GC–FID Tests Alcohol Stability in Autopsy Samples

Author(s)John Chasse
Listen
0:00 / 0:00

Key Takeaways

  • Paired blood–vitreous ethanol testing mitigates postmortem confounding from redistribution and endogenous/microbial ethanol production, supporting more defensible intoxication assessment in diverse medicolegal deaths.
  • Vitreous humor ethanol trails blood by ~1–2 hours, enabling approximate reconstruction of preterminal blood alcohol and contextualizing perimortem drinking kinetics.
SHOW MORE

Gas chromatography with flame ionization detection (GC–FID) shows eye fluid alcohol stays stable for six months, unlike blood.

When someone dies, measuring how much alcohol was in their body can help investigators piece together what happened and figure out the cause of death. Alcohol levels in the blood tend to track closely with levels in the vitreous humor, the clear gel that fills the eyeball, so the two are often tested side by side.

Researchers in Serbia wanted to know whether these alcohol readings hold steady over time. They tested blood and eye fluid samples taken at autopsy, stored them properly, and then tested them again six months later. Their goal was to see whether the alcohol levels changed, whether the blood and eye fluid readings still matched up, and whether their testing method was reliable. To measure the alcohol, they used gas chromatography with flame ionization detection (GC–FID). Their findings were published in the journal Legal Medicine.¹

Why Test Both Blood and Eye Fluid for Alcohol After Death?

Alcohol is one of the most widely available mood-altering substances, and it has powerful effects on both the body and behavior. When someone dies, testing for alcohol in their body is a key step in understanding what happened and why they died. Alcohol is well known to play a role in many kinds of deaths, including poisonings, car crashes, suicides, and violent incidents.2,3

Measuring the alcohol in a person’s blood is still considered the best way to tell whether they were intoxicated when they died, but it isn’t foolproof. After death, the blood goes through chemical changes, and alcohol can shift from one part of the body to another. On top of that, bacteria, yeasts, and molds can actually produce new alcohol in the body after death. All of this must be kept in mind when interpreting the results.4

Alcohol levels in the blood are often found to be higher than those in the eye fluid, which may depend on how and under what circumstances the person died. This is partly because alcohol reaches the eye fluid more slowly: the eye fluid lags about one to two hours behind the blood. That delay is actually useful, because the alcohol level in the eye fluid can offer a glimpse of what the blood alcohol level was roughly one to two hours before death.5

How Stable Are Alcohol Levels in Stored Blood and Eye Fluid Samples?

The study ran from January 2023 to November 2024 and drew on autopsies performed at several different centers. In total, the researchers analyzed 248 samples, testing the alcohol levels in each one at four points: at the time of autopsy, and again after one, three, and six months in storage.1

At every one of these checkpoints, the alcohol levels in blood and eye fluid lined up closely: when one was high, the other was too. The match was strongest at the one-month mark and stayed very strong throughout, dipping only slightly by the three- and six-month tests.1

The eye fluid turned out to be especially dependable. Its alcohol levels held steady over the full six months, with no meaningful change. The blood samples were a different story. Even though they were kept frozen at a constant –20 °C, their alcohol levels rose noticeably by the six-month mark. This suggests that eye fluid may be the more trustworthy sample when alcohol tests need to be repeated long after an autopsy.1

“Concentration of ethanol in the blood increased by 29% from moment of autopsy to six months,” write the authors of the paper.1 “ Validated method demonstrated precision, accuracy, specificity and sensitivity. Due to the statistically significant change in the concentration of ethanol in blood samples, it is not suitable for recalculating the concentration of ethanol in blood from vitreous humor, after storage for more than six months.”

Read More on Similar Topics
Using UHPLC-MS/MS to Help Pinpoint Time of Death

References

  1. Mihajlović, F.; Slović, Ž.; Todorović, M. et al. Forensic Assessment of Ethanol Toxicity and Validation of an Analytical Method for Ethanol Determination in Blood and Vitreous Humor After Autopsy. Leg Med (Tokyo) 2026, 87, 103048. DOI: 10.1016/j.legalmed.2026.103048
  2. Oshaug, K.; Kronstrand, R.; Kugelberg, F. C. et al. Frequency of Postmortem Ethanol Formation in Blood, Urine and Vitreous Humor - Improving Diagnostic Accuracy with the Use of Ethylsulphate and Putrefactive Alcohols. Forensic Sci Int. 2022, 331, 111152. DOI: DOI: 10.1016/j.forsciint.2021.111152
  3. Pigaiani, N.; Bertaso, A.; De Palo, E. F. et al. Vitreous Humor Endogenous Compounds Analysis for Post-mortem Forensic Investigation. Forensic Sci Int. 2020, 310, 110235. DOI: 10.1016/j.forsciint.2020.110235
  4. Zekan, P.; Ljubičić, N.; Blagaić, V. et al. Pharmacokinetic Analysis of Ethanol in a Human Study: New Modification of Mathematic Model. Toxics 2023, 11 (9), 793. DOI: 10.3390/toxics11090793
  5. Issa, S. Y.; Khattab, A. Ethanol Concentrations in Various Biological Specimens: Living and Postmortem Forensic Toxicology Analysis and Comprehensive Literature Review. J Forensic Leg Med. 2024, 106, 102737. DOI: 10.1016/j.jflm.2024.102737


Related to this article

Magnetic Aerogel Extracts Pesticides for LC–MS/MS Testing
Kazhal Khazaei of K.N. Toosi University of Technology (Iran) discusses her research team’s development of a magnetic copper-based metal-organic gel aerogel that captures tiny amounts of pesticides from plant-based milks for liquid chromatography.–tandem mass spectrometry (LC–MS/MS) testing with LCGC International.