Can Tinier Particles Make Better LC Columns?
Key Takeaways
- Particle downsizing improves speed in different efficiency regimes, with slow-diffusing peptides/proteins benefiting where routine work targets ~10⁵ plates, while small molecules realize gains mainly at lower plate counts.
- Extreme-pressure LC is constrained by fixed-cycle pump flow fluctuations and impractical gradient-formation strategies, plus high-pressure injection, scarce rated fittings/unions, and heightened mechanical safety risks.
LCGC International spoke to participants in a study using theoretical modeling to explore how much better chromatography columns packed with ultra-small (sub-1 micrometer) particles could perform under current and future high-pressure instruments, while weighing the practical challenges that have so far kept them from widespread use.
For decades, one of the main way scientists have made liquid chromatography (LC) columns better is by shrinking the size of particles packed inside them. In the 1960s, those particles were about 50 μm across; today, the smallest commercial-available particles packed in commercial column are 1.3 μm. Smaller particles give the liquid shorter, more even flowpaths through the column, so mixtures get separated more quickly with shaper peaks. The catch is that smaller particles also make it harder to push liquid through, so the instruments need much higher pressure.
Even so, commercial particles have not gotten any smaller in nearly twenty years. Researchers have experimented with particles under 1 μm, but it's still rare because there are several stubborn problems. Today's instruments can only handle so much pressure, which limits how long these columns can be. The ultra-sharp peaks they produce are easily blurred by small imperfections elsewhere in the system, and it's simply very hard to pack such tiny particles into a stable, even bed.
A recent study conducted at Vrije Universiteit Brussel (VUB) used theoretical modeling to ask how well columns with these ultra-small particles could perform. The researchers looked at what is possible with current equipment and with future instruments that could reach roughly double today's pressure limits. They also compared a normal, randomly packed column with an ideal, perfectly ordered arrangement of particles, and they outline the main technical hurdles that still stand in the way.
LCGC International spoke to Hanrong Wen, Ken Broeckhoven, and Sebastiaan Eeltink of VUB, authors of the paper resulting from this study,1 about their work.
Your kinetic-plot analysis shows a "time gain factor larger than 2" specifically for slow-diffusing molecules like peptides and proteins with 500 nm particles. Why does particle size reduction benefit these large biomolecules more than small-molecule analytes, and what does this mean for the future of protein/peptide separations?
When we reduce the particle size, both small molecules and larger molecules can benefit from a similar magnitude of improvement in separation speed, but the improvement occurs in different efficiency regimes. For example, if we compare 750 nm particles with 1.5 μm particles, for peptides with a diffusion coefficient of around 10⁻¹⁰ m²/s, the time-gain factor can approach to 2 when the required efficiency is below 2 × 10⁵ plates. For small molecules, which typically have a higher diffusion coefficient of around 10⁻⁹ m²/s, a similar time gain is achieved at a lower efficiency, below 2 × 10⁴ plates.1
The reason is that the difference in diffusion coefficient leads to different linear velocities when the separation is performed at the same reduced velocity. When we operate at the same maximum pressure, a fast-diffusing analyte will be separated at a higher linear velocity, leading to the use of a shorter column, and therefore achieving a lower efficiency. In contrast, a slowly diffusing analyte requires a lower linear velocity, resulting in a longer separation time, but allows to use a longer column and thus maintaining a higher separation efficiency. This is why reducing particle size can be particularly valuable for larger molecules. Their effective time-gain region overlaps with the efficiency range that is commonly relevant in practice, roughly below 10⁵ theoretical plates. At the same time, the lower diffusion coefficients of these analytes mean that reducing the particle size can translate into a substantial reduction in separation time due to reduction of diffusion path length.
Looking toward the future of peptide and protein separations, smaller particles could certainly provide greater kinetic benefits, particularly when higher throughput is desired. However, we also need to consider the potential drawbacks. Smaller particles generate a higher shearing environment and introduce potential hydrodynamic or size-exclusion effects, which might deteriorate the separation of large biomolecules such as intact protein/long nucleotide. So, when it comes to determining the optimal particle size for large biomolecules, further experimental studies using submicron particle-packed columns will be essential.
