Analysis of Saccharides by Hydrophilic Interaction Chromatography (HILIC) Using TSKgel NH2-100 Columns

Article

The Application Notebook

The Application NotebookThe Application Notebook-02-01-2012
Volume 0
Issue 0

Saccharides are fundamental substances that express various bioactivities and may exist independently or form complexes with proteins or lipids.

Saccharides are fundamental substances that express various bioactivities and may exist independently or form complexes with proteins or lipids. They can be classified into monosaccharides, disaccharides, oligosaccharides, polysaccharides, etc., based upon the degrees of polymerization and condensation. Sugar alcohols are a class of polyols. A polyol is an alcohol containing multiple hydroxyl groups. Sugar alcohols are commonly added to foods since they are of lower caloric content than the corresponding sugars. The analysis of saccharides and sugar alcohols provides valuable information for the medical research, food industries and regulatory agencies.

In the past, various analytical techniques have been used to analyze saccharides, including all modes of high performance liquid chromatography (HPLC). Normal phase chromatography, in tandem with a differential refractometer as a detector, has long been used for the analysis of saccharides, as it provides good selectivity with relatively short analysis times. Hydrophilic interaction liquid chromatography (HILIC) selectively retains saccharides and polyhydric alcohols, such as sugar alcohols, while most of the substances with low polarity, as well as monohydric alcohols, elute in or very close to the void volume of the column.

We report the successful analysis of different kinds of saccharides and sugar alcohols using a TSKgel NH2-100 HILIC HPLC column. TSKgel NH2-100 columns are packed with 3 µm silica particles containing 100-Å pores. A novel bonding strategy was adopted to improve chemical stability of the bonded phase. First the silica is reacted with a trimethylsilane endcapping reagent at a low stoichiometric ratio, before reacting residual and accessible silanol groups with trifunctional alkylaminosilane reagent. The resulting bonded phase provides a better safeguard against hydrolysis of the underlying silica. TSKgel NH2-100 columns are unique in that the bonded phase ligand not only, as expected, has a terminal primary amino group, but that the spacer also incorporates secondary as well as tertiary amino groups (see Figure 1). TSKgel NH2-100 columns are well suited for the analysis of all types of hydrophilic compounds, including saccharides and sugar alcohols.

Figure 1: Structure of TSKgel NH2-100.

All the standards and samples were filtered through a 0.45 µm PVDF filter (Gelman) before injecting onto the column. High purity chemicals and HPLC grade solvents were used for the preparation of stock standards, samples, and mobile phases.

Results and Discussion

The symmetry and efficiency of the eluted sucrose standard, a disaccharide, from the TSKgel NH2-100 column is shown in Figure 2.

Figure 2: Analysis of sucrose standard.

To determine the suitability of the system and method, three consecutive injections of sucrose were loaded onto the TSKgel NH2-100 column. The study yielded a very consistent result across all peak parameters (see Table II). This column shows very high reproducibility for the analysis of sucrose.

Table I: Preparation of standards

Table II: System suitability results - sucrose analysis

Mannitol, a sugar alcohol, was successfully analyzed on a TSKgel NH2-100 column within 5 min (Figure 3).

Figure 3: Analysis of mannitol standard.

As in the case of sucrose, the analysis of mannitol could be reproduced with a high degree of consistency (Table III).

Conclusions

This study shows that a TSKgel NH2-100 column is suitable for the analysis of sucrose, a disaccharide, and mannitol, a sugar alcohol. Both standards separated on this column with good symmetry and efficiency. System suitability studies show that the analyses of sucrose and mannitol can be reproduced with very low %RSD in peak parameters using the TSKgel NH2-100 column.

Tosoh Bioscience and TSKgel are registered trademarks of Tosoh Corporation.

Table III: System suitability results - mannitol analysis

Tosoh Bioscience LLC

3604 Horizon Drive, Suite 100, King of Prussia, PA 19406

tel. (484) 805-1219; fax (610) 272-3028

Website: www.tosohbioscience.com

Related Videos
Related Content