The use of superficially porous particles (SPPs) for modern high performance liquid chromatography (HPLC) is now very common. Initially, SPPs rose as an alternative to sub-2-µm fully porous particles (FPPs). In recent years, many column manufacturers have developed 2-µm and smaller SPP-based products. This article investigates the practical utility of these smaller SPP designs.
The use of superficially porous particles (SPPs) for modern high performance liquid chromatography (HPLC) is now very common. Initially, SPPs rose as an alternative to sub-2-µm fully porous particles (FPPs). In recent years, many column manufacturers have developed 2-µm and smaller SPP-based products. This article investigates the practical utility of these smaller SPP designs.
The introduction of modern superficially porous particles (SPPs) by Advanced Materials Development in 2006 revolutionized high performance liquid chromatography (HPLC). Interest in these particles, characterized by a solid core and a porous outer layer, stemmed from the very high efficiency they produce at modest back pressures (1). During the time of SPP emergence, efforts to improve efficiency focused on decreasing the size of packed particles. The term “ultrahigh-pressure liquid chromatography”, or UHPLC, was coined by Jorgenson in 1997, and was based on studies of nanobore columns packed with 1–1.5 µm particles (2). In 2004, Waters introduced the first commercial liquid chromatography system capable of operating at pressures as high as 1000 bar. Together with the availability of columns packed with sub-2-µm particles, the age of UHPLC was born. The small sub-2-µm particle size resulted in high system back pressures that required instrumentation capable of operating under such conditions. The SPP design proved to be a viable alternative to smaller fully porous particles (FPPs), because columns packed with subâ3âµm SPP materials provided similar efficiencies to subâ2âµm FPP columns at back pressures that were attainable on standard HPLC systems (3).
Since the initial rise of SPP columns with particle sizes around 2.7 µm, many manufacturers have developed stationary phases using smaller and larger particles, as well as particles with various pore structures. In 2014, Fekete and coauthors predicted that one future trend for UHPLC would be the incorporation of subâ2âµm SPP columns (4). The authors note that, because of significant improvements in eddy diffusion, moderate decrease in longitudinal diffusion, and some improvement in mass transfer resistance, SPPs offer a 30–50% increase in efficiency over same particle size FPPs, regardless of the particle size. Efficiency gains for smaller SPPs should thus trend similarly with what has been observed for FPPs. It was also noted that only two sub-2-µm columns were commercially available at that time.
In another paper, Fekete and Guillarme evaluated a 1.3-µm SPP phase, noting kinetic advantages over fully porous packed columns. Efficiencies of greater than 500,000 plates/m were observed as compared to 300,000 typically generated by a well-packed subâ2âµm FPP column. The authors noted that the performance of the column was limited by the current instrumentation. Lower extracolumn variance and higher pressure limits for both the column and the instrument were noted as potential improvements (5).
DeStefano and coauthors noted that the advantages of smaller SPPs include higher efficiency, improved detection sensitivity, less mobile-phase consumption per analysis, higher peak capacity, and the perception of stateâof-the-art technology. The authors also noted the following disadvantages:
The authors noted that smaller particles do provide utility for some applications, but the efficiency advantage often does not overcome the disadvantages (6).
The availability of commercial 2-µm and sub-2-µm columns midâyear 2015 was still limited. Advanced Material Technologies and Supelco had introduced 2-µm column lines with several surface modifications. Phenomenex offered both a 1.3âµm (C18) and a 1.7-µm version of the Kinetex brand. Thermo Fisher Scientific commercialized a 1.4-µm C18 phase, and Waters launched a 1.6-µm platform with bare silica and C18 surface chemistries (7). In a 2016 review, no new providers entered into the 2-µm and subâ2âµm SPP market; however, two additional stationary phases, Kinetex EVOâC18 and HALO Peptide ES C18 (MacâMod), were made available (8). In 2017, Agilent introduced a number of chemistries on 1.9-µm SPPs, including bare silica for HILIC. ChromaNik introduced SunShell in a 2-µm format, Waters introduced 1.6-µm Cortecs with additional chemistries, and Phenomenex introduced a 1.7-µm SPP column intended for characterization of RNA and DNA (9). It is clear from this list that the trend towards smaller SPPs has indeed continued.
There have been many publications in recent years about studies employing SPP phases of all sizes for applications in a variety of disciplines. Within the literature scanned for this instalment of “Column Watch”, no “applied” publications could be found that compared sub-3-µm SPP phases to sub-2-µm SPP phases explicitly. In a systematic evaluation of commercially available UHPLC columns for drug metabolite profiling, Dubbelman and coauthors found 1.3-µm and 1.7âµm SPPs to provide highly efficient separations. However, the system pressure experienced using the 1.3-µm particles was noted as an issue when higher flow rates were employed in an attempt to shorten run times (10).
