to Test GPC/SEC Columns - - Chromatography Online
Tips & Tricks GPC/SEC: How to Test GPC/SEC Columns


The Column
Volume 10, Issue 14, pp. 710

Gel permeation chromatography/size-exclusion chromatography (GPC/SEC) columns are at the heart of a GPC/SEC system. Even when advanced detection, such as on-line light scattering, viscometry, or mass spectrometry (MS), is applied the quality of the results strongly depends on the least expensive part of the system, the separation columns. An easy test to determine the plate count, the peak symmetry, and the resolution of the columns is therefore an essential tool to ensure GPC/SEC system functionality.


PHOTO CREDIT: MARRIO31/GETTY IMAGES
Column performance is often described using plate count (or plate count/m), peak symmetry or asymmetry, and specific resolution. It is good practice to regularly monitor these parameters and several national and international standards — such as ISO 13885, DIN 55672, ASTM D 5296-05, and others — provide acceptance criteria for these parameters in dedicated solvents. If column certificates with test data are available to demonstrate column performance, it is good practice to repeat the tests with the specified conditions after column installation. Please note that comparisons of overall column performance can only be done if pertinent variables are kept constant. Mobile phase, test material, flow-rate, temperature, column loading, detector, and system including tubings should be kept the same, otherwise numerical results cannot be compared.


Table 1: Potential test materials to determine plate count, peak symmetry, and resolution.
Plate Count and Asymmetry: To measure plate count and asymmetry, a monodisperse test material can be injected onto the gel permeation chromatography/size-exclusion chromatography (GPC/SEC) instrument. Detailed experimental conditions can be found either in a corresponding GPC/SEC standard or on the column quality certificate. Alternatively, suitable test materials for a variety of typical GPC/SEC solvents are presented in Table 1. For these materials an injection volume of 20 ÁL (or if possible less) can be used, whereas concentration should be adapted to the number of columns used: One column = 0.2 g/L; two columns = 0.5 g/L; three columns = 1 g/L. As mentioned before, it is important to use the same test substance with the same instrument and analytical conditions for all tests.




The calculation of the theoretical plate count per metre (Nth [1/m]) uses the peak position and the peak width when at half peak height as according to:

where σ is the variance estimated by the half-height method, w1/2 and L is the column length in cm.

The asymmetry calculation strongly depends on the GPC/SEC standard applied. The peak asymmetry in DIN 56672 and ISO/EN 13885 are defined as A = wl/wr where wl and wr are the peak widths on the left and right side of the peak maximum (measured in 10% of the peak height). In ASTM asymmetry is defined as A' = wr/wl, therefore A' = 1/ A.

In general, resolution is the most important factor because it gives direct information about the performance of a separation in a dedicated molar mass region. To measure resolution, a mixture of polymer standards should be injected onto the system and analyzed. Alternatively, if a calibration curve is available the required information, such as the slope (D), can also be taken from this calibration.




Specific resolution (Rsp) specifies the quality of resolution of two peaks (denoted as Rs), whose molecular weight differs by one order in magnitude:


Figure 1: Example of plate count, peak symmetry, and resolution determination based on the acceptance criteria from ISO 13885/DIN 55672.
Figure 1 shows the results of a test to determine plate count, peak symmetry, and resolution using butylated hydroxytoluene (BHT) in tetrahydrofuran (THF). Numbers in green indicate that the acceptance criteria of the applied standard ISO 13885/DIN 55672 are met.

Regular plate count determination is important if a lot of different samples and samples with unknown composition or origin are analyzed. Testing before and after analysis can help to identify column damage because of inappropriate sample preparation or can detect samples that cannot be analyzed using the applied conditions. In quality control, where the same type of sample is analyzed, a check-out sample of the same chemistry can be analyzed in addition to ensure that the actual application is running properly.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: Avoiding Reversed-Phase Chromatography Problems Through Informed Method Development Practices: Choosing the Stationary-Phase Chemistry


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Seven Common Faux Pas in Modern HPLC


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Estimating Resolution for Marginally Separated Peaks


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Column,
Click here