Antioxidants protect cells from the damaging effects of free radicals and offer numerous benefits for human health. Many phenolic compounds found in plants, as well as some vitamins, exhibit antioxidant activity. Several colorimetric assays exist to measure the total antioxidant capacity, typically expressed as Trolox or Gallic acid equivalents.
Antioxidants protect cells from the damaging effects of free radicals and offer numerous benefits for human health. Many phenolic compounds found in plants, as well as some vitamins, exhibit antioxidant activity. Several colorimetric assays exist to measure the total antioxidant capacity, typically expressed as Trolox or Gallic acid equivalents.
An increased interest in antioxidants has created a demand for methods that are not only capable of determining the total antioxidant activity of the sample, but also are able to identify and quantify individual compounds known for their biological benefits. Our method abstract demonstrates that the well-known colorimetric reagents, such as Folin-Ciocalteu and 2,2'-azinobis (3-ethylbenzothiazoline)-6-sulfonate (ABTS), can be successfully used for analysis of antioxidants in foods and dietary supplements by HPLC with post-column derivatization.
Folin-Ciocalteu Reagent (FCR) reacts with reducing compounds to form a chromophore that can be detected using UV-vis detection. Phenolic and non-phenolic antioxidants, including Vitamin C, Vitamin ± and Folic acid, can be analyzed using this reagent with high sensitivity. Many antioxidants can be easily detected in solution at levels as low as 1 µg/mL.
Figure 1: Mixture of 10 common antioxidants, analyzed using Folin-Ciocalteu reagent.
The ABTS reagent is used in the Trolox-equivalent antioxidant capacity (TEAC) assay. The reaction is based on suppression of the absorbance of radical cations by antioxidants, and can be detected by UV-vis detection as negative peaks on the chromatograms. This indirect detection method allows for analyzing common antioxidants at levels between 40 and 200 µg/mL.
Figure 2: Fresh raspberries extract, analyzed using Folin-Ciocalteu reagent.
Using both reagents, it is possible both to identify and quantify specific target compounds, as well as to calculate the total antioxidant capacity of the sample by comparing the total area of all the peaks on the chromatogram with Trolox or Gallic Acid calibration curves.
Figure 3: Hot water extract of green tea, analyzed using Folin-Ciocalteu reagent.
Sample preparation
Fresh fruit: Homogenize 10 g of fresh fruit with 20 mL of water/methanol (50:50) extraction solution for 5 min. Centrifuge and filter through 0.45 µm Nylon filter.
Dried fruit: Mix 4 g of finely ground dried fruit with 20 mL of 100% Methanol. Shake for 1 h on a mechanical shaker. Centrifuge and filter through 0.45 µm Nylon filter.
Dietary supplements: Mix 100 mg of finely ground sample with 10 mL of 100% Methanol. Shake for 30 min on a mechanical shaker. Centrifuge and filter through 0.45 µm Nylon filter.
Dry tea leaves: Homogenize 1 g of dry tea leaves with 25 mL of water/methanol (50:50) extraction solution for 5 min. Centrifuge and filter through 0.45 µm Nylon filter. Alternatively, mix 1 g of dry tea leaves with 25 mL of hot water, stir on a hot plate for 5 min. Cool down, centrifuge and filter through 0.45 µm Nylon filter.
All extracts can be diluted if needed with water/methanol (50:50).
Figure 4: Extract of dietary supplement, analyzed using ABTS reagent.
Column: Reversed-phase C18 column, 4.6 x 150 mm
Temperature: 40 °C
Flow Rate: 1 mL/min
Mobile Phase: 4.8% Acetic acid in water, Methanol
Injection Volume: 10–50 uL
Table I: HPLC gradient
Post-column System: Pinnacle PCX or Vector PCX
Reactor Volume: 1.4 mL
Reactor Temperature: 130 °C
Reagents: Option 1: 40% of Folin-Ciocalteu reagent in water
Option 2: 15% of ABTS stock in Methanol.
ABTS stock solution: 7 mmole/L of 2,2'- azinobis(3-ethylbenzothiazoline)-6-sulfonate is dissolved in a 2.45 mmole/L solution of Sodium Persulphate in water. Stock solution should be stored in the dark for at least 12 h before making a working reagent.
Flow Rate: 0.1 mL/min
Detection: UV-vis 635 nm
Table II: Calculated total antioxidant capacity of selected samples
(1) B. Kusznierewicz, A. Piasek, A. Bartoszek, and J. Namiesnik, "The Optimization of Analytical Parameters for Routine Profiling of Antioxidants in Complex Mixtures by HPLC Couples with Post-column Derivatization," Phytochem. Anal. 22, 392–402 (2011).
Pickering Laboratories
1280 Space Park Way, Mountain View, California 94043
tel. (800) 654-3330, fax (408) 694-6700
Website: www.pickeringlabs.com
Analytical Challenges in Measuring Migration from Food Contact Materials
November 2nd 2015Food contact materials contain low molecular weight additives and processing aids which can migrate into foods leading to trace levels of contamination. Food safety is ensured through regulations, comprising compositional controls and migration limits, which present a significant analytical challenge to the food industry to ensure compliance and demonstrate due diligence. Of the various analytical approaches, LC-MS/MS has proved to be an essential tool in monitoring migration of target compounds into foods, and more sophisticated approaches such as LC-high resolution MS (Orbitrap) are being increasingly used for untargeted analysis to monitor non-intentionally added substances. This podcast will provide an overview to this area, illustrated with various applications showing current approaches being employed.
Investigating PFAS in Plastic Food Storage Bags Using LC–MS/MS
May 8th 2024Yelena Sapozhnikova from the United States Department of Agriculture spoke to The Column about her innovative research investigating PFAS in plastic food storage bags using targeted and non-targeted liquid chromatography–tandem mass spectrometry.