Higher Order Mass Spectrometry Techniques Applied to Biopharmaceuticals

Higher Order Mass Spectrometry Techniques Applied to Biopharmaceuticals

The recent trends in mass spectrometric techniques - including native mass spectrometry (MS), ion mobility spectrometry (IMS), hydrogen–deuterium exchange MS (HDX MS), and chemical cross-linking MS (CXMS) - employed to elucidate higher‑order structures of protein complexes and the practical implementations of these methods are discussed.

 

  Since its invention in the early 20th century (1), mass spectrometry (MS) has been used to discover new chemical elements and their isotopes (2), explore martian soil for organic matter (3), and study biological processes by profiling proteomes or metabolomes (4,5). Biopolymers, and especially proteins, are the subject of intensive investigation. They are characterized by different levels of structural organization, ranging from the primary structure represented by the amino acid sequence, over secondary structural elements such as alpha-helices and beta-sheets, and the three-dimensional orientation of the polypeptide chain, and finally to the assembly of subunits into protein complexes.   Analysis of protein structure using MS was first possible in the mid 1980s, when the soft ionization techniques of electrospray ionization (ESI) (6) and matrix-assisted laser desorption–ionization (MALDI) (7) were introduced. Implementation of MS to protein analysis initially focused on large-scale identification (8) and the determination or confirmation of primary structure (9), whereas newer technologies have laid the ground for the study of tertiary and even higher order structures of protein molecules and complexes.    The analysis of biopharmaceuticals (therapeutic proteins developed for disease treatment) requires analytical techniques able to elucidate the various structural levels to ensure their efficacy and safety in patients. Several MS techniques are indispensable in the toolbox of physico-chemical characterization methods available for the analysis of therapeutic proteins (10). Some of the methods used in the elucidation of higher order structural elements of proteins - including native MS, ion mobility MS, hydrogen-deuterium exchange MS, and chemical cross‑lining MS - will be discussed in this feature article.  

Native Mass Spectrometry

 

Experiments using electrospray ionization mass spectrometry (ESI–MS) for the analysis of intact proteins were first performed by J.B. Fenn’s group (6). The study used strongly denaturing conditions created by using 50–90% organic solvent containing acetic acid or trifluoroacetic acid and was beneficial for the detection of multiple-charge protein ions with a quadrupole mass spectrometer of

m/z

1500 upper nominal mass limit. However, it was soon discovered that the charge-state distribution of electrosprayed proteins was significantly influenced by the protein structure prevalent under non-denaturing or denaturing conditions (11). This then led to the implementation of native MS.   Native MS aims to maintain the three-dimensional structure of proteins or protein complexes as much as possible during an experiment (12) by using conditions that reflect the protein’s native environment. The overall charge of a protein ion is limited by the number of ionizable functionalities that are accessible on the surface for charging, predominantly through protonation or deprotonation, leading to the observation of low-charged species in mass spectra. Weakly bound, noncovalent complexes - including proteins interacting with inhibitors, cofactors, metal ions, carbohydrates, or peptides - can be preserved during the electrospray process facilitating the study of the structure, stoichiometry, and association constant of such biomolecular complexes (13). The reduction of charge requires the use of mass spectrometers with an extended mass range such as time-of-flight (TOF) (14), or more recently orbitrap (15) mass analyzers to detect the low-charged protein species.  

(a) and 25+ (c) charge states with resolution of the different glycoforms: (a) 147 917.1 ± 1.1 Da (G0/G0F), (b) 148 061.7 ± 0.8 Da (G0F/G0F), (c) 148 222.4 ± 0.9 Da (G0F/G1F), (d) 148 383.8 ± 0.8 Da (G1F/G1F), and (e) 148 544.3 ± 1.0 Da (G1F/G2F). (b) and (d): IMS analysis of trastuzumab. Ion mobility mobilograms of trastuzumab in denaturing (b) or native (d) conditions. IMS data obtained in native conditions (d) reveal small amounts of dimeric mAb. Adapted and reproduced with permission from Analytical Chemistry84, Alain Beck, Sarah Sanglier-Cianférani, and Alain Van Dorsselaer, Biosimilar, biobetter, and Next Generation Antibody Characterization by Mass Spectrometry, 4637–4646 (2012) © American Chemical Society.

