Column Diameter, Linear Velocity and Column Efficiency - - Chromatography Online
Column Diameter, Linear Velocity and Column Efficiency


LCGC Europe
Volume 2, Issue 23, pp. 72-75

I recently had a question regarding some changes that took place in the 2009 edition of the US Pharmacopeia. In USP 32, General Chapter 621,1 the adjustments allowed to meet system suitability for liquid chromatography (LC) methods include a change in column internal diameter of 25%. A change was made in USP 32 Supplement 2.2 This stated that column internal diameter "can be adjusted provided that the linear velocity is kept constant." The person was not sure how this adjustment of linear velocity should be made and wondered about what effect it would have on column efficiency.

Adjustment of Flow Rate

Much of the band broadening that takes place within an LC column is a result of slow diffusion of the analyte in and out of the pores in the column packing. This means that the mobile phase flow rate can have an affect on the volumetric width of a peak. By volumetric width, I mean the peak width in units of volume (for example, millilitres or microlitres), not time. In the days when 10 μm diameter particles were used as column packing, a noticeable reduction in column efficiency could be observed when the flow rate was increased.

The smaller the particle diameter, the less the dependence of column efficiency upon flow rate. Although it is easy to demonstrate a loss in efficiency with an increase in flow rate for very well-behaved analytes with 3 and 5 μm diameter particles, a two-fold change in flow rate is rarely noticed with real applications. With sub-3 μm particles, flow rate has little influence on column efficiency, even with well-behaved compounds.

In this discussion, I have referred to flow rate, but the important variable is not flow rate, but linear velocity. Linear velocity is the speed at which the mobile phase travels through the column, for example, in millimetres per second. For comparison of equivalent conditions between columns of different internal diameters, the linear velocity should be kept constant. To keep linear velocity constant, the flow rate should be adjusted in proportion to the column cross-sectional area, which is directly proportional to the square of the ratio of column diameters. For the current discussion, let's consider two cases. The first is a change from a 4.6 mm i.d. column to a 2.1 mm i.d. column:

(4.6/2.1)2 = 4.8 ≈ 5

and the change from 4.6 mm to 1.0 mm:

(4.6/1.0)2 = 21.2 ≈ 20

In both cases, we'll consider the approximations as close enough for practical work and certainly easier to remember and use for mental calculations.

This means that for equivalent linear velocities, a change from a 4.6 mm i.d. column operated at 1 mL/min to a 2.1 mm i.d. column would require a flow rate adjustment of five-fold, or to 1.0/5 = 0.2 mL/min. A change to a 1.0 mm i.d. column would mean a new flow rate of 1.0/20 = 50 μL/min.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors:

LCGC Columnist Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments in existing technology lines.

LATEST: The Role of Selectivity in Extractions: A Case Study

History of Chromatography | Industry Veterans:

With each installment of this column, a different industry veteran covers an aspect of the evolution and continued development of the science of chromatography, from its birth to its eventual growth into the high-powered industry we see today.

LATEST: Georges Guiochon: Separation Science Innovator

MS — The Practical Art| Kate Yu:
Kate Yu is the editor of 'MS-The Practical Art' bringing her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers.

LATEST: Mass Spectrometry for Natural Products Research: Challenges, Pitfalls, and Opportunities


LC Troubleshooting | John Dolan:

LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan, Vice President of LC Resources and world renowned expert on HPLC, is able to highlight common problems and provide remedies for them.

LATEST: LC Method Scaling, Part I: Isocratic Separations

More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC Europe,
Click here