LC–MS/MS Method Quantifies Psilocin in Brain Tissue

Author(s)John Chasse
Listen
0:00 / 0:00

Key Takeaways

  • Matrix-related ion suppression from phospholipids and salts is a major barrier to reproducible brain LC–MS/MS quantitation of psilocin and necessitates optimized chromatographic separation and instrument plumbing.
  • High sample dilution (1:10), reduced flow, and a shallower solvent gradient enabled baseline separation of psilocin/4-HIAA from phospholipid interferences without specialty cleanup sorbents.
SHOW MORE

Tuned liquid chromatography–tandem mass spectrometry (LC–MS/MS) separation beats brain matrix effects to quantify psilocin.

Psilocin is the chemical that psilocybin, the active ingredient in “magic mushrooms,” turns into in the body, and it is what causes the mushrooms’ mind-altering effects. The chemical is gaining attention as a possible treatment for depression and anxiety that haven’t responded to other therapies. To use it safely and wisely, scientists need to know exactly how much psilocin, and how much of its main breakdown product (4-HIAA), ends up in the brain, and how those amounts change over time.

Measuring small molecules in brain tissue is tricky, however. The usual tool for the job is liquid chromatography–tandem mass spectrometry (LC–MS/MS), but brain tissue is full of fatty molecules called phospholipids that can blur the instrument’s readings. Leftover salts can also weaken the signal. Both problems make results harder to reproduce.

Researchers at the University of Saskatchewan’s College of Pharmacy and Nutrition set out to solve these problems and build a new, fully tested LC–MS/MS method for measuring psilocin in different parts of the brain. They tackled the interference both in how the sample is separated and in how the instrument is set up. According to the team, no fully validated method had existed for measuring psilocin in the brain, the organ where it acts. They published the work in the journal Analytical Methods.¹

Why Do Scientists Need Better Ways to Measure Psilocin?

Researchers are increasingly testing psilocin as a possible treatment for major depression, anxiety, and post-traumatic stress disorder (PTSD).2-4 As research on psilocybin-based treatments grows, scientists need precise, reliable ways to measure psilocin in the body. That lets them track how the drug is absorbed, spread, and cleared, and check that it is safe.5

How Did the Team Get Around Brain Tissue Interference, and What Did the New Method Reveal?

Instead of using special cleanup materials to strip out the fatty phospholipids, the team solved the problem with the separation step itself. First, they diluted the brain samples, mixing one part sample with nine parts solvent. Second, they slowed the flow through the instrument and changed the solvent mix more gradually, so the drug and its breakdown product came out separately from the phospholipids. Third, they carefully timed a valve that sends the salty portion of each run to waste instead of into the detector.Thanks to these changes, the method could measure psilocin at extremely low levels, down to 0.05 nanograms per milliliter, with only one simple sample-prep step. It also passed the standard tests for picking out the right compound, giving accurate results, and giving consistent results, so it can reliably measure psilocin in mouse brain tissue.1

The researchers then used the method to compare two forms of the drug: regular psilocybin and a new formulation that combines psilocin and psilocybin in a single crystal (a “co-crystal”). Psilocin levels in the brain did not differ between male and female mice or between brain regions. Ten minutes after dosing, mice given regular psilocybin had more psilocin in their brains than mice given the co-crystal. The co-crystal seemed to keep brain psilocin levels steadier over time, but that difference was not large enough to be statistically meaningful. The breakdown product 4-HIAA was not found in the brain at all, which suggests it does not build up there and is cleared quickly.1

“Overall,” wrote the authors of the paper,1 “this method establishes a robust platform for preclinical investigations of psilocin pharmacokinetics, brain distribution, and metabolic fate.”

Read More on Similar Topics
Applications of Untargeted Chemical Analysis to Determine Psilocybin and Psilocin Content in Multiple Psilocybe cubensis Mushroom Strains Using LC–MS/MS: An Interview with Kevin Schug


References

  1. Khajavinia, A.; Reekie, J.; Michel, D. et al. Novel Quantification Method for Psilocin in Brain Tissues Using Liquid Chromatography-Tandem Mass Spectrometry. Anal Methods 2026.DOI: 10.1039/d6ay01614f
  2. García-Cabrerizo, R.; Beruete-Fresnillo, I.; Bergas-Cladera, P. et al. Psilocybin as a Fast-acting and Long-lasting Antidepressant for Adolescence: Proposing NeuroD1 as a Biomarker of its Long-term Plasticity. Biomed Pharmacother. 2025, 193, 118720. DOI: 10.1016/j.biopha.2025.118720
  3. Davis, A. K.; Barrett, F. S.; May, D. G. et al. Effects of Psilocybin-Assisted Therapy on Major Depressive Disorder: A Randomized Clinical Trial. JAMA Psychiatry 2021, 78 (5), 481-489. DOI: 10.1001/jamapsychiatry.2020.3285
  4. Chen, J.; Wang, Z.; Yong, C. Y. et al. Elucidating the Phase I Metabolism of Psilocin in vitro. Arch Toxicol. 2025, 99, 1085–1094. DOI: 10.1007/s00204-024-03952-7
  5. Straumann, I.; Holze, F.; Becker, A. M. et al. Safety Pharmacology of Acute Psilocybin Administration in Healthy Participants. Neurosci Appl. 2024, 3, 104060. DOI: 10.1016/j.nsa.2024.104060

Related to this article

Can Tinier Particles Make Better LC Columns?
LCGC International spoke to participants in a study using theoretical modeling to explore how much better chromatography columns packed with ultra-small (sub-1 micrometer) particles could perform under current and future high-pressure instruments, while weighing the practical challenges that have so far kept them from widespread use.