 

Trastuzumab (INN; trade name Herceptin) is a monoclonal antibody that interferes with the human epidermal growth factor receptor 2 (HER2) and is used to treat HER2-positive breast cancers. Figure 1(a) and 1(c) illustrate the difference between mass spectra of trastuzumab when analyzed under (a) denaturing versus (c) non-denaturing conditions. Under denaturing conditions (a) charge states from 33+ to 60+ are detected in an

m/z

range of 2200–4500, whereas non-denaturing conditions (c) yield charge states of 22+ to 28+ at

m/z

5200–6700. Deconvolution of both mass spectra gives equivalent masses for the uncharged species with masses in the range of 147–148 kDa and also reveals several different protein species that represent the different glycoforms of the monoclonal antibody. Such analysis can therefore readily reveal the glycosylation pattern of the protein, a highly important quality parameter of recombinant biopharmaceuticals.   Native MS has been shown to be a highly efficient tool for determining binding stoichiometry of a monoclonal antibody with its antigen. Humanized murine monoclonal antibody (hzmAb), directed against the junctional adhesion molecule A (JAM-A) to have antiproliferative and antitumoural properties, was titrated with its antigen and then analyzed using native MS to reveal non‑covalent complex stoichiometries (17). Three species were detected when equimolar amounts of antibody and its target antigen were incubated including free antibody, 1:1, and 1:2 antibody:antigen complex. Two molar equivalents of antigen led to an almost quantitative formation of the 1:2 complex, while eightfold molar excess yielded a 1:4 complex with a small portion of 1:3 complex. 
   

Ion Mobility Spectrometry

 

  Developed in the 1960s, ion mobility spectrometry (IMS) enables the generation of size- and conformation-dependent information that is not possible using MS alone. When coupled to mass spectrometry this technique has the potential to separate isomers, isobars, and conformers; significantly reduce chemical background; and detect aggregates of biopharmaceuticals. IMS separation of ions is possible using differing separation powers, analyte detection, and hyphenation to mass spectrometry (18) including: (1) drift-time ion mobility spectrometry (DTIMS); (2) aspiration ion mobility spectrometry (AIMS); (3) field-asymmetric waveform ion mobility spectrometry (FAIMS); or (4) travelling-wave ion mobility spectrometry (TWIMS). DTIMS and TWIMS are the two principles most often used in commercial instruments. In a DTIMS device ions are moved through a uniform, linear-field drift tube filled with a so-called “buffer gas” through a small, uniform electric field. The moving ions are attenuated by collisions with the buffer gas depending on their overall charge and collision cross-section (18). Ions with multiple charges and a small cross-sectional area move faster through the drift cell than low‑charge ions with large collisional cross-sections. TWIMS is a type of IMS that utilizes travelling waves created by a series of ring-shaped electrodes that split the structure of the drift cell into a series of segments (Figure 2). Instead of a uniform linear field, a high field is applied to one segment of the cell that is subsequently swept through the cell in the direction of ion migration. Consequently, movement and separation of ions in the mobility cell is accomplished by means of pulses of an electric field passing through.   An example for the application of TWIMS to the analysis of monoclonal antibodies is illustrated in Figure 1(b) and (d). IMS analysis of trastuzumab under denaturing conditions reveals a large number of highly charged species clustering at drift times between 10 and 15 ms, while native conditions clearly distinguish between the different, low-charged species in a drift time range of 7–25 ms. This contrast in drift behaviour is advantageous for the analysis of more complex mixtures of biopharmaceuticals, particularly when looking at sequence variants or other post-translational modifications such as oxidation or pyroglutamate formation.  

22, Petra Olivova, Weibin Chen, Asish B. Chakraborty, and John C. Gebler, Determination of N-glycosylation sites and site heterogeneity in a monoclonal antibody by electrospray quadrupole ion-mobility time-of-flight mass spectrometry, 29–40 (2007) © John Wiley and Sons.

 

Another practical example of IMS characterization of biopharmaceuticals (less directed towards higher order elucidation) is outlined in Figure 3. Here, a reduced mouse monoclonal antibody (IgG1, κ) sample comprising of both heavy and light chains was introduced into an ESI-quadrupole-IMS-TOF system (19). The two-dimensional ion mobilogram-mass spectrum depicted in Figure 3(a) clearly shows that light and heavy chains can be readily separated as different species without any other upfront separation technique. Multiple charged species related to the light and heavy chains were differentiated between using mass spectra extracted from the encircled areas in Figure 3(a) without mutually interfering signals. Extracted mass spectra (Figure 3[b] and [d]) were deconvoluted using a maximum entropy algorithm, yielding spectra of uncharged species, as illustrated in Figure 3(c) and (e). As expected, the light chain is detected as a single species, while the spectrum of the heavy chain reveals at least three glycoforms, characterized by different galactose content in the attached N-glycan (162 Da mass difference). This example nicely demonstrates the benefits of an additional dimension of separation, although an upfront separation method such as high performance liquid chromatography (HPLC) or capillary electrophoresis (CE) may be necessary - especially for the quantitative analysis of trace amounts of impurities that is often indispensable for quality control.  

