Many "minor" variables in GC–MS analysis can be optimized to improve selectivity or sensitivity.
There are many "minor" variables in gas chromatography–mass spectrometry (GC–MS) analysis that can be optimized to improve selectivity or sensitivity, and knowledge of these is important in getting the most from your analytical methods. What follows is a brief summary of some of the simpler ways to improve your GC–MS analyses — go to ChromAcademy.com for a more detailed description.
Column bleed should be avoided in GC–MS because it significantly lowers detection limits and adds to the spectral noise when qualitative analysis is required. Ensure you have high-capacity oxygen and hydrocarbon traps fitted to the carrier-gas line, use "MS" designated (low-bleed) columns where possible, and condition columns properly before installation into the detector — this will mean passing carrier gas through the column at room temperature for at least 10 min before thermally cycling the column, which will usually not require more than 30 min at the isothermal temperature limit. To check for column bleed one would investigate a spectrum of the baseline of a total ion chromatogram (TIC) and note ions at 207, 267, and 281 — these ions indicate the presence of siloxane column bleed products.
Ensure that Vespel (Dupont)/graphite ferrules are used to prevent oxygen permeation, and avoid wider bore or short columns that require very low carrier gas head pressures because the carrier flow may not be reproducible with vacuum outlet conditions. Always work within the specified tolerances of the mass spectrometer pumping equipment to maintain a good level of vacuum within the instrument and improve sensitivity (this will typically be a 2–4 mL/min flow of carrier gas depending upon instrument design). The quality of the vacuum (and hence instrument sensitivity) will be directly influenced by the quality of the roughing pump oil, which should be replaced and "ballasted" to expel impurities on a weekly basis to optimize instrument sensitivity.
Most systems use perfluorotributylamine (PFTBA) as a tuning compound to optimize the ion source settings and mass analyzer calibration. This compound is suitable because it is volatile at room temperature under vacuum, contains no mass deficit due to the lack of hydrogen atoms, and fragments reliably to give ions over a wide mass range. Most manufacturers will use ions at m/z 69, 219, and 502 to perform the sensitivity optimization and mass axis calibration of the instrument; although these ions represent a wide range of masses and relative abundance, they may not be suitable for "target tuning" the equipment for high sensitivity at masses relevant to your particular analysis. The relative "normative" benefits of autotuning are outside the scope of this piece; however, to obtain the maximum performance from GC–MS equipment, one needs to appreciate the "manual" tuning benefits.
Many electrostatic components of a modern ion source will be "tunable" by monitoring ion abundance against applied voltage for a range of ions, which are typically user selectable. Selecting the voltage corresponding to optimum ion abundance for an ion whose mass is close to your target ions for the repeller or pusher electrode, for example, can often lead to an increase in instrument sensitivity in selected ion monitoring (SIM) mode. Other ion source components may be optimized using such voltage "ramping" techniques, and one may need to iterate ion source voltages for beam formation, focusing, and acceleration to find the optimum combination of settings.
The energy of the ionizing electrons is of great importance and when using spectral libraries for qualitative work, the electron energy should be set at 70 eV. For all other purposes, the use of 70 eV is purely arbitrary and it may be that altering the electron energy will aid optimization of a method. Reducing electron energy can lead to the promotion of a more prominent molecular ion, although one should note that the absolute abundance of all ions will decrease exponentially. Increasing the electron energy may improve the abundance of target ions in SIM analysis and hence improve the sensitivity of a method. These parameters should be explored as part of the method optimization where better qualitative or quantitative performance is required. Note that altering electron energy may also affect relative abundances of target and qualifier ions.
The mass analyzing device will almost certainly be capable of tuning to a higher specification than achieved by the autotune routine. For example, most quadrupole mass analyzing devices can be tuned for resolution and sensitivity by altering the magnitude of the applied voltages (DC and AC). An increase in the applied DC voltage will increase spectral resolution, but it will reduce sensitivity and affect all masses to the same extent. An increase in the gain (slope) of applied DC versus applied AC voltages will also reduce sensitivity and improve resolution, but it will affect higher masses to a greater extent. Of course, an increase in sensitivity and a reduction in resolution can be achieved by lowering the applied DC and the gain. These techniques are important approaches to optimizing a GC–MS analysis and more information should be sought from your manufacturer.
For scan data acquisition, keep the mass range as narrow as possible and the number of sampling iterations at each mass as high as possible to obtain good quality spectra; consider only spectra that have been properly background subtracted and are an average of all spectra across the peak apex from half height. For SIM analysis, optimize the ion dwell or recording time and keep the numbers of ions per group at a minimum to ensure optimum sensitivity.
Ensure that the transfer device, ion source, and mass analyzer are at the correct temperature. The optimum temperatures will vary between manufacturer and between electron ionization and chemical ionization modes.
Reflecting on the Influence of the Current State of Sample Preparation on GC, Part 2: Techniques
May 9th 2024In this installment, we examine trends in the use of sample preparation techniques through the lens of instrumental analysis by gas chromatography (GC) and GC–mass spectrometry (GC–MS).
TD-GC–MS and IDMS Sample Prep for CRM to Quantify Decabromodiphenyl Ether in Polystyrene Matrix
April 26th 2024At issue in this study was the certified value of decabromodiphenyl ether (BDE 209) in a polystyrene matrix CRM relative to its regulated value in the EU Restriction of Hazardous Substances Directive.