How Raw Are Your Data — 2012? - - Chromatography Online
How Raw Are Your Data — 2012?


LCGC Europe
Volume 25, Issue 2, pp. 88-107

This column revisits the topic of raw data and electronic records for a chromatography data system (CDS) in light of recent updates in regulations and guidance issued by regulatory agencies and industry bodies. We look at the ways to define what raw data are for a CDS and pose the question: Is it still possible to define paper records as chromatographic raw data?

In this Questions of Quality column I want to return to a topic that I have discussed twice before in the past few years: What are the raw data and/or electronic records for a chromatography data system (CDS)? In December 1996 I published the first Questions of Quality on raw data (1) and, with the timing that snatched defeat out of the jaws of victory, I was in print before the final rule on electronic records and electronic signatures was issued by the FDA in March 1997 (2). Therefore in 2000, I returned to the subject of raw data in light of industry and my experience with Part 11 to discuss and define the scope of electronic records that you could have with a CDS (3). More recently, in 2011, there was the revision of EU GMP Annex 11 on computerized systems (4) and Chapter 4 on documentation (5) that I reviewed in this column last year (6). Although the Questions of Quality column focused on Annex 11, there were significant changes to Chapter 4 on documentation that I want to focus on in this column. This is the new requirement to define raw data for systems involved in making quality decisions (5). As a CDS will be used to release raw materials, finished products and stability testing, it will be involved in making quality decisions and the raw data needs to be defined. Therefore what should we do for a CDS? Has anything changed since I wrote the first article on this topic over fifteen years ago?

To give you an idea of what we will be discussing I have presented the regulations, regulatory guidance and industry guidance documents to be discussed in Figure 1. First we'll look back and I'll give you a summary of the points from the Questions of Quality columns from 1996 and 2000 (1, 3), the latter article being greatly influenced by the publication of 21 CFR 11 regulations (2). Then we'll begin by looking at the FDA Guidance on Part 11 Scope and Application published in 2003 (7), which was expanded with the FDA interpretation published in 2004 (8) for electronic versus paper records for a CDS.


Figure 1: Outline of this month’s column on raw data.
Next, we will look in more detail at the specific EU GMP requirements for raw data from the new version of Chapter 4 on documentation (5). However, as Chapter 4 does not give a definition of raw data, we will look at the definitions in the US GLP regulations (9) and a Swiss GLP Guidance (10). Finally, we'll end with a look at what is said in the FDA's Compliance Programme Guide 7346.832 on Pre-Approval Inspections (11) and the second edition of the GAMP Good Practice Guide: RiskBased Approach to GxP Compliant Laboratory Computerized Systems (12).


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC Europe,
Click here