Hydrogen-Deuterium Exchange Mass Spectrometry

  The native state of proteins is generally characterized by a tightly folded, compact structure that exposes a well-defined surface to its environment. Denaturation under conditions - such as high temperature, extreme pH, adsorption to surfaces, or dissolution in organic solvent - results in the protein unfolding and forming a significantly different surface. Similarly, interactions of a protein with other molecules such as small drugs, nucleic acids, or other proteins can lead to a significant change in surface properties.
    Denaturing and complex formation can also have a profound influence on the exchangeability of protons at the surface of a protein molecule - the acidic protons of the carboxyl groups or acidic side chains of aspartate and glutamate are normally the most easily and rapidly exchanged while the amide protons of the protein backbone are much less prone to exchange. Exchange can be monitored by dissolution of a protein in heavy water (D2O), which leads to a hydrogen exchange by deuterium (H/D exchange) in a few minutes to hours. In proteins, the exchange rates for the different hydrogen atoms strongly depend on accessibility and therefore on protein conformation or association into higher order structures. The substitution of exchangeable hydrogen atoms with deuterium atoms forms the basis of hydrogen‑deuterium exchange mass spectrometry (HDX-MS) (20).  

47, Kaspar D. Rand, Martin Zehl, and Thomas J.D. Jørgensen, Measuring the Hydrogen/Deuterium Exchange of Proteins at High Spatial Resolution by Mass Spectrometry: Overcoming Gas-Phase Hydrogen/Deuterium Scrambling, 3018–3027 (2014) © American Chemical Society.

 

A schematic workflow of HDX-MS is depicted in Figure 4. In brief, a protein with exchangeable hydrogens is dissolved for different periods of time at ambient or elevated incubation temperature (20–40 °C) in deuterated water (buffered to pH around 7). Depending on exchangeability, hydrogen atoms are replaced by deuterium atoms during the incubation time, before the exchange is quenched upon acidification and cooling to 0 °C. Proteins are then digested under quenching conditions, and the resulting peptides are separated by low-temperature HPLC, and finally analyzed by tandem mass spectrometry (MS–MS) upon fragmentation either by collision‑induced dissociation (CID) or electron-transfer dissociation (ETD). Characteristic mass shifts in the fragment ions are indicative for the presence and position(s) of deuterium atoms. Analysis of the kinetics of deuterium uptake yields information about the accessibility of hydrogens at different positions in the protein, which allows valuable insights into the three-dimensional structure of proteins or protein complexes.  

-1a (IFN) peptic peptides from four different H/DX MS comparability experiments. In each graph, the reference IFN data are the black lines with closed triangles, while the experimental IFN data, to which it is being compared, is the red line with open circles. Row a: Comparison of two different large scale IFN batches prepared over eight years apart; row b: Comparison of IFN versus N-terminally PEGylated IFN; row c: Comparison of IFN produced using different cell culture media and growth conditions; row d: Comparison of IFN versus oxidized IFN (oxidation of Met and Cys residues C17, M1, M36, M62, and M117 was 100%). Adapted and reproduced with permission from Journal of Pharmaceutical Science100, Damian Houde, Steven A. Berkowitz, and John R. Engen, The utility of hydrogen/deuterium exchange mass spectrometry in biopharmaceutical comparability studies, 2071–2086 (2010) © John Wiley and Sons.

 

 