You mention that at ≥3000 bar, applicable particle sizes could drop below 300 nm. What specific engineering barriers currently prevent commercial LC systems from reaching these pressures and which do you see being solved first?
Ultrahigh-pressure liquid chromatography (UHPLC) setups capable of operating at pressures above 3000 bar were already reported in the late 1990s. To the best of our knowledge, the highest-pressure limit reported in the literature is 5000 bar.2 Such extremely high pressures can be generated using either a syringe pump2 or a pneumatic-driven liquid pump.3 The major issue with these pumps is that they operate with a fixed duty cycle, which can cause a large fluctuation in flow rate/back pressure when the pumping cycle ends. We previously established a dual-pump system to reduce this fluctuation during continuous operation at 2000 bar.4 When a capillary column is applied at nanoflow rates, we can expect a much longer effective duty cycle. However, the flow fluctuation can still be significant, meaning that the separation would need to avoid the end of the pumping cycle.
Another major engineering challenge is how to implement these extreme-pressure pumping systems for gradient separations. Several approaches have been reported in the literature, including mixing the strong solvent into the weak-solvent reservoir and pre-loading the gradient using a conventional HPLC instrument.2,5,6 However, these approaches require relatively long preparation or re-equilibration times, which is not ideal for high-throughput separations. Furthermore, a pneumatic pump operates as a constant-pressure pumping system. As a result, the flow rate during a gradient separation will continuously change as the viscosity of the mobile phase changes. This makes it more difficult to maintain a constant linear velocity throughout the gradient.
There are also several other engineering challenges in designing an extreme-pressure LC system, including the large size of the pumping system, sample injection at such high pressures, the limited availability of commercial fittings and unions rated for extreme pressures, and, of course, safety.7 At 3000 bar, the mechanical failure of a column or fitting could potentially result in the column being ejected from the system. Overall, we would say that when designing a 3000-bar LC system, two issues should probably be addressed first: minimizing the influence of the pump duty cycle on flow stability and developing a more efficient strategy for introducing gradients. Solving these two problems would be particularly important if the goal is to achieve high-throughput separations under extreme-pressure conditions.
The paper distinguishes between randomly packed beds and "perfectly ordered" face-centered cubic (FCC)-packed nanoparticle columns, citing a potential 20-35x speed improvement. What is currently preventing FCC self-assembly from being realized in practice, and how close is the field to overcoming this?
Achieving long-range order in a periodic FCC structure inside a cylindrical tube is almost an impossible task. Because of the presence of the circular wall and the relatively fast packing process, it is extremely difficult to avoid small defects, such as colloidal crystal misalignment, missing particles, or the formation of voids/cracks.8 The Wirth group reported colloidal crystal columns with an FCC packing structure in the HPLC literature. However, their column lengths were mainly limited to around 2 cm.9,10 Only a few studies have reported longer columns with FCC structures,11 and many of these results have not yet been independently reproduced by other research groups. Therefore, we would expect that even a well-prepared FCC column under practical conditions would provide a smaller improvement in separation speed than our theoretical predictions suggest.
Achieving reproducible self-assembly-based column packing requires sophisticated control over the assembly process and a highly stable packing environment. In principle, these requirements should be achievable by applying knowledge from materials science and colloidal physics. However, developing a robust and reproducible packing protocol would likely require extensive trial and error. Another practical limitation is the packing time. Compared with conventional high-pressure slurry packing, self-assembly may require substantially longer packing times because the particles need to assemble under relatively undisturbed conditions. This could become an important bottleneck for the practical implementation of FCC-packed columns.
You calculate that extra-column dispersion tolerance for sub-1 µm columns falls to just 2-6 nL2—far below what current commercial systems can achieve. Can you walk us through why this tolerance shrinks so dramatically at these particle sizes, and what specific system components contribute most to exceeding it?