So the question still remains, “Does one need to use 2-µm or subâ2âµm SPP phases, or do subâ3âµm SPP phases suffice?” The answer depends on the analysis at hand. For example, Figure 1 shows a reasonably complex separation of perfluorinated compounds (PFCs) on both a 1.8âµm SPP and a 2.7âµm SPP. With very little difference observed in terms of sensitivity, resolution, or peak shape, it would be difficult to justify running this method at the higher back pressure using the 1.8-µm phase. Figure 2 shows a comparison of 1.8-µm and a 2.7âµm SPP-based biphenyl columns performing a separation of artificial sweeteners. In this case, the 1.8-µm phase shows additional resolution over the 2.7-µm phase. The more complex the sample, the higher the probability that higher efficiency will be needed. Figure 3 provides a comparison of a bovine serum albumin (BSA) tryptic digest run on both a 1.8-µm and a 2.7-µm SPPâbased C18 column. As indicated by the higher peak capacity, using the smaller SPP column, the increased efficiency may become more important for such complex samples.
As with many so many questions when it comes to chromatography, the answer to whether one needs to enter into the 2-µm and subâ2âµm SPP realm is “It depends”. It appears that, as the complexity of the separation increases, the probability of needing the extra efficiency of a smaller SPP increases. In many cases, however, closely eluted compounds, even in a two- or threeâanalyte separation, may require added efficiency and thus smaller SPP-based phases. In practice, the use of subâ3âµm SPP columns is typically a suitable compromise for method development. Adjustment to smaller (or larger) particle sizes during method optimization may provide a more suitable final method.
References
Landon Wiest is an LC Applications Scientist at Restek. Over the last five years, Landon has worked developing LC products and now focuses his attention on LC applications in the food safety area. He completed his Ph.D. in analytical chemistry at Brigham Young University (Provo, Utah, USA) where he received instruction from Milton Lee and studied the viability of diamond, and other carbonâbased materials, as stationary phase supports for liquid chromatography under Professor Matthew Linford. Landon’s primary focuses have
been on developing novel chromatography phases on various supports and finding new applications for them that can increase the capabilities of the chromatography community. He has presented his research at numerous international conferences and published in multiple peer-reviewed journals. He has also been awarded patents for his inventions.
Shun-Hsin Laing is a Senior LC Applications Scientist at Restek. He received his bachelor’s degree from the National Taiwan University in 1988 and obtained his Ph.D. degree from the Michigan State University in 1996. He performed postdoctoral research at the University of Michigan from 1996 to 2000 in oncology studies. In 2001, he was appointed as Research Faculty in Penn State University and focused on molecular toxicology research. In 2006, he joined MPI Research Inc. as a Senior Research Scientist and played a role as Study Director for GLP analytical projects. In 2013, Dr. Liang joined the LC Solution of Restek Corporation and specialized in developing application methods across the fields of environmental, food safety, and life sciences.
Dan Li is an Application Scientist at Restek. Her work focuses on pharmaceutical, food, and environmental applications using LC–MS and GC–MS. Dan received her Ph.D. degree in analytical chemistry under the supervision of Professor Milton Lee from Brigham Young University. She was working in Covance performing bioanalytical method development and validation before joining Restek in 2015.
David S. Bell is a director of Research and Development at Restek. He also serves on the Editorial Advisory Board for LCGC and is the Editor for “Column Watch”. Over the past 20 years, he has worked directly in the chromatography industry, focusing his efforts on the design, development, and application of chromatographic stationary phases to advance gas chromatography, liquid chromatography, and related hyphenated techniques. His undergraduate studies in chemistry were completed at the State University of New York at Plattsburgh (SUNY Plattsburgh). He received his Ph.D. in analytical chemistry from The Pennsylvania State University and spent the first decade of his career in the pharmaceutical industry performing analytical method development and validation using various forms of chromatography and electrophoresis. His main objectives have been to create and promote novel separation technologies and to conduct research on molecular interactions that contribute to retention and selectivity in an array of chromatographic processes. His research results have been presented in symposia worldwide, and have resulted in numerous peerâreviewed journal and trade magazine articles. Direct correspondence to: LCGCedit@ubm.com
Sustainable Green Solvents in Microextraction: A Review of Recent Advancements
March 27th 2024Conventional sample preparation can be time- and resource-consuming, and a green analytical methodology can be a game-changer for scientists, in addition to facilitating selective and sensitive separations.
Transferring Methods to Compact and Portable HPLC
February 14th 2024The current trend in laboratory equipment design is the miniaturization of laboratory instruments. Smaller-scale HPLC instruments offer benefits that cannot be matched by analytical-scale equipment, especially in the areas of portability, reduced fluid volumes, and reduced operating costs. Yet, the miniaturization of laboratory equipment has brought with it a unique set of challenges, including transferring methods to compact LC. Capillary LC expands the use of LC to applications not currently done using conventional LC in a wide array of application areas, including pharmaceutical, food and beverage, petrochemical, environmental, and oil and gas. Greg Ward, Axcend’s CEO wrote, “Customers want an HPLC system with a small footprint, low flow rates and green chemistry.” Join his podcast where he shares method transfer in these application areas.
High-Throughput Analysis of Volatile Compounds in Air, Water, and Soil Using SIFT-MS (Apr 2024)
March 27th 2024This study demonstrates high-throughput analysis of BTEX compounds from several matrices (air, water and soil). Detection limits in the single-digit part-per-billion concentration range (by volume) are readily achievable within seconds using SIFT-MS, because sample analysis is achieved without chromatography, pre-concentration, or drying. We also present a calibration approach that enables speciation of ethylbenzene from the xylenes in real time.