In this study, the scientists developed a cost-effective and time-efficient laboratory method designed to simulate emissions from the mineral wool curing process using direct thermal extraction gas chromatography–mass spectrometry (DTE-GC–MS). Factors affecting VOC emissions, such as flow, residence time, and the atmosphere during the thermal desorption step, were investigated to identify the dominant factor influencing the chemical composition of emissions. Industrial processes could be better replicated. Additionally, a laboratory method was created to mimic factory-produced materials.
Analyzing the results from the emission simulation studies indicated that only the atmosphere proved to be a statistically significant parameter. The negligible effects of the flow rate, residence time, and their interactions can be attributed to the design of the DTE-GC–MS setup utilized in this research. Namely, the small sample size of uncured wool, introduced in glass tubes with approximately 1.9 mL volume, allowed for efficient heat transfer between the hot gas and the sample. Studied flow rates also varied from 7.8 to 60 tube volumes per minute, with sample temperatures closely following the thermal desorption unit’s temperature program, even at high flow rates.
The results showed that only the type of atmosphere in a thermal desorber during thermal desorption significantly affects emission compositions, with residence time and flow rates having statistically negligible impacts. Further, a short (6 min) method with a low flow rate (15 mL/min) was deemed sufficient to simulate curing-oven emissions effectively.
Strong correlations between laboratory and industrial emissions confirmed the method’s reliability for studying VOC emissions from real-world emissions. This allows for more accurate assessments of the environmental behavior of novel lignin-based binders while supporting the development of low-emission, sustainable materials. Using lignin-based binders can potentially reduce the environmental footprint compared to traditional PUF binders, contributing to more sustainable manufacturing practices. By establishing reliable methods for early-stage emission assessment, this study can facilitate the adoption of greener binder technologies aimed at minimizing unintended emissions.
References
(1) Homepage. Rockwool 2025. https://www.rockwool.com/north-america/ (accessed 2025-6-18)
(2) Salionov, D.; Nikolic, M.; Glasius, M. Emission Simulation Using Direct Thermal Extraction Gas Chromatography-Mass Spectrometry: Case Study with Curing of Lignin-Based Binder. J. Chromatogr. A 2025, 1753, 465970. DOI: 10.1016/j.chroma.2025.465970