This extremely small tolerance for extra-column dispersion is mainly caused by the very high column efficiency per unit length, or, in other words, the smaller theoretical plate height. This improvement comes from two factors: the smaller particle size and the use of non-porous particles. The tolerance for extra-column band broadening is typically defined as 10% of the column-only band broadening.12 Therefore, when the column efficiency increases and the column-only band broadening becomes smaller, the allowable extra-column band broadening decreases accordingly. Most mainstream UHPLC instrumentation is designed primarily for conventional analytical HPLC columns with a 2.1 mm I.D., as well as microLC applications. These columns typically produce a column band broadening on the order of a few μL². Even with the best instrument configurations,13 the extra-column band broadening is usually around 1-2 μL², which is far above the tolerance required for capillary columns packed with submicron particles.
The major contributors to extra-column band broadening are the connections and tubing. According to Hara et al.,14 a poorly selected union or an improperly made fitting can easily introduce an extra-column band broadening of a few hundred nL². When using 20 μm i.d. tubing, the contribution from the tubing itself can be relatively small, which is sufficient for many microLC applications. However, if we consider the length of tubing required to make connections in a conventional high-performance LC system, the dispersion introduced by the tubing can easily exceed the very small tolerance of a submicron particle column. Therefore, for submicron particle columns, optimizing the connections and fittings and minimizing the tubing length will be essential for preserving the efficiency of the column.
You propose a column-in-valve modular interface design as a solution to extra-column dispersion. How would a column-in-valve format practically change instrument architecture, and what trade-offs (cost, robustness, ease of use) come with abandoning the traditional hi-fi tower configuration?
The column-in-valve design inserts the column directly into the injection valve, like put the plug into a socket. If all the modules in an HPLC system could be connected directly to one another in this “plug-and-fit” configuration, as Pepermans et al. suggested,15 we could significantly reduce extra-column band broadening by eliminating most of the connection tubing and unions. The column cartridges developed for the Axcend Focus LC can be considered a commercially available example of this kind of design. Ultimately, this type of architecture could make the entire LC system much more compact, and we might eventually see far fewer external tubes on the instrument.
However, there are several significant barriers to implementing a “plug-and-fit” design across an entire LC system and bringing it to the market. First, such a system would require the redesign of almost every module in a conventional LC instrument. This would involve substantial development and manufacturing investment, while the target market would likely remain relatively small because only a limited number of applications require extremely tight control of extra-column band broadening. Second, the pressure resistance and long-term reliability of these direct interfaces would need to be carefully demonstrated, particularly for ultra-high-pressure applications. Third, such an architecture would require a dedicated and highly isolated supply chain, which could inevitably increase the cost/price of the associated modules, interfaces, and accessories. Finally, although this highly integrated design could substantially improve system compactness and performance, it would also reduce flexibility. Users would have fewer options to modify, replace, or repair individual components themselves. So, there is a trade-off between achieving a highly optimized, integrated LC system and maintaining the modularity and flexibility that users currently value.
Viscous heating is flagged as a key limitation restricting nanoparticle columns to short capillaries under 0.2 mm i.d. How does viscous heating specifically degrade separation efficiency at these length and pressure combinations, and are there mobile phase or thermal management strategies that could mitigate it?
If we consider a separation performed at a fixed pressure and a given efficiency (that is a specific point on the kinetic performance limit), a nanoparticle-packed column will generally require a much higher pressure drop per unit length, or ΔP/L, because the optimal column is much shorter. This leads to a substantial increasing in viscous heating.16 Increasing the column length while keeping the pressure constant would push the separation into the B-term-dominated region of the van Deemter curve. This is unfavorable from a chromatographic efficiency perspective. Thus, it is not an effective way to solve the problem of viscous heating.
Therefore, the most straightforward way to minimize viscous heating is to reduce the inner diameter of the column. The temperature increases due to viscous heating scales with the square of the column diameter. For example, reducing the column i.d. from 2.1 mm to 0.2 mm could theoretically reduce the temperature increase by more than a factor of 100. If we still want to operate nanoparticle-packed columns at the microLC or analytical scale, pre-cooling the mobile phase, using active cooling unit17 or operating at a lower oven temperature,18 could help compensate for the effects of viscous heating. However, these approaches do not effectively eliminate the temperature gradient that develops across the column. Therefore, they are no ideal solutions for addressing the fundamental problem of viscous heating in nanoparticle-packed columns.
