Questions From Afar - - Chromatography Online
Questions From Afar


LCGC North America
Volume 30, Issue 8, pp. 634-639

Temperature Control

One of the topics that comes up regularly in this column is the importance of controlling the column temperature. Retention times can change by 1–2% for each 1 C change in the column temperature, and peak spacing can change, as well. One of the illustrations I like to use to emphasize the importance of using the column oven rather than ambient temperature relates to a couple trips I took in 2008. In January I was in central China in a laboratory that was not heated, and the ambient temperature was 10 C; six months later I was in a temporary laboratory in Israel where, if the temperature wasn't 35 C, it certainly felt like it! These temperatures are drastically different from the 20–25 C that most of us would quote when asked to put a number on "ambient" temperature. Imagine trying to transfer a method between those two laboratories. In addition to control of the column temperature, it is important that the mobile phase be preheated sufficiently so that there is no more than a maximum of ≈6 C difference between the column inlet and outlet.

Following this discussion in one of the classes, an attendee asked about the influence of the temperature of the solvent reservoir and of the autosampler. My feeling is that the temperature of the solvent in the reservoir is not important, because by the time the solvent travels through the pump and autosampler and all the associated connecting tubing, any preheating or cooling provided in the reservoir will be canceled out by the thermal mass of the system. The control of the temperature of the autosampler is most commonly accomplished by heating or, more commonly, cooling the sample tray. The goal with autosampler cooling is to prevent sample degradation before injection, and because it only cools the sample it will have little or no influence on the mobile-phase temperature.

So the conclusion of the discussion is that you should always use the column oven to control the column temperature, even if it is just to set it a few degrees above room temperature — for example at 30–35 C — to ensure a constant column temperature and thus more stable retention times. You should cool the autosampler tray if your samples are likely to degrade at room temperature before they are injected. Heating or cooling the mobile-phase reservoir has little, if any, effect on the chromatography.

Broad Peaks with a New Column

Another class attendee wondered why he observed broad peaks for his sample after installing a new column. This is a good example of a time to apply the divide-and-conquer principle of troubleshooting — do a physical or mental experiment to divide the problem up into smaller parts so that you can eliminate as many possible causes as you can. Before doing anything, I would reinject the sample, and if the peaks were still broad, I would reinject a reference standard. If both of these injections are still bad, it is not a problem with the particular sample.

My first question would be to determine if the problem occurred because the column was replaced. Were the peaks of normal width before column replacement, but broader after? If so, then either the new column is bad, a nonequivalent replacement was used, or something happened in connecting the column to cause the problem. The connections at each end of the column may be the source of the problem, especially if polyether ether ketone (PEEK) tubing and fittings are used. If these are not seated properly, a small gap in the connection can result in extracolumn volume and corresponding peak broadening. When adjusting PEEK fittings, I recommend turning the pump flow off, loosening the fitting, pushing the tubing into the connection to ensure it hits the bottom, and then tightening the nut. If you try to tighten the fitting with the flow on, it can slip during the tightening process, creating exactly the problem you are trying to correct. Once again, reinject a reference standard to see if the problem is now corrected. If the problem persists, the next easy step would be to replace the column with another new column, double-checking that the manufacturer and part number is the same as the original column. While it is rare to receive a bad column today, it can happen. No harm is done by double-checking with another new column, and if it is good, you can just put the spare column in stock for later use.

Another check that should be made is to examine the chromatogram (or chromatograms) for other changes. Has the retention time of one or more peaks changed? Has the peak spacing (relative retention) changed? If either of these situations has occurred, it suggests a change in chemistry of the system. We've already eliminated the column as the source of problems through substitution of another new column. The remaining chemical possibilities are the mobile phase and the column temperature. If the problem origination coincided with a change in mobile phase, this is a likely source of the problem. For example, if acids or bases are present, an error in pH adjustment of the mobile phase might be the problem source. Because it is easy to check by substitution, make up a new batch of mobile phase and see if the problem is corrected. The final chemical possibility is a change in column temperature. Make sure the column oven is on and set to the proper temperature.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Sample Prep Perspectives | Ronald E. Majors: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments.
LATEST: UV Detector Problems


Perspectives in Modern HPLC | Michael W. Dong: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: Superficially Porous Particles: Perspectives, Practices, and Trends


MS — The Practical Art | Kate Yu: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting | John Dolan: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: Problems with Large-Molecule Separations


More LCGC Chromatography-Related Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC North America,
Click here