Multidimensional liquid chromatography strategies are the most widely used method for increasing the number of spatially resolved components and reducing stress on mass spectrometric detection. However, the stress placed on a secondary dimension in a comprehensive on-line methodology is very high. An increasingly attractive approach is the coupling of high performance liquid chromatography (HPLC) with ion mobility spectrometry hyphenated to mass spectrometry (IMS-MS). Tim Causon and Stephan Hann of the University of Natural Resources and Life Sciences in Vienna, Austria, spoke to The Column about their work evaluating this approach and exploring its possibilities for metabolomics.
Photo Credit: Antonio M. Rosario/Getty Images
Multidimensional liquid chromatography strategies are the most widely used method for increasing the number of spatially resolved components and reducing stress on mass spectrometric detection. However, the stress placed on a secondary dimension in a comprehensive on-line methodology is very high. An increasingly attractive approach is the coupling of high performance liquid chromatography (HPLC) with ion mobility spectrometry hyphenated to mass spectrometry (IMS-MS). Tim Causon and Stephan Hann of the University of Natural Resources and Life Sciences in Vienna, Austria, spoke to The Column about their work evaluating this approach and exploring its possibilities for metabolomics. - Interview by Lewis Botcherby
Q. Your research explored the potential of high performance liquid chromatography (HPLC) with ion mobility spectrometry hyphenated to mass spectrometry (IMS-MS) for nonâtargeted analysis strategies. Why has the combination of HPLC and IMSâMS garnered attention in this area?
Tim Causon: There are already a number of interesting “flavours” of IMS-MS instruments around that rely on different IMS mechanisms, some of which have been available already for a few years on the market. Some of these approaches are already being used in combination with HPLC and there are plenty of options out there. Drift-tube (DT-) IMSâMS has only recently become commercially available for routine coupling with HPLC and I think that this low-field IMS approach has some quite attractive properties for adding on to conventional HPLC in combination with high resolution MS (HRMS). In particular, the ability to accurately determine a collisional cross section (CCS) value under low-field conditions via the low-field equation (MasonâSchamp) is particularly interesting as a physical property that we can eventually use in databases.
Stephan Hann: The introduction of a technique allowing the measurement of a “new” physical property of a molecule and its integration into a LC–MS-based workflow is highly interesting for analytical chemists and follow up interdisciplinary research. Of course, the new drift time dimension leads to an increased selectivity in many cases, but there are also other benefits that can be exploited, especially in non-targeted analysis; for example the fact that drift-tube IMS is a generic approach that works for all ions entering the drift tube. Additionally, it allows binning and feature alignment algorithms to make use of drift spectra (in addition to chromatographic signals) for recognition and combination of ion signals as molecular features belonging to one compound with more confidence. This will significantly increase the quality and reliability of compound detection within non-targeted workflows.
Q. What does this approach offer against other column-based multidimensional strategies for liquid chromatography?
TC: I am a big fan of multidimensional HPLC in combination with MS and I think it is going to continue to be an important topic for many applications, which need that really high degree of chemical separation prior to ionization such as proteomics. I don’t think that using drift-tube IMS can be a replacement in all cases because the separation occurs after the ionization interface, but there are other aspects to consider for this type of multidimensional approach, particularly for identity confirmation across different samples.
SH: On the practical side, while columnâbased HPLC×HPLC is starting to become realistic for users with commercial instrumentation, method development, data assessment, and longâterm robustness are going to remain challenging for users. HPLC×HPLC and parallel column approaches1 are beneficial if oneâdimensional (1D) chromatography suffers from matrix problems, low selectivity, or low retention factors. The latter drawback cannot be compensated by IMS, but we already know that driftâtube IMS can improve signal-to-noise ratios in highâmatrix samples and enable the separation of compounds, which cannot be distinguished in the LC-domain, or by accurate mass measurement. In addition to retention time, drift-tube IMS delivers a new “qualifier”, specifically drift time (or CCS), with a precision below 0.5% - this is often better than the precisions we achieve for retention times in two-dimensional (2D) chromatography. Furthermore, HPLC×IMSâMS doesn’t require any additional analysis time or solvent for the analysis - we can just switch the IMS functionality on or off as we require.
Q. Your research approached HPLC–IMS-MS theoretically. What were your major theoretical considerations and findings?
TC: We rely a great deal on the combination of HPLC and high resolution MS in our laboratory and with the acquisition of a new drift-tube IMS-MS instrument, we wanted to explore how this would help our metabolomics workflows in particular. I originally trained in the area of HPLC, so the idea was to explore these concepts using the language of chromatographers in a theoretical framework. What we found was that sample peak capacity is not so difficult to describe in the ideal case with values of 10–30 possible in most cases, and it is actually rather like describing isocratic HPLC in a simplistic sense. Secondly, we focused on determining how much difference in CCS values would be required to identify or quantify isomeric compounds that are coâeluting.2 If experimental parameters that influence IMS resolution outside of the drift tube are also included (as shown by work from Vanderbilt University),3 it is a nice way to estimate whether adding drift-tube IMS is really going to be able to solve our analytical goals in a given case. Eventually, we mustn’t forget that shapeâtoâcharge (IMS) and size-to-charge (MS) display only a limited orthogonality, so it is really important to have a good handle on what we can and cannot resolve with this combination before we invest our time in method development.
Q. Did your research uncover theoretical limitations to the combination of HPLC and IMS (-MS)?
TC: The understanding of peak capacity is a great place to start a conversation about what drift-tube ion mobility can add as an additional on-line separation, but it does not describe the full situation for this combination. Firstly, we know that a high-resolution mass spectrometer can resolve a lot of the separation problems that we are facing, which can make IMS separation seem not so valuable at first glance. Secondly, we have to keep in mind that this combination will not help with ion suppression issues, which is quite different to the situation with HPLC×HPLC–MS.