You note that ultraviolet (UV) detectors remain important given cost and industry adoption, despite mass spectrometry (MS) offering better post-column dispersion control. What specific design changes would UV detectors need to become viable for sub-1 µm particle columns, and is this a solvable optics/flow-cell problem or a more fundamental sensitivity limitation?
A 3 nL UV flow cell (smallest UV cell that is commercially available) could introduce an extra-column dispersion of only about 0.75 nL²,19 which sounds quite promising. However, this value is assumed without accounting for the additional dispersion introduced by the connections and tubing. Considering that we only have a tolerance of roughly 2-6 nL² for a column packed with 750 nm particles, even 0.75 nL² from the flow cell alone represents a significant fraction of the total dispersion budget. This means that we would ideally need to reduce the flow-cell volume even further, either by shortening the optical path length or reducing the inner diameter of the flow cell. However, this creates a direct trade-off with UV sensitivity because reducing the optical path length also reduces the absorbance signal. Another factor limiting sensitivity is the use of non-porous particles in a capillary-column format. Because non-porous particles have a lower surface area, and the capillary format has a relatively small column volume, the allowable sample load is more limited, which means that the analyte concentration reaching the detector may be lower.
These are primarily engineering challenges that could potentially be addressed by redesigning the UV flow cell and its interface with the column. In addition, the development of pressure-resistant porous nanoparticles could potentially improve UV sensitivity in two ways: by providing a higher sample-loading capacity and by increasing the tolerance for extra-column band broadening. Although LC-MS analysis of complex biological samples may ultimately be the most compelling application for nanoparticle-packed columns, developing a practical LC-UV method could significantly broaden their industrial value.
The modified Knox model underlies your kinetic-plot predictions. What assumptions in this model might not hold at the sub-1 µm scale, and how confident are you that experimental data would match these theoretical performance gains?
We expect this model to work well for randomly packed particle columns. The modification to the eddy-dispersion term was developed based on experimental results from commercial particle-packed columns,20 and we have already obtained experimental results for submicron particle-packed columns that agree surprisingly well with the theoretical predictions.
However, for self-assembled columns, this modification may no longer be applicable because of the substantially different packing morphology. The more ordered packing structure could fundamentally change the mechanisms contributing to eddy dispersion, so this will need to be validated experimentally.
Packing stable beds with sub-1 µm particles is described as "technically challenging." What specifically goes wrong during packing at this scale (e.g., aggregation, bed heterogeneity, void formation), and what packing strategies show the most promise for producing reproducible, stable beds?
Smaller particles build up pressure much more rapidly than microparticles during conventional high-pressure slurry packing. This can lead to a sharp reduction in packing flow rate and make it difficult to maintain a stable and consistent packing process using conventional methods.21
To address this issue, a packing system capable of operating at a higher-pressure limit would be preferable for constant-flow-rate packing.22,23 Alternatively, a low-flow-rate packing protocol could be used to maintain a stable packing process within the available pressure limit.24 Centrifugal or sedimentation-based packing methods provide another possible solution because they can avoid the need for precise flow-rate control and eliminate back-pressure-related limitations.25 This approach has recently been demonstrated for nanoparticle column packing in a study by Sun et al.26
If both the pressure and dispersion challenges were solved tomorrow, what analytical application do you think would benefit most immediately from sub-1 µm/nanoparticle-packed columns, and why?
The key advantage of nanoparticle-packed columns is their potential to achieve significantly faster separations while maintaining high resolving power. This advantage is particularly relevant for analytes with low diffusion coefficients, but it may be less applicable to very large biomolecules, such as intact proteins or long nucleic acids, because of potential shear-induced and hydrodynamic effects. Therefore, the most promising applications for nanoparticle-packed columns would involve samples containing moderate size of molecules, highly complex mixtures that require high separation efficiency, and large sample amounts that demand high-throughput analysis. All these requirements point toward proteomics, particularly emerging applications such as single-cell proteomics, spatial proteomics, and clinical proteomics.
