A study in the Journal of Chromatography A used gas chromatography coupled to a time-of-flight mass spectrometer (GC–TOF-MS) to build an electron ionization mass spectra (EI-MS) database of more than 250 chemicals classified as either volatile or semi-volatile compounds (1). An additional, confirmatory layer of liquid chromatography–mass spectrometry (LC–MS) analysis was subsequently performed.
This report out of the Universidade de Santiago de Compostela, Spain, aimed to improve upon gas chromatography–mass spectrometry (GC–MS) detection considered the popular standard for analysis of “e-liquids” (1). The authors said both LC–MS and nuclear magnetic resonance (NMR) spectroscopy have been proposed for determination, but that only approximately 140 volatile compounds were previously identified.
Certain compounds produced by tobacco combustion in conventional cigarettes are not present in e-cigarettes, burnishing their reputation as a “safe” alternative (1). But the researchers list factors and symptoms indicating lung issues in consumers, from flavors that may cause oxidative stress to reactions between carbonyls and alcohols producing acetals, known respiratory irritants. Other addictive substances can be found in electronic smoking devices, such as nicotine and cannabinoids.
The research team sought a nontargeted approach, analyzing the e-liquids using two capillary columns with different polarities (1). The complementary LC–MS analysis was reserved for compounds identified at the highest of three levels of confidence, having satisfied three identification criteria. Study results showed while concentration ratios between delta-9-tetrahydrocannabinol (Δ9-THC) and cannabidiol (CBD) in e-liquids fell into a relatively narrow range (between 0.02% and 0.3%), the ratio of propylene glycol acetals to parent aldehydes ranged from 2%, for ethyl vanillin, to upwards of 80% for benzaldehyde.
(1) Cobo Golpe, M.; Ramil, M.; Rodríguez, I. Comprehensive Characterization of Volatile and Semi-Volatile Compounds in E-Liquids for Electronic Cigarette Using Gas Chromatography Accurate Mass Spectrometry. J. Chromatogr. A 2023, 1703, 464114. DOI: 10.1016/j.chroma.2023.464114
Carol Robinson Awarded 2024 Lifetime Achievement European Inventor Award
July 24th 2024Carol Robinson of the University of Oxford has received the European Inventor Award 2024 for Lifetime Achievement from the European Patent Office for her work bringing mass spectrometry to structural biology.
Microseparations and Analysis Covered at HPLC 2024
July 23rd 2024At HPLC 2024 in Denver, Colorado, various sessions will be led by experienced scientists, covering different aspects of chromatography techniques and maximizing their effectiveness. Here is some information about tomorrow’s sessions.
Automated LC Method Development: An HTC-18 Interview with Bob Pirok
July 23rd 2024At HTC-18 in Leuven, Executive Editor of LCGC International, Alasdair Matheson, spoke to this year’s winner of the 2024 HTC-18 Innovation Award, Bob Pirok from the University of Amsterdam, about his innovative research on automated LC method development.