SH: As usual, a new technique faces limitations and challenges. As we are one of the first groups transferring metabolomic methods from LC–quadrupole time-of-flight (QTOF)-MS systems to LC×IMS–QTOF-MS, we encounter intrinsic problems of small molecules, for example space charge effects leading to losses during the trapping process. An additional challenge is more related to data processing and concerns the separation of certain molecular features, such as doubly charged ions or dimers from the original ions (singly charged isotopologues) in the drift tube - the re-alignment of these features is desirable and needs fitâforâpurpose software algorithms.
Q. What could be done to address these challenges?
SH: Together with other IMS users we see that trapping for a shorter time prior to the IMS separation but more often (so-called multiplexing) can solve such problems, especially in the low molecular mass region, enhancing signal-to-noise ratios and metabolite coverage. Evidently, the interlaced signals obtained by multiplexing are complex and the de-multiplexing needs more processor time.
TC: While the separation of IMS does not always resolve separation problems that the combination of HPLC–MS cannot, it helps make up for chromatographic variability especially from sample matrix issues. The ability to extract a CCS value with high precision really gives an extra dimension of confidence in feature alignment across samples. Now we have retention time, driftâtime/CCS, and a filtered high-resolution mass spectrum associated with each extracted feature. We need to standardize some details on how we annotate all of this information, but I think that we are already on the way to making the best use of this information.
The precision of the drift time and, ultimately, the collisional cross-section value is a parameter that really starts to help eliminate false positives. We need to work hard to look at reproducibility across instruments and figure out the best way to add drift information to HPLC–MS databases, but I believe that these are things that can be achieved.
Q. You extended your theoretical thinking towards an even more powerful approach of HPLC×HPLC×IMS-MS. How viable do you think this approach could be and what would be the major challenges if it were to be considered for use?
TC: More separation before the MS is definitely something that should be considered for extremely complex and dirty samples as long as we can manage to develop tools to handle the data extraction, assist with interpretation, and eventually make a workflow that is accessible for users. Improved preparation of the sample for the IMS-MS, whether by selective sample preparation methods or additional separation steps, will continue to be important development areas for non-targeted methods and trace analysis in complex samples. What drift-tube IMS can really offer is the addition of simple but highly repeatable drift spectrum information that we can use to describe and confirm the presence of a compound.
Q. Your research hinted towards future work revolving around using different drift gases for IMS-MS experiments. What are your thoughts on this area of research and are there any plans to investigate them further?
SH: I think that helium and nitrogen are going to remain the drift gases of choice for most users for large and small molecules, respectively. Helium seems to be the best choice for studying large molecules and matching up with computational predictions of drift behaviour, which has been a big topic in the ion mobility community for some time. In the small molecule world there are some very interesting works going back a few years covering different drift gases such as argon, carbon dioxide, and sulphur hexafluoride for drift-tube IMS. The group of Prof. Herbert Hill Jr. at Washington State University, for example, have contributed a number of interesting works in this area.4,5 A recent publication seemed to indicate that the selectivity differences obtainable from using alternative gases is not going to offer too much improvement for LC with DT-IMS-MS,5 but there are also other options including dopants that could be interesting for particular applications. Innovation can always be surprising, so I am looking forward to see what happens in this area.
Q. What are you working on next?
TC: Although fast screening with flowâinjection and IMS-MS alone is a fantastic tool for high-throughput workflows, I believe that separation methods will remain essential for getting the most reliable information out for assessment and interpretation with IMS-MS. With that in mind, I can see opportunities for developing really robust non-targeted workflows in metabolomics. The possibility to identify isomeric and isobaric species is also going to be a route to elucidate new biological information and there is plenty of unexplored space there.
SH: Currently, we are implementing DT-IMS into our workflows. This implies that we attempt to validate the results obtained for the “new measurand”, specifically CCS, in our targeted and non-targeted methods. We have determined long- and short-term precisions, but the trueness of the CCS values is, of course, unknown. This is why we are currently conducting an inter-laboratory study together with Vanderbilt University and Pacific Northwest National Laboratory in the USA to determine method reproducibility at different laboratories and with different instruments. In addition to our non-targeted work in biotechnological projects, we want to apply the new technique in rhizosphere research to identify important compounds and elucidate the chemical processes taking place at the soil–root interface.
References
Tim Causon is a Senior Scientist in the group of Instrumental Analytical Chemistry and Metabolomics at the Department of Chemistry of the University of Natural Resources and Life Sciences (BOKU - Vienna). His current research interests centre on mass spectrometry and separation science, including the metabolomic study of cell factories, fundamental analytical studies, and application developments for both targeted and non-targeted workflows.
Stephan Hann heads the group of Instrumental Analytical Chemistry and Metabolomics at BOKU - Vienna. The research of his group aims at the development and application of analytical methods in mass spectrometry-based metabolomics and ultra-trace analysis applied to a broad range of research topics.
Analysis of Greenhouse Gases by Gas Chromatography
May 15th 2024This application note demonstrates the use of SCION's 8500GC system for the analysis of key greenhouse gases—carbon dioxide, methane, and nitrous oxide—in a single atmospheric air matrix. Highlighting the system's excellent sensitivity and repeatability, this method is crucial for understanding emission sources and combating climate change.
Analysis of DEHP in Drinking Water by HPLC-DAD
May 15th 2024This application note outlines a method for detecting low levels of 1,4-Dioxane, a likely carcinogenic compound, in drinking water using SCION GC-MS technology. It details the procedure for analyzing water samples spiked with 1,4-Dioxane, achieving excellent sensitivity and low detection limits, demonstrating the method's effectiveness for environmental monitoring.