It is also worth noting that the challenges associated with pressure and extra-column dispersion are not necessarily insurmountable problems that would require many years of development. Even at the pressure limits currently available, around 1500 bar, our theoretical prediction suggests that nanoparticle-packed columns could provide a significant improvement in separation speed more than two times. Furthermore, in a nanoLC-MS setup operated under gradient conditions, the extra-column dispersion problem can be substantially mitigated through sample focusing and a well configurated MS interface, as has already been demonstrated in the literature.27 This compatibility with nanoLC-MS is particularly well aligned with the requirements for proteomics.
References
- Wen, H.; Broeckhoven, K.; Eeltink, S. Opportunities and Challenges of Sub-1μm Particle Columns in Ultra-high-pressure Liquid Chromatography: A Theoretical Framework. J. Chromatogr. A 2026, 1785, 467310. DOI:
10.1016/j.chroma.2026.467310 - MacNair, J. E.; Patel, K. D.; Jorgenson, J. W. Ultrahigh-pressure Reversed-phase Capillary Liquid Chromatography: Iisocratic and Gradient Elution Using Columns Packed with 1.0-μm Particles, Anal. Chem. 1999, 71, 700–708. DOI:
10.1021/ac9807013 - MacNair, J. E.; Lewis, K. C.; Jorgenson, J. W. Ultrahigh-pressure Reversed-phase Liquid Chromatography in Packed Capillary Columns, Anal. Chem. 1997, 69, 983–989. DOI:
10.1021/ac961094r - Pauw, R. D.; Degreef, B.; Ritchie, H. et al. Extending the Limits of Operating Pressure of Narrow-bore Column Liquid Chromatography Instrumentation, J. Chromatogr. A 2014, 1347, 56–62. DOI:
10.1016/j.chroma.2014.04.056 - Grinias, K. M.; Godinho, J. M.; Franklin, E. G. et al. Development of a 45kpsi Ultrahigh Pressure Liquid Chromatography Instrument for Gradient Separations of Peptides Using Long Microcapillary Columns and Sub-2μm Particles. J Chromatogr A2016, 1469, 60-67.
10.1016/j.chroma.2016.09.053 - De Pauw, R.; Swier, T.; Degreef, B. et al. On the Feasibility to Conduct Gradient Lquid Chromatography Separations in Narrow-bore Columns at Pressures up to 2000 Bar. J Chromatogr A2016,1473, 48–55. DOI:
10.1016/j.chroma.2016.10.008 - Xiang, Y.; Maynes, D. R.; Lee, M. L. Safety Concerns in Ultrahigh Pressure Capillary Liquid Chromatography Using Air-driven Pumps. J Chromatogr A 2003, 991 (2), 189-196. DOI:
10.1016/S0021-9673(03)00171-7 - Marlow F; Muldarisnur; Sharifi P, Brinkmann R, Mendive C. Opals: Status and Prospects. Angew Chem Int Ed Engl. 2009, 48 (34), 6212-6233. DOI:10.1002/anie.200900210
- Wei, B.; Rogers, B. J.; Wirth, M. J. Slip Flow in Colloidal Crystals for Ultraefficient Chromatography. J Am Chem Soc. 2012, 134 (26), 10780-10782. DOI:
10.1021/ja304177m - Wu, Z.; Wei, B.; Zhang, X. et al. Efficient Separations of Intact Proteins Using Slip-flow with Nano-liquid Chromatography-Mass Spectrometry. Anal Chem. 2014, 86 (3), 1592-1598. DOI:
10.1021/ac403233d - Liu, Q.; Yan, K.; Chen, Q. et al. Preparation of Silica Colloidal Crystal Column and its Application in Pressurized Capillary Electrochromatography. J Chromatogr A 2019, 1587, 172-179. DOI:
10.1016/j.chroma.2018.12.016 - Vissers, J. P. C.; Claessens,H. A.; Cramers, C. A. Microcolumn Liquid Chromatography: Instrumentation, Detection and Applications, J. Chromatogr. A 1997, 779, 1–28.
