Transient Protein Self-Association Determined by SEC–MALS


The Application Notebook
pp. 169

Domain antibodies (dAbs) are the smallest functional binding units of antibodies, corresponding to the variable regions of either the heavy (VH) or light (VL) chains of human antibodies. Different dAbs can show different degrees of self-association, which drives the need for a cheap, quick — yet robust — technique to study self-association behaviour in drug discovery. This study outlines the use of SEC–MALS to determine self-association, and validates the work by comparison with analytical ultracentrifugation (AUC).

Methods

Shimazdu LC-20AD Prominence was used in series with miniDAWN TREOS and Optilab REX for this study. The columns used were TSK-GEL3000SWXL and Superdex S200.


Table 1: List of the samples injected, Mw and concentration as determined by SEC–MALS.
dAb_1 was expressed, purified, and dialyzed into PBS, pH 7.4 and submitted to SEC–MALS and AUC. For MALS the sample was concentrated to 2.4 mg/mL, 7.0 mg/mL, 75 mg/mL, and 130 mg/mL and followed by appropriate dilution to obtain nine different concentrations that were run on SEC–MALS (Table 1).


Figure 1: Overlay of UV, dRI, and MALS signal from SEC–MALS for the sample with stock concentration 15 mg/mL.
K2,1 (Propensity of dAb_1 to dimerize) was determined by plotting a graph of Mw versus log10 concentration obtained from SEC–MALS mid peak.


Figure 2: Overlay of the five different sample results from SEC–MALS highlighting the ’Determined Mw’. Samples with stock concentration 5 mg/mL, 15 mg/mL, 30 mg/mL, 60 mg/mL, and 120 mg/mL represent the smaller to bigger peak, respectively.
The K2,1 values for dAb_1 from SEC–MALS was calculated as 300 μM compared to 391 μM from the AUC experiment.

Conclusion

While AUC is the gold standard for the determination of self-association in solution, scientists are looking for alternatives as a result of AUC's high costs, low throughput, and expertise requirements. Here, SEC–MALS has shown excellent agreement with AUC results and can be used as a cheaper, easier, and quicker alternative.

This note was graciously submited by M. Burman and O. Schon, GlaxoSmithKline, 315 Cambridge Science Park, Cambridge CB4 0WG, United Kingdom.

Wyatt Technology Corporation
6300 Hollister Avenue, Santa Barbara, California 93117, USA
Tel: +1 (805) 681 9009 fax: +1 (805) 681 0123
Website: http://www.wyatt.com/

ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: The Application Notebook,
Click here