Questions From Afar - - Chromatography Online
Questions From Afar


LCGC North America
Volume 30, Issue 8, pp. 634-639

The remaining items on my list fall in the extracolumn effects category. We've already looked at the column connections, but extracolumn peak broadening can occur because of poor tubing connections wherever the sample contacts, so connections at the autosampler and detector should be checked as well, especially if PEEK tubing and fittings are used. The detector flow cell, time constant (noise filter), and data system data rate also can be sources of peak broadening. I'm assuming that the detector was not changed, so the flow cell is not likely the source. The detector time constant is an electronic filter that helps to reduce the noise through signal averaging, but if it is set to too large a value, it can result in broadened peaks. The rule of thumb is that it should be set to ≈0.1 times the width of the narrowest peak of interest. Similarly, the data system sampling rate should be set so that at least 15–20 data points are collected across the peak. Although neither of these settings is likely to change without intervention by the operator, a power outage or other factor might cause these to be reset to default values that are not appropriate for the present samples. It is easy to check both the time constant and data rate to be sure they aren't the source of the problem.

The final item on the extracolumn effects list is the possibility of the use of too strong of an injection solvent, too large of an injection, or both. My rule of thumb is that you can inject ≈15% of the volume of the first peak of interest if you inject in mobile phase. So, for example, if the first peak is 0.1-min wide and the flow rate is 1 mL/min, the peak volume is 100 μL. This translates to ≈15 μL for an injection volume if mobile phase is used as the injection solvent. If a stronger solvent is used for injection, smaller volumes should be used; if a weaker solvent is injected, larger volumes may be possible. The simplest way to check for injection problems is to inject a very small volume (for example, ≤5 μL) to see if the problem is corrected. I've seen a case where a method worked well for years, then one day started giving broad or split peaks for no apparent reason. In examining the conditions, it was discovered that too much of too strong a solvent was injected. By diluting the injection solvent and increasing the injection volume to maintain the same injection mass, the problem was solved. Unfortunately, too many methods are in use where the operating conditions are "right on the edge" of reliability — it takes only a minor, and often unidentified, change in conditions to cause them to fail.

The preceding discussion followed a specific series of steps to solve the problem. In any particular case, some of these steps could be eliminated by mental experiments or it may be more logical in your case to perform the test steps in a different order. Additional questions might also be appropriate, such as asking if the method ever worked, or if it was being transferred from another lab or a literature method. Do your best to eliminate possible causes by a bit of mental work before you start doing physical experiments to correct the problem.

References

(1) J.W. Dolan, LCGC North Am. 29(11), 982–986 (2011).

(2) L.R. Snyder, J.J. Kirkland, and J.W. Dolan, Introduction to Modern Liquid Chromatography, 3rd ed. (Wiley, Hoboken, New Jersey, 2010).


John W. Dolan
John W. Dolan
"LC Troubleshooting" Editor John Dolan has been writing "LC Troubleshooting" for LCGC for more than 25 years. One of the industry's most respected professionals, John is currently the Vice President of and a principal instructor for LC Resources, Walnut Creek, California. He is also a member of LCGC's editorial advisory board. Direct correspondence about this column via e-mail to
.


ADVERTISEMENT

blog comments powered by Disqus
LCGC E-mail Newsletters
Global E-newsletters subscribe here:




 

LCGC COLUMNISTS 2014

Column Watch: Ron Majors, established authority on new column technologies, keeps readers up-to-date with new sample preparation trends in all branches of chromatography and reviews developments. LATEST: When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration


Perspectives in Modern HPLC: Michael W. Dong is a senior scientist in Small Molecule Drug Discovery at Genentech in South San Francisco, California. He is responsible for new technologies, automation, and supporting late-stage research projects in small molecule analytical chemistry and QC of small molecule pharmaceutical sciences. LATEST: HPLC for Characterization and Quality Control of Therapeutic Monoclonal Antibodies


MS — The Practical Art: Kate Yu brings her expertise in the field of mass spectrometry and hyphenated techniques to the pages of LCGC. In this column she examines the mass spectrometric side of coupled liquid and gas-phase systems. Troubleshooting-style articles provide readers with invaluable advice for getting the most from their mass spectrometers. LATEST: Radical Mass Spectrometry as a New Frontier for Bioanalysis


LC Troubleshooting: LC Troubleshooting sets about making HPLC methods easier to master. By covering the basics of liquid chromatography separations and instrumentation, John Dolan is able to highlight common problems and provide remedies for them. LATEST: How Much Can I Inject? Part I: Injecting in Mobile Phase


More LCGC Columnists>>

LCGC North America Editorial Advisory Board>>

LCGC Europe Editorial Advisory Board>>

LCGC Editorial Team Contacts>>


Source: LCGC North America,
Click here