HDX-MS has been successfully used to compare three-dimensional structures of biopharmaceuticals, which is essential to demonstrate manufacturing consistency to regulatory agencies or provide a proof of structural equivalence between an originator biopharmaceutical and its biosimilar. The advantage of this approach is that it probes the whole molecule instead of just certain substructures. The results of an interrogation of the three-dimensional structure of interferon-β-1a, a 20-kDa cytokine used to treat multiple sclerosis (traded under the names Avonex (Biogen), Rebif (Merck Serono or Pfizer), or CinnoVex (CinnaGen) as a biosimilar), are shown in Figure 5 (22). Hydrogen exchange rates determined for five different peptic peptides effectively show the impact of different manufacturing conditions as well as post-translational modifications - modification with poly(ethylene glycol) (PEG); or oxidation at C17, M1, M36, M62, and M117 - on protein structure. 
    No significant alteration in hydrogen-deuterium exchange profile was observed, even though production involved different batches, using different cell media, and was subjected to N-terminal modification with PEG. A significant impact was however found for methionine or cysteine oxidation, and because the oxidized peptides incorporated more deuterium compared to the reference analogues, it could be concluded that oxidized interferon-β-1a is more solvent exposed and less hydrogen bonded.    Although this example convincingly demonstrated the applicability of HDX-MS for revealing structural differences in homogenous biopharmaceuticals, the authors also pointed out that it is not capable, at the moment, to detect conformational differences in coexisting, low-level (<10%) components of the population (22).  

Chemical Cross-Linking Mass Spectrometry (CXMS)

  Three-dimensional structures of proteins can be determined with atomic resolution by using high‑resolution methods such as X-ray crystallography or nuclear magnetic resonance (NMR) spectroscopy, but the high amount of sample required for these methods (typically in the milligram range) makes them impractical for biological studies. In comparison, low‑resolution structural data generated by chemical cross-linking MS (CXMS) uses much less sample amounts (in the nanogram to picogram range) to generate highly valuable data (23). Low-resolution structural information is obtained by chemically cross-linking functional groups in a protein by means of a bifunctional cross-linker, which gives information about the distance of the cross-linked functional groups in a protein molecule or a protein complex.   The most common functional groups available for cross‑linking in proteins are the lysine amino groups. Sulphydryl groups of cysteines are another possibility, but they can become involved in the three-dimensional structure of a protein particularly when created by reduction of disulphide‑bridges in the native protein. Although formaldehyde is the oldest cross‑linking reagent, the most commonly utilized reagents are based on bifunctional N-hydroxy‑succinimide esters, which readily react with free amino groups (and in a side reaction also with hydroxyl groups of tyrosine) to create a stable amide or imide bond upon release of N-hydroxysuccinimide. Depending on the length of the cross-linking spacer, different distances of amino acids can be probed, ranging from (almost) zero for formaldehyde to 6.4 Å for disulphosuccinimidyl tartrate, 11.4 Å for bis(sulphosuccinimidyl)suberate (10 atoms), and 16.1 Å for ethylene glycol bis(sulphosuccinimidyl succinate (14 atoms) (24).   

 

Figure 6 shows an outline of a cross-linking experiment. After the formation of intramolecular or intermolecular cross-links, the protein or protein complex is proteolytically digested and the resulting peptides are analyzed via HPLC–MS–MS. Because the crosslink remains unaffected by the proteolysis, cross-linked amino acids are revealed through the corresponding cross‑linked peptides. To more easily identify cross-linked products, isotope-labelled cross-linking reagents with 50% heavy isotope exchange can be used. Thus, cross‑linked peptides are recognized by 1:1 doublets of mass signals for the light and heavy versions, which is usually achieved with the help of computer‑based searching algorithms (25). The distance information obtained from the cross‑linking experiment is then used to build and verify structural models for proteins or protein complexes.   Chemical cross-linking can also be utilized to directly analyze stabilized protein complexes. For example, disuccinimidyl suberate, as well as 1,1’-(suberoyldioxyl)bisazabenzotriazole) were used as cross-linkers to stabilize the complex between the bovine prion protein and a specific antibody against it, the antibody 3E7 (26). Direct analysis of the reaction products by matrix-assisted laser desorption–ionization MS revealed both the free prion protein and the free antibody together with 1:1 and 1:2 antibody:prion protein complexes.   

Conclusions

  In conclusion, MS, traditionally regarded as one of the most important analytical methods for the determination of the primary structure of proteins, is increasingly contributing to the elucidation of diverse fundamental aspects of the tertiary and even quaternary structure of proteins and protein complexes. In spite of providing less spatial resolution, the major strength of MS-based investigations over NMR spectroscopy or X-ray crystallography lies within the comparatively low amounts of sample required for successful analysis, typically a few picograms to nanograms. Such studies are, however, only feasible with substantial support through elaborate computational algorithms and workflows, which requires significant involvement of bioinformatics into data evaluation.       

Acknowledgements

  The financial support by the Austrian Federal Ministry of Economy, Family, and Youth, the National Foundation of Research, Technology, and Development, and by a Start-up Grant of the State of Salzburg is gratefully acknowledged.  