10.1016/S0021-9673(97)00422-6 - Desmet, G.; Broeckhoven, K. Extra-column Band Broadening Effects in Contemporary Liquid Chromatography: Causes and Solutions, TrAC Trends Anal. Chem. 2019, 119, 115619. DOI:
10.1016/j.trac.2019.115619 - Hara, T.; Nakahara, K.; Broeckhoven, K. et al. Investigation and Optimization of the Extra-Column Band Broadening in Micro-flow Capillary Liquid Chromatography. J. Chromatogr. A 2025, 1748, 465805. DOI:
10.1016/j.chroma.2025.465805 - Pepermans, V.; Rerick, M. T.; Degreef, B. et al. Column-in-valve Designs to Minimize Extra-column Volumes. J Chromatogr A 2021, 1637, 461779. DOI:
10.1016/j.chroma.2020.461779 - de Villiers, A.; Lauer, H.; Szucs, R. et al. Influence of Frictional Heating on Temperature Gradients in Ultra-high-pressure Liquid Chromatography on 2.1mm I.D. Columns. J Chromatogr A 2006, 1113 (1-2), 84-91. DOI:
10.1016/j.chroma.2006.01.120 - Broeckhoven, K.; Billen, J.; Verstraeten, M. et al. Towards a Solution for Viscous Heating in Ultra-high pressure Liquid Chromatography Using Intermediate Cooling. J Chromatogr A 2010, 1217 (13), 2022-2031. DOI:
10.1016/j.chroma.2010.01.072 - Nováková, L.; Veuthey, J. L.; Guillarme, D. Practical Method Transfer from High Performance Liquid Chromatography to Ultra-high Performance Liquid Chromatography: The Importance of Frictional Heating. J Chromatogr A 2011, 1218 (44), 7971-7981. DOI: DOI:
10.1016/j.chroma.2011.08.096 - Song, H.; Vanderheyden, Y.; Adams, E. et al. Extensive Database of Liquid Phase Diffusion Coefficients of Some Frequently Used Test Molecules in Reversed-phase Liquid Chromatography and Hydrophilic Interaction Liquid Chromatography. J Chromatogr A 2016, 1455, 102-112. DOI:
10.1016/j.chroma.2016.05.054 - Desmet, G.; Song, H.; Makey, D. et al. Experimental Investigation of the Retention Factor Dependency of Eddy Dispersion in Packed Bed Columns and Relation to Knox's Empirical Model Parameters. J Chromatogr A 2020, 1626, 461339. DOI:
10.1016/j.chroma.2020.461339 - Blue, L. E.; Jorgenson, J. W. 1.1 μm Superficially Porous Particles for Liquid Chromatography: Part II: Column Packing and Chromatographic Performance. J Chromatogr A 2015, 1380, 71-80. DOI:
10.1016/j.chroma.2014.12.055 - Müller-Reif, J. B.; Hansen, F. M.; Schweizer, L. et al. A New Parallel High-Pressure Packing System Enables Rapid Multiplexed Production of Capillary Columns. Mol Cell Proteomics 2021, 20, 100082. DOI:
10.1016/j.mcpro.2021.100082 - Shishkova, E.; Hebert, A. S.; Westphall, M. S. et al. Ultra-High Pressure (>30,000 psi) Packing of Capillary Columns Enhancing Depth of Shotgun Proteomic Analyses. Anal Chem. 2018, 90 (19), 11503-11508. DOI:
10.1021/acs.analchem.8b02766 - Kovalchuk, S. I.; Jensen, O. N.; Rogowska-Wrzesinska, A. FlashPack: Fast and Simple Preparation of Ultrahigh-performance Capillary Columns for LC-MS. Mol Cell Proteomics 2019, 18 (2), 383-390. DOI:
10.1074/mcp.TIR118.000953 - Liu, Y.; Wen, H.; Chen, S. et al. Mass Fabrication of Capillary Columns Based on Centrifugal Packing. Anal Chem.2022, 94 (23), 8126-8131. DOI:
10.1021/acs.analchem.2c00442 - Sun, H.; Wang, C.; Guo, K. et al. High-Throughput Single-Cell-Resolved Spatial Proteomics Enabled by an Ordered Colloidal Crystal Column. Angew Chem Int Ed Engl. 2026, 65 (24), e8045649. DOI:
10.1002/anie.8045649 - Rogers, B. A.; Wu, Z.;Wei, B. et al. Submicrometer Particles and Slip Flow in Liquid Chromatography. Anal Chem. 2015, 87 (5), 2520-2526. DOI:
10.1021/ac504683d