Dionex Application note
Kanamycin and amikacin are aminoglycoside antibiotics used to treat serious bacterial infections. Amikacin is used for infections resistant to other aminoglycosides because it is less susceptible to enzymatic reactions. Purified kanamycin from Streptomyces kanamyceticus is mainly kanamycin A, from which amikacin is synthesized by acylation of the kanamycin A amino group with L-(-)-γ-amino-α-hydroxybutyric acid (L-HABA). Kanamycin A and L-HABA are, therefore, expected impurities in amikacin synthesis.
These antibiotics must meet specified purity criteria before clinical use. The United States Pharmacopeia (USP) monographs for the assay of kanamycin and amikacin use high-performance, anion-exchange chromatography with pulsed amperometric detection (HPAE-PAD) (1,2). Here we evaluate the USP assays using disposable gold on polytetrafluoroethylene (PTFE) working electrodes with a 4-potential waveform suitable for use with these electrodes. Disposable electrodes are more convenient than conventional electrodes because they are easier to install and do not require time consuming electrode polishing. Compared to other disposable gold electrodes, the gold on PTFE electrodes have longer lifetimes and can operate at the hydroxide concentration used in the USP assays.
A Dionex ICS-3000 system and Chromeleon Chromatography Data System software were used in this study. Kanamycin and amikacin (20 µL) were separated using a CarboPac MA1 (USP L47) column set with 115 mM sodium hydroxide at 0.5 mL/min.
Figure 1 shows the <10 min separation of kanamycin and amikacin. Peak resolution was >4, (exceeding the USP requirement of 3); asymmetry for both antibiotics was 1.1 (USP requirement of <2). Retention time RSDs were 0.16 for kanamycin and 0.07 for amikacin for nine replicate injections (USP requirement of <0.3). Intra- and between-day peak area precisions (RSDs) were 0.99 and 1.3 for kanamycin and 1.2 and 2.3 for amikacin. These values suggest that this method effectively assays these antibiotics without column regeneration, and that using a disposable electrode with its associated waveform meets the current USP assay requirements.
Figure 1: Typical chromatograms of: (a) resolution solution (kanamycin 0.008 mg/mL and amikacin 0.02 mg/mL); (b) commercial kanamycin A sulphate sample; (c) commercial amikacin sample. Peaks: 1. kanamycin (8 µg/mL), 2. amikacin (20 µg/mL).
To demonstrate method capability for stability assays, these antibiotics were studied after exposure to elevated temperatures under acidic or basic conditions. For both antibiotics, most of the degradation products eluted within 10 min. An unidentified late-eluting peak at ~21 min is similar to a late-eluting peak reported for degraded streptomycin (3), which could interfere with quantification if a shorter run time is used. Under basic conditions, amikacin loses its acetylated group resulting in a kanamycin-like molecule.
This method matches and exceeds USP requirements, achieves good sensitivity and has high sample throughput. Additionally, using disposable electrodes provide shorter equilibrium time and better electrode-to-electrode reproducibility.
1. United States Pharmacopeia (USP), Kanamycin Sulfate, USP34-NF29, 3244.
2. United States Pharmacopeia (USP), Amikacin Sulfate, USP34-NF29, 1846.
3. Dionex Corporation, Application Note 181; Sunnyvale, California, USA
CarboPac, Chromeleon, and UltiMate are registered trademarks of Dionex Corporation.
Dionex Corporation
1228 Titan Way, PO Box 3603, Sunnyvale Calfornia 94088, USA
tel: +1 408 737 0700 fax: +1 408 730 9403
Website: www.dionex.com
Monitoring 5-Fluorouracil Levels in Dried Blood Spots with UHPLC-MS/MS
May 3rd 2024Researchers have utilized ultra-high performance liquid chromatography-tandem mass spectrometry (UH PLC-MS/MS) to monitor 5-Fluorouracil, a medication used to treat various cancers, in dried blood spots. This process offers a simpler, more efficient method than those in place involving the monitoring of plasma.
Metabolomics Meets Machine Learning: The Future of Nature-Inspired Drug Discovery
March 29th 2024Researchers at Enveda Biosciences are transforming nature into new medicines with their artificial intelligence (AI)-enabled industrial-scale drug discovery platform incorporating tandem mass spectrometry (MS/MS).
Element Launches New Pharmaceutical Testing Service
March 29th 2024Element Materials Technology introduced its Rapid Response Pharmaceutical Testing Service, providing expedited testing and focusing on various requirements including active ingredients and final drug products, extractables and leachables, degradation production, and elemental impurities.