References

(1) J.J. Thomson,

Proceedings of the Royal Society of London Series a-Containing Papers of a Mathematical and Physical Character

89

, 1–20 (1913). (2) F.W. Aston,

Nature

105

, 547–547, (1920). (3) K. Biemann,

Proc. Natl. Acad. Sci. U.S.A.

104

, 10310–10313 (2007). (4) R. Aebersold and M. Mann,

Nature

422

, 198–207 (2003). (5) O. Fiehn,

Plant Mol. Biol.

48

, 155–171 (2002). (6) C.K. Meng, M. Mann, and J.B. Fenn,

Z. Phys.

10

, 361–368 (1988). (7) M. Karas and F. Hillenkamp,

Anal. Chem.

60

, 2299–2301 (1988). (8) J.K. Eng, A.L. McCormack, and J.R.I. Yates,

J. Am. Soc. Mass Spectrom.

5

, 976–989 (1994). (9) H. Nau and K. Biemann,

Abstracts of Papers of the American Chemical Society

62–62 (1974). (10) R.J. Falconer, D. Jackson-Matthews, and S.M. Mahler,

J. Chem. Technol. Biotechnol.

86

, 915–922 (2011). (11) J.A. Loo, H.R. Udseth, and R.D. Smith,

Biomed. Environ. Mass Spectrom.

17

, 411–414 (1988). (12) M. Przybylski and M.O. Glocker,

Angew. Chem. Int. Ed.

35

, 806–826 (1996). (13) J.A. Loo,

Int. J. Mass spectrom.

200

, 175–186 (2000). (14) M.C. Fitzgerald, I. Chernushevich, K. G. Standing, C. P. Whitman, and S. B. Kent,

Proc. Natl. Acad. Sci. U.S.A.

93

, 6851–6856 (1996). (15) S. Rosati, R.J. Rose, N.J. Thompson, E. van Duijn, E. Damoc, E. Denisov, A. Makarov, and A. J. R. Heck,

Angew. Chem.-Int. Ed.

51

, 12992–12996 (2012). (16) A. Beck, S. Sanglier-Cianferani, and A. Van Dorsselaer,

Anal. Chem.

84

, 4637–4646 (2012). (17) C. Atmanene, E. Wagner-Rousset, M. Malissard, B. Chol, A. Robert, N. Corvaia, A. Van Dorsselaer, A. Beck, and S. Sanglier-Cianferani,

Anal. Chem.

81

, 6364–6373 (2009). (18) A.B. Kanu, P. Dwivedi, M. Tam, L. Matz, and H.H. Hill,

J. Mass Spectrom.

43

, 1–22 (2008). (19) P. Olivova, W. Chen, A.B. Chakraborty, and J.C. Gebler,

Rapid Commun. Mass Spectrom.

22

, 29–40 (2008). (20) V. Katta and B.T. Chait,

J. Amer. Chem. Soc.

115

, 6317–6321 (1993). (21) K.D. Rand, M. Zehl, and T.J.D. Jorgensen,

Acc. Chem. Res.

47

, 3018–3027 (2014). (22) D. Houde, S.A. Berkowitz, and J.R. Engen,

J. Pharm. Sci.

100

, 2071–2086 (2011). (23) A. Sinz,

Mass Spectrom. Rev.

25

, 663–682 (2006). (24) A. Sinz,

J. Mass Spectrom.

38

, 1225–1237 (2003). (25) A. Leitner, T. Walzthoeni, and R. Aebersold,

Nature Protocols

9

, 120–137 (2014). (26) C. Bich, S. Maedler, K. Chiesa, F. DeGiacomo, N. Bogliotti, and R. Zenobi,

Anal. Chem.

82

, 172–179 (2010).  

Christian Huber

was educated as an analytical chemist from 1985 to 1993 at the University of Innsbruck, Austria. Following a lecturing qualification at the University of Innsbruck in 1997, he held the chair of analytical chemistry position at Saarland University in Germany from 2002 to 2008. In 2008, he was made a professor of chemistry for biosciences at the Department of Molecular Biology of the University of Salzburg, Austria. His research interests include bioanalytical chemistry and proteome and metabolome analysis, as well as in-depth characterization of therapeutic proteins.


Related to this article

Can Tinier Particles Make Better LC Columns?
LCGC International spoke to participants in a study using theoretical modeling to explore how much better chromatography columns packed with ultra-small (sub-1 micrometer) particles could perform under current and future high-pressure instruments, while weighing the practical challenges that have so far kept